An investigation of temporal resolution parameters in cine‐mode four‐dimensional computed tomography acquisition

The accuracy of four‐dimensional computed tomography (4DCT) imaging depends on temporal characteristics of the acquisition protocol—for example, the temporal spacing of the reconstructed images (also known as cine duration between images) and the gantry rotation speed. These parameters affect the te...

Descripción completa

Detalles Bibliográficos
Autores principales: Mutaf, Yildirim D., Brinkmann, Debra H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722349/
https://www.ncbi.nlm.nih.gov/pubmed/19020488
http://dx.doi.org/10.1120/jacmp.v9i4.2819
_version_ 1783284992722337792
author Mutaf, Yildirim D.
Brinkmann, Debra H.
author_facet Mutaf, Yildirim D.
Brinkmann, Debra H.
author_sort Mutaf, Yildirim D.
collection PubMed
description The accuracy of four‐dimensional computed tomography (4DCT) imaging depends on temporal characteristics of the acquisition protocol—for example, the temporal spacing of the reconstructed images (also known as cine duration between images) and the gantry rotation speed. These parameters affect the temporal resolution of 4DCT images, and a single default acquisition protocol, as commonly used in most clinics, may be suboptimal for a subset of respiratory motion characteristics. It could lead to substantial inaccuracies in target delineation. The aim of the present study was to evaluate the interplay between parameters affecting temporal resolution and the accuracy of the resulting images. We acquired 4DCT images of cylindrical phantoms under repetitive motion induced by a translation platform. Acquisition settings varied with respect to temporal spacing, gantry rotation speed, and motion period of the phantoms. Reconstructed images were sorted into 10 phase bins and were compared to static phantom images acquired at corresponding positions of the respiration phase. Acquisitions with different temporal spacing did not play a significant role in the amount of motion observed in full‐cycle maximum intensity projection images. Target delineation accuracy at end‐of‐inhalation phase was observed to be constant up to a threshold in the value of the reconstruction interval, beyond which it varied arbitrarily. This threshold was found to be correlated with the number of phase bins and the motion period. No observable variations were noted with images from the end of exhalation when temporal spacing was varied. Target delineation accuracy was observed to be enhanced in acquisitions using faster gantry rotation speeds. An evaluation of the acquisition parameters needs to be performed depending on the period of the motion and limiting factors such as the availability of acquisition settings, X‐ray tube workload, image storage, and processing power. PACS numbers: 87.53.Xd, 87.57.‐s, 87.57.Gg, 87.59.Fm
format Online
Article
Text
id pubmed-5722349
institution National Center for Biotechnology Information
language English
publishDate 2008
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57223492018-04-02 An investigation of temporal resolution parameters in cine‐mode four‐dimensional computed tomography acquisition Mutaf, Yildirim D. Brinkmann, Debra H. J Appl Clin Med Phys Medical Imaging The accuracy of four‐dimensional computed tomography (4DCT) imaging depends on temporal characteristics of the acquisition protocol—for example, the temporal spacing of the reconstructed images (also known as cine duration between images) and the gantry rotation speed. These parameters affect the temporal resolution of 4DCT images, and a single default acquisition protocol, as commonly used in most clinics, may be suboptimal for a subset of respiratory motion characteristics. It could lead to substantial inaccuracies in target delineation. The aim of the present study was to evaluate the interplay between parameters affecting temporal resolution and the accuracy of the resulting images. We acquired 4DCT images of cylindrical phantoms under repetitive motion induced by a translation platform. Acquisition settings varied with respect to temporal spacing, gantry rotation speed, and motion period of the phantoms. Reconstructed images were sorted into 10 phase bins and were compared to static phantom images acquired at corresponding positions of the respiration phase. Acquisitions with different temporal spacing did not play a significant role in the amount of motion observed in full‐cycle maximum intensity projection images. Target delineation accuracy at end‐of‐inhalation phase was observed to be constant up to a threshold in the value of the reconstruction interval, beyond which it varied arbitrarily. This threshold was found to be correlated with the number of phase bins and the motion period. No observable variations were noted with images from the end of exhalation when temporal spacing was varied. Target delineation accuracy was observed to be enhanced in acquisitions using faster gantry rotation speeds. An evaluation of the acquisition parameters needs to be performed depending on the period of the motion and limiting factors such as the availability of acquisition settings, X‐ray tube workload, image storage, and processing power. PACS numbers: 87.53.Xd, 87.57.‐s, 87.57.Gg, 87.59.Fm John Wiley and Sons Inc. 2008-10-29 /pmc/articles/PMC5722349/ /pubmed/19020488 http://dx.doi.org/10.1120/jacmp.v9i4.2819 Text en © 2008 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Mutaf, Yildirim D.
Brinkmann, Debra H.
An investigation of temporal resolution parameters in cine‐mode four‐dimensional computed tomography acquisition
title An investigation of temporal resolution parameters in cine‐mode four‐dimensional computed tomography acquisition
title_full An investigation of temporal resolution parameters in cine‐mode four‐dimensional computed tomography acquisition
title_fullStr An investigation of temporal resolution parameters in cine‐mode four‐dimensional computed tomography acquisition
title_full_unstemmed An investigation of temporal resolution parameters in cine‐mode four‐dimensional computed tomography acquisition
title_short An investigation of temporal resolution parameters in cine‐mode four‐dimensional computed tomography acquisition
title_sort investigation of temporal resolution parameters in cine‐mode four‐dimensional computed tomography acquisition
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722349/
https://www.ncbi.nlm.nih.gov/pubmed/19020488
http://dx.doi.org/10.1120/jacmp.v9i4.2819
work_keys_str_mv AT mutafyildirimd aninvestigationoftemporalresolutionparametersincinemodefourdimensionalcomputedtomographyacquisition
AT brinkmanndebrah aninvestigationoftemporalresolutionparametersincinemodefourdimensionalcomputedtomographyacquisition
AT mutafyildirimd investigationoftemporalresolutionparametersincinemodefourdimensionalcomputedtomographyacquisition
AT brinkmanndebrah investigationoftemporalresolutionparametersincinemodefourdimensionalcomputedtomographyacquisition