The use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device
A scanning liquid‐filled ionization chamber electronic portal imaging device (SLIC‐EPID) and extended dose range (EDR2) films were used to evaluate transmitted dose profiles for homogeneous and inhomogeneous phantoms. Calibrated ionization chamber measurements were used to convert the pixel values a...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2007
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722406/ https://www.ncbi.nlm.nih.gov/pubmed/17592449 http://dx.doi.org/10.1120/jacmp.v8i1.2305 |
_version_ | 1783285006524743680 |
---|---|
author | Mohammadi, Mohammad Bezak, Eva Reich, Paul |
author_facet | Mohammadi, Mohammad Bezak, Eva Reich, Paul |
author_sort | Mohammadi, Mohammad |
collection | PubMed |
description | A scanning liquid‐filled ionization chamber electronic portal imaging device (SLIC‐EPID) and extended dose range (EDR2) films were used to evaluate transmitted dose profiles for homogeneous and inhomogeneous phantoms. Calibrated ionization chamber measurements were used to convert the pixel values acquired from the electronic portal images to dose. Because SLIC‐EPID was developed to have a uniform response for all liquid ionization chambers, the off‐axis dose values were reconstructed using a correction factor matrix, defined as the ratio of the relative EDR2 film and the corresponding EPID dose values measured in air. The transmitted dose distributions in the EPID detector layer were also modeled using a [Formula: see text] treatment planning system (TPS: Philips Radiation Oncology Systems, Milpitas, CA). The gamma function algorithm was then used to assess agreement between transmitted dose distributions measured using a SLIC‐EPID and EDR2 film, and those calculated using the TPS. For homogenous and inhomogeneous phantoms, more than 90% agreement was achieved using gamma criteria of 2% and 3 mm and 3% and 2.5 mm respectively. Our results indicate that the calibration procedure proposed in the present study should be performed if SLIC‐EPID is to be used as a reliable two‐dimensional transmitted dosimeter for clinical purposes. PACS numbers: 87.53.Tf, 87.53.Oq |
format | Online Article Text |
id | pubmed-5722406 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2007 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57224062018-04-02 The use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device Mohammadi, Mohammad Bezak, Eva Reich, Paul J Appl Clin Med Phys Radiation Oncology Physics A scanning liquid‐filled ionization chamber electronic portal imaging device (SLIC‐EPID) and extended dose range (EDR2) films were used to evaluate transmitted dose profiles for homogeneous and inhomogeneous phantoms. Calibrated ionization chamber measurements were used to convert the pixel values acquired from the electronic portal images to dose. Because SLIC‐EPID was developed to have a uniform response for all liquid ionization chambers, the off‐axis dose values were reconstructed using a correction factor matrix, defined as the ratio of the relative EDR2 film and the corresponding EPID dose values measured in air. The transmitted dose distributions in the EPID detector layer were also modeled using a [Formula: see text] treatment planning system (TPS: Philips Radiation Oncology Systems, Milpitas, CA). The gamma function algorithm was then used to assess agreement between transmitted dose distributions measured using a SLIC‐EPID and EDR2 film, and those calculated using the TPS. For homogenous and inhomogeneous phantoms, more than 90% agreement was achieved using gamma criteria of 2% and 3 mm and 3% and 2.5 mm respectively. Our results indicate that the calibration procedure proposed in the present study should be performed if SLIC‐EPID is to be used as a reliable two‐dimensional transmitted dosimeter for clinical purposes. PACS numbers: 87.53.Tf, 87.53.Oq John Wiley and Sons Inc. 2007-02-28 /pmc/articles/PMC5722406/ /pubmed/17592449 http://dx.doi.org/10.1120/jacmp.v8i1.2305 Text en © 2007 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Mohammadi, Mohammad Bezak, Eva Reich, Paul The use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device |
title | The use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device |
title_full | The use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device |
title_fullStr | The use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device |
title_full_unstemmed | The use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device |
title_short | The use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device |
title_sort | use of extended dose range film for dosimetric calibration of a scanning liquid‐filled ionization chamber electronic portal imaging device |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722406/ https://www.ncbi.nlm.nih.gov/pubmed/17592449 http://dx.doi.org/10.1120/jacmp.v8i1.2305 |
work_keys_str_mv | AT mohammadimohammad theuseofextendeddoserangefilmfordosimetriccalibrationofascanningliquidfilledionizationchamberelectronicportalimagingdevice AT bezakeva theuseofextendeddoserangefilmfordosimetriccalibrationofascanningliquidfilledionizationchamberelectronicportalimagingdevice AT reichpaul theuseofextendeddoserangefilmfordosimetriccalibrationofascanningliquidfilledionizationchamberelectronicportalimagingdevice AT mohammadimohammad useofextendeddoserangefilmfordosimetriccalibrationofascanningliquidfilledionizationchamberelectronicportalimagingdevice AT bezakeva useofextendeddoserangefilmfordosimetriccalibrationofascanningliquidfilledionizationchamberelectronicportalimagingdevice AT reichpaul useofextendeddoserangefilmfordosimetriccalibrationofascanningliquidfilledionizationchamberelectronicportalimagingdevice |