Simple and rapid validated HPLC-fluorescence determination of perampanel in the plasma of patients with epilepsy
We present a simple and fast high-performance liquid chromatography method with fluorescence detection for the determination of the antiepileptic drug perampanel in human plasma. The chromatographic separation was performed on a Kinetex PFP (100 × 2.6 mm, 4.6 µm) column, using a mobile phase of sodi...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722279/ https://www.ncbi.nlm.nih.gov/pubmed/29234708 http://dx.doi.org/10.1016/j.plabm.2017.11.003 |
Sumario: | We present a simple and fast high-performance liquid chromatography method with fluorescence detection for the determination of the antiepileptic drug perampanel in human plasma. The chromatographic separation was performed on a Kinetex PFP (100 × 2.6 mm, 4.6 µm) column, using a mobile phase of sodium acetate 0.03 M pH 3.7 and acetonitrile (40/60, v/v), at a flow rate of 0.8 mL/min. Total chromatography time for each run was 5 min. Sample preparation (250 µL) involved only one simple precipitation step by acetonitrile spiked with mirtazapine as internal standard. The method was validated over a concentration range of 20–1000 ng/mL and successfully applied to measure perampanel concentrations in plasma samples obtained from patients with epilepsy. This assay combines the high specificity of fluorescence detection with a very simple and fast sample pretreatment and can offer real advantages over existing methods in terms of simplicity and transferability to a therapeutic drug monitoring setting. |
---|