Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop

Carbon fiber is commonly used in radiation therapy for treatment tabletops and various immobilization and support devices, partially because it is generally perceived to be almost radiotransparent to high‐energy photons. To avoid exposure to normal tissue during modern radiation therapy, one must de...

Descripción completa

Detalles Bibliográficos
Autores principales: Myint, W. Kenji, Niedbala, Malgorzata, Wilkins, David, Gerig, Lee H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2006
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722426/
https://www.ncbi.nlm.nih.gov/pubmed/17533341
http://dx.doi.org/10.1120/jacmp.v7i3.2247
_version_ 1783285011506528256
author Myint, W. Kenji
Niedbala, Malgorzata
Wilkins, David
Gerig, Lee H.
author_facet Myint, W. Kenji
Niedbala, Malgorzata
Wilkins, David
Gerig, Lee H.
author_sort Myint, W. Kenji
collection PubMed
description Carbon fiber is commonly used in radiation therapy for treatment tabletops and various immobilization and support devices, partially because it is generally perceived to be almost radiotransparent to high‐energy photons. To avoid exposure to normal tissue during modern radiation therapy, one must deliver the radiation from all gantry angles; hence, beams often transit the couch proximal to the patient. The effects of the beam attenuation by the support structure of the couch are often neglected in the planning process. In this study, we investigate the attenuation of 6‐MV and 18‐MV photon beams by a Medtec (Orange City, IA) carbon fiber couch. We have determined that neglecting the attenuation of oblique treatment fields by the carbon fiber couch can result in localized dose reduction from 4% to 16%, depending on energy, field size, and geometry. Further, we investigate the ability of a commercial treatment‐planning system (Theraplan Plus v3.8) to account for the attenuation by the treatment couch. Results show that incorporating the carbon fiber couch in the patient model reduces the dose error to less than 2%. The variation in dose reduction as a function of longitudinal couch position was also measured. In the triangular strut region of the couch, the attenuation varied [Formula: see text] following the periodic nature of the support structure. Based on these findings, we propose the routine incorporation of the treatment tabletop into patient treatment planning dose calculations. PACS numbers: 87.53.Dq, 87.53.Mr
format Online
Article
Text
id pubmed-5722426
institution National Center for Biotechnology Information
language English
publishDate 2006
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57224262018-04-02 Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop Myint, W. Kenji Niedbala, Malgorzata Wilkins, David Gerig, Lee H. J Appl Clin Med Phys Radiation Oncology Physics Carbon fiber is commonly used in radiation therapy for treatment tabletops and various immobilization and support devices, partially because it is generally perceived to be almost radiotransparent to high‐energy photons. To avoid exposure to normal tissue during modern radiation therapy, one must deliver the radiation from all gantry angles; hence, beams often transit the couch proximal to the patient. The effects of the beam attenuation by the support structure of the couch are often neglected in the planning process. In this study, we investigate the attenuation of 6‐MV and 18‐MV photon beams by a Medtec (Orange City, IA) carbon fiber couch. We have determined that neglecting the attenuation of oblique treatment fields by the carbon fiber couch can result in localized dose reduction from 4% to 16%, depending on energy, field size, and geometry. Further, we investigate the ability of a commercial treatment‐planning system (Theraplan Plus v3.8) to account for the attenuation by the treatment couch. Results show that incorporating the carbon fiber couch in the patient model reduces the dose error to less than 2%. The variation in dose reduction as a function of longitudinal couch position was also measured. In the triangular strut region of the couch, the attenuation varied [Formula: see text] following the periodic nature of the support structure. Based on these findings, we propose the routine incorporation of the treatment tabletop into patient treatment planning dose calculations. PACS numbers: 87.53.Dq, 87.53.Mr John Wiley and Sons Inc. 2006-08-24 /pmc/articles/PMC5722426/ /pubmed/17533341 http://dx.doi.org/10.1120/jacmp.v7i3.2247 Text en © 2006 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Myint, W. Kenji
Niedbala, Malgorzata
Wilkins, David
Gerig, Lee H.
Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop
title Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop
title_full Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop
title_fullStr Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop
title_full_unstemmed Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop
title_short Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop
title_sort investigating treatment dose error due to beam attenuation by a carbon fiber tabletop
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722426/
https://www.ncbi.nlm.nih.gov/pubmed/17533341
http://dx.doi.org/10.1120/jacmp.v7i3.2247
work_keys_str_mv AT myintwkenji investigatingtreatmentdoseerrorduetobeamattenuationbyacarbonfibertabletop
AT niedbalamalgorzata investigatingtreatmentdoseerrorduetobeamattenuationbyacarbonfibertabletop
AT wilkinsdavid investigatingtreatmentdoseerrorduetobeamattenuationbyacarbonfibertabletop
AT gerigleeh investigatingtreatmentdoseerrorduetobeamattenuationbyacarbonfibertabletop