Measurement of the interplay effect in lung IMRT treatment using EDR2 films

Intrafraction organ motion during the dynamic delivery of intensity‐modulated radiation therapy (IMRT) treatment of lung tumors may cause unexpected hot/cold spots within the target volume, due to the interplay effect between tumor motion and multileaf collimator (MLC) leaf motion. In the past, this...

Descripción completa

Detalles Bibliográficos
Autores principales: Berbeco, Ross I., Pope, Cynthia J., Jiang, Steve B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2006
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722391/
https://www.ncbi.nlm.nih.gov/pubmed/17533350
http://dx.doi.org/10.1120/jacmp.v7i4.2222
_version_ 1783285002837950464
author Berbeco, Ross I.
Pope, Cynthia J.
Jiang, Steve B.
author_facet Berbeco, Ross I.
Pope, Cynthia J.
Jiang, Steve B.
author_sort Berbeco, Ross I.
collection PubMed
description Intrafraction organ motion during the dynamic delivery of intensity‐modulated radiation therapy (IMRT) treatment of lung tumors may cause unexpected hot/cold spots within the target volume, due to the interplay effect between tumor motion and multileaf collimator (MLC) leaf motion. In the past, this has been investigated through theoretical analysis, computer simulation, and experimental measurement using an ionization chamber dosimeter. In the work presented here, the interplay effect was studied experimentally in 2D, using Kodak EDR2 films. A five‐field lung IMRT plan was delivered to a solid water phantom with embedded film. The phantom was placed on a motor‐driven platform with a sinusoidal motion to simulate the respiration‐induced tumor motion. The delivery of each field began at one of eight equally spaced initial breathing phases. The dose distribution for each treatment fraction was estimated by combining the dose distributions for all fields with randomly sampled initial breathing phases. The dose variation caused by the interplay effect was estimated by looking at the dose values from 1000 trials of 30 fractions. It was found that, on a day‐to‐day basis, the standard deviation of the dose to a given pixel in the high‐dose region could be as high as 2% to 4% due to the motion interplay effect. After 30 fractions, the standard deviation in the dose to each pixel is reduced to 0.4% to 0.7%. However, compared to the static delivery, the dose distribution from a 30‐fraction case in the presence of motion shows some underdosing in the region of interest. We found that the maximum dose in the target remains within 1% of the maximum dose in the static case, but the minimum dose in the target is most likely to be about 6% lower than the static case. Our results indicate that there can be some underdosing of the tumor due to the interplay effect in lung IMRT delivery over the entire course of a 30‐fraction treatment. PACS number: 87.53.Mr
format Online
Article
Text
id pubmed-5722391
institution National Center for Biotechnology Information
language English
publishDate 2006
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57223912018-04-02 Measurement of the interplay effect in lung IMRT treatment using EDR2 films Berbeco, Ross I. Pope, Cynthia J. Jiang, Steve B. J Appl Clin Med Phys Radiation Oncology Physics Intrafraction organ motion during the dynamic delivery of intensity‐modulated radiation therapy (IMRT) treatment of lung tumors may cause unexpected hot/cold spots within the target volume, due to the interplay effect between tumor motion and multileaf collimator (MLC) leaf motion. In the past, this has been investigated through theoretical analysis, computer simulation, and experimental measurement using an ionization chamber dosimeter. In the work presented here, the interplay effect was studied experimentally in 2D, using Kodak EDR2 films. A five‐field lung IMRT plan was delivered to a solid water phantom with embedded film. The phantom was placed on a motor‐driven platform with a sinusoidal motion to simulate the respiration‐induced tumor motion. The delivery of each field began at one of eight equally spaced initial breathing phases. The dose distribution for each treatment fraction was estimated by combining the dose distributions for all fields with randomly sampled initial breathing phases. The dose variation caused by the interplay effect was estimated by looking at the dose values from 1000 trials of 30 fractions. It was found that, on a day‐to‐day basis, the standard deviation of the dose to a given pixel in the high‐dose region could be as high as 2% to 4% due to the motion interplay effect. After 30 fractions, the standard deviation in the dose to each pixel is reduced to 0.4% to 0.7%. However, compared to the static delivery, the dose distribution from a 30‐fraction case in the presence of motion shows some underdosing in the region of interest. We found that the maximum dose in the target remains within 1% of the maximum dose in the static case, but the minimum dose in the target is most likely to be about 6% lower than the static case. Our results indicate that there can be some underdosing of the tumor due to the interplay effect in lung IMRT delivery over the entire course of a 30‐fraction treatment. PACS number: 87.53.Mr John Wiley and Sons Inc. 2006-11-28 /pmc/articles/PMC5722391/ /pubmed/17533350 http://dx.doi.org/10.1120/jacmp.v7i4.2222 Text en © 2006 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Berbeco, Ross I.
Pope, Cynthia J.
Jiang, Steve B.
Measurement of the interplay effect in lung IMRT treatment using EDR2 films
title Measurement of the interplay effect in lung IMRT treatment using EDR2 films
title_full Measurement of the interplay effect in lung IMRT treatment using EDR2 films
title_fullStr Measurement of the interplay effect in lung IMRT treatment using EDR2 films
title_full_unstemmed Measurement of the interplay effect in lung IMRT treatment using EDR2 films
title_short Measurement of the interplay effect in lung IMRT treatment using EDR2 films
title_sort measurement of the interplay effect in lung imrt treatment using edr2 films
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722391/
https://www.ncbi.nlm.nih.gov/pubmed/17533350
http://dx.doi.org/10.1120/jacmp.v7i4.2222
work_keys_str_mv AT berbecorossi measurementoftheinterplayeffectinlungimrttreatmentusingedr2films
AT popecynthiaj measurementoftheinterplayeffectinlungimrttreatmentusingedr2films
AT jiangsteveb measurementoftheinterplayeffectinlungimrttreatmentusingedr2films