Cargando…
Initial evaluation of a four‐dimensional computed tomography system, using a programmable motor
A dynamic lung tumor phantom was used to investigate the geometric reconstruction accuracy of a commercial four‐dimensional computed tomography (4D‐CT) system. A ball filled with resin, embedded in a cork cube, was placed on a moving platform. Various realistic antero‐posterior (AP) motions were pro...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2006
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722388/ https://www.ncbi.nlm.nih.gov/pubmed/17533356 http://dx.doi.org/10.1120/jacmp.v7i4.2301 |
_version_ | 1783285002135404544 |
---|---|
author | Simon, Luc Giraud, Philippe Servois, Vincent Rosenwald, Jean‐Claude |
author_facet | Simon, Luc Giraud, Philippe Servois, Vincent Rosenwald, Jean‐Claude |
author_sort | Simon, Luc |
collection | PubMed |
description | A dynamic lung tumor phantom was used to investigate the geometric reconstruction accuracy of a commercial four‐dimensional computed tomography (4D‐CT) system. A ball filled with resin, embedded in a cork cube, was placed on a moving platform. Various realistic antero‐posterior (AP) motions were programmed to reproduce the respiratory motion of a lung tumor. Several three‐dimensional (3D) CT and 4D‐CT images of this moving object were acquired and compared using different acquisition parameters. Apparent volume and diameter of the ball were measured and compared to the real values. The position of two points (the AP limits of the ball) during the motion in the coordinate system of the CT scanner were also compared with the expected values. Volume error was shown to increase with object speed. However, although the volume error was associated with intraslice artifacts, it did not reflect large interslice inconstancies in object position and should not be used as an indicator of image accuracy. The 3D‐CT gave a random position of the tumor along the phantom excursion; accuracy in the assessment of position by 4D‐CT ranged from 0.4 mm to 2.6 mm during extreme phases of breathing. We used average projection (AVE) and maximum intensity projection (MIP) algorithms available on the commercial software to create internal target volumes (ITVs) by merging gross tumor volume (GTV) images at various respiratory phases. The ITVs were compared to a theoretical value computed from the programmed ball excursion. The ITVs created from the MIP algorithm were closer to the theoretical value (within 12%) than were those created from the AVE algorithm (within 40%). PACS numbers: 87.53.Xd, 87.56.Mp, 87.57.Ce, 87.59.Fm |
format | Online Article Text |
id | pubmed-5722388 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2006 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57223882018-04-02 Initial evaluation of a four‐dimensional computed tomography system, using a programmable motor Simon, Luc Giraud, Philippe Servois, Vincent Rosenwald, Jean‐Claude J Appl Clin Med Phys Radiation Oncology Physics A dynamic lung tumor phantom was used to investigate the geometric reconstruction accuracy of a commercial four‐dimensional computed tomography (4D‐CT) system. A ball filled with resin, embedded in a cork cube, was placed on a moving platform. Various realistic antero‐posterior (AP) motions were programmed to reproduce the respiratory motion of a lung tumor. Several three‐dimensional (3D) CT and 4D‐CT images of this moving object were acquired and compared using different acquisition parameters. Apparent volume and diameter of the ball were measured and compared to the real values. The position of two points (the AP limits of the ball) during the motion in the coordinate system of the CT scanner were also compared with the expected values. Volume error was shown to increase with object speed. However, although the volume error was associated with intraslice artifacts, it did not reflect large interslice inconstancies in object position and should not be used as an indicator of image accuracy. The 3D‐CT gave a random position of the tumor along the phantom excursion; accuracy in the assessment of position by 4D‐CT ranged from 0.4 mm to 2.6 mm during extreme phases of breathing. We used average projection (AVE) and maximum intensity projection (MIP) algorithms available on the commercial software to create internal target volumes (ITVs) by merging gross tumor volume (GTV) images at various respiratory phases. The ITVs were compared to a theoretical value computed from the programmed ball excursion. The ITVs created from the MIP algorithm were closer to the theoretical value (within 12%) than were those created from the AVE algorithm (within 40%). PACS numbers: 87.53.Xd, 87.56.Mp, 87.57.Ce, 87.59.Fm John Wiley and Sons Inc. 2006-11-28 /pmc/articles/PMC5722388/ /pubmed/17533356 http://dx.doi.org/10.1120/jacmp.v7i4.2301 Text en © 2006 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Simon, Luc Giraud, Philippe Servois, Vincent Rosenwald, Jean‐Claude Initial evaluation of a four‐dimensional computed tomography system, using a programmable motor |
title | Initial evaluation of a four‐dimensional computed tomography system, using a programmable motor |
title_full | Initial evaluation of a four‐dimensional computed tomography system, using a programmable motor |
title_fullStr | Initial evaluation of a four‐dimensional computed tomography system, using a programmable motor |
title_full_unstemmed | Initial evaluation of a four‐dimensional computed tomography system, using a programmable motor |
title_short | Initial evaluation of a four‐dimensional computed tomography system, using a programmable motor |
title_sort | initial evaluation of a four‐dimensional computed tomography system, using a programmable motor |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722388/ https://www.ncbi.nlm.nih.gov/pubmed/17533356 http://dx.doi.org/10.1120/jacmp.v7i4.2301 |
work_keys_str_mv | AT simonluc initialevaluationofafourdimensionalcomputedtomographysystemusingaprogrammablemotor AT giraudphilippe initialevaluationofafourdimensionalcomputedtomographysystemusingaprogrammablemotor AT servoisvincent initialevaluationofafourdimensionalcomputedtomographysystemusingaprogrammablemotor AT rosenwaldjeanclaude initialevaluationofafourdimensionalcomputedtomographysystemusingaprogrammablemotor |