Evaluation of the analytical anisotropic algorithm in an extreme water–lung interface phantom using Monte Carlo dose calculations
Our study compares the performance of the analytical anisotropic algorithm (AAA), a new superposition–convolution algorithm recently implemented in the Eclipse (Varian Medical Systems, Palo Alto, CA) Integrated Treatment Planning System (TPS), to that of the pencil beam convolution (PBC) algorithm i...
Autores principales: | Gagné, Isabelle Marie, Zavgorodni, Sergei |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2007
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722400/ https://www.ncbi.nlm.nih.gov/pubmed/17592451 http://dx.doi.org/10.1120/jacmp.v8i1.2324 |
Ejemplares similares
-
Dosimetry of small bone joint calculated by the analytical anisotropic algorithm: a Monte Carlo evaluation using the EGSnrc
por: Chow, James C. L., et al.
Publicado: (2013) -
Validation of the RayStation Monte Carlo dose calculation algorithm using realistic animal tissue phantoms
por: Schreuder, Andries N., et al.
Publicado: (2019) -
Validation of the RayStation Monte Carlo dose calculation algorithm using a realistic lung phantom
por: Schreuder, Andries N., et al.
Publicado: (2019) -
Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method
por: Rodríguez, Miguel L., et al.
Publicado: (2002) -
A fast jaw‐tracking model for VMAT and IMRT Monte Carlo simulations
por: Townson, Reid, et al.
Publicado: (2018)