Observed intercamera variability in clinically relevant performance characteristics for Siemens Symbia gamma cameras

We conducted an evaluation of the intercamera (i.e., between cameras) variability in clinically relevant performance characteristics for Symbia gamma cameras (Siemens Medical Solutions, Malvern, PA) based on measurements made using nine separate systems. The significance of the observed intercamera...

Descripción completa

Detalles Bibliográficos
Autores principales: Kappadath, S. Cheenu, Erwin, William D., Wendt, Richard E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2006
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722386/
https://www.ncbi.nlm.nih.gov/pubmed/17533358
http://dx.doi.org/10.1120/jacmp.v7i4.2376
_version_ 1783285001643622400
author Kappadath, S. Cheenu
Erwin, William D.
Wendt, Richard E.
author_facet Kappadath, S. Cheenu
Erwin, William D.
Wendt, Richard E.
author_sort Kappadath, S. Cheenu
collection PubMed
description We conducted an evaluation of the intercamera (i.e., between cameras) variability in clinically relevant performance characteristics for Symbia gamma cameras (Siemens Medical Solutions, Malvern, PA) based on measurements made using nine separate systems. The significance of the observed intercamera variability was determined by comparing it to the intracamera (i.e., within a single camera) variability. Measurements of performance characteristics were based on the standards of the National Electrical Manufacturers Association and reports 6, 9, 22, and 52 from the American Association of Physicists in Medicine. All measurements were performed using [Formula: see text] (except [Formula: see text] used for extrinsic resolution) and low‐energy, high‐resolution collimation. Of the nine cameras, four have crystals 3/8 in. thick and five have crystals 5/8 in. thick. We evaluated intrinsic energy resolution, intrinsic and extrinsic spatial resolution, intrinsic integral and differential flood uniformity over the useful field‐of‐view, count rate at 20% count loss, planar sensitivity, single‐photon emission computed tomography (SPECT) resolution, and SPECT integral uniformity. The intracamera variability was estimated by repeated measurements of the performance characteristics on a single system. The significance of the observed intercamera variability was evaluated using the two‐tailed F distribution. The planar sensitivity of the gamma cameras tested was found be variable at the 99.8% confidence level for both the 3/8‐in. and 5/8‐in. crystal systems. The integral uniformity and energy resolution were found to be variable only for the 5/8‐in. crystal systems at the 98% and 90% confidence level, respectively. All other performance characteristics tested exhibited no significant variability between camera systems. The measured variability reported here could perhaps be used to define nominal performance values of Symbia gamma cameras for planar and SPECT imaging. PACS numbers: 87.62.+n, 87.58.Pm, 87.58.Ce
format Online
Article
Text
id pubmed-5722386
institution National Center for Biotechnology Information
language English
publishDate 2006
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57223862018-04-02 Observed intercamera variability in clinically relevant performance characteristics for Siemens Symbia gamma cameras Kappadath, S. Cheenu Erwin, William D. Wendt, Richard E. J Appl Clin Med Phys Medical Imaging We conducted an evaluation of the intercamera (i.e., between cameras) variability in clinically relevant performance characteristics for Symbia gamma cameras (Siemens Medical Solutions, Malvern, PA) based on measurements made using nine separate systems. The significance of the observed intercamera variability was determined by comparing it to the intracamera (i.e., within a single camera) variability. Measurements of performance characteristics were based on the standards of the National Electrical Manufacturers Association and reports 6, 9, 22, and 52 from the American Association of Physicists in Medicine. All measurements were performed using [Formula: see text] (except [Formula: see text] used for extrinsic resolution) and low‐energy, high‐resolution collimation. Of the nine cameras, four have crystals 3/8 in. thick and five have crystals 5/8 in. thick. We evaluated intrinsic energy resolution, intrinsic and extrinsic spatial resolution, intrinsic integral and differential flood uniformity over the useful field‐of‐view, count rate at 20% count loss, planar sensitivity, single‐photon emission computed tomography (SPECT) resolution, and SPECT integral uniformity. The intracamera variability was estimated by repeated measurements of the performance characteristics on a single system. The significance of the observed intercamera variability was evaluated using the two‐tailed F distribution. The planar sensitivity of the gamma cameras tested was found be variable at the 99.8% confidence level for both the 3/8‐in. and 5/8‐in. crystal systems. The integral uniformity and energy resolution were found to be variable only for the 5/8‐in. crystal systems at the 98% and 90% confidence level, respectively. All other performance characteristics tested exhibited no significant variability between camera systems. The measured variability reported here could perhaps be used to define nominal performance values of Symbia gamma cameras for planar and SPECT imaging. PACS numbers: 87.62.+n, 87.58.Pm, 87.58.Ce John Wiley and Sons Inc. 2006-11-28 /pmc/articles/PMC5722386/ /pubmed/17533358 http://dx.doi.org/10.1120/jacmp.v7i4.2376 Text en © 2006 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Kappadath, S. Cheenu
Erwin, William D.
Wendt, Richard E.
Observed intercamera variability in clinically relevant performance characteristics for Siemens Symbia gamma cameras
title Observed intercamera variability in clinically relevant performance characteristics for Siemens Symbia gamma cameras
title_full Observed intercamera variability in clinically relevant performance characteristics for Siemens Symbia gamma cameras
title_fullStr Observed intercamera variability in clinically relevant performance characteristics for Siemens Symbia gamma cameras
title_full_unstemmed Observed intercamera variability in clinically relevant performance characteristics for Siemens Symbia gamma cameras
title_short Observed intercamera variability in clinically relevant performance characteristics for Siemens Symbia gamma cameras
title_sort observed intercamera variability in clinically relevant performance characteristics for siemens symbia gamma cameras
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722386/
https://www.ncbi.nlm.nih.gov/pubmed/17533358
http://dx.doi.org/10.1120/jacmp.v7i4.2376
work_keys_str_mv AT kappadathscheenu observedintercameravariabilityinclinicallyrelevantperformancecharacteristicsforsiemenssymbiagammacameras
AT erwinwilliamd observedintercameravariabilityinclinicallyrelevantperformancecharacteristicsforsiemenssymbiagammacameras
AT wendtricharde observedintercameravariabilityinclinicallyrelevantperformancecharacteristicsforsiemenssymbiagammacameras