Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation
A low‐energy (50 kV) X‐ray tube used for the stereotactic irradiation of intracranial lesions has been in use since 1999. The unit is calibrated prior to every procedure and during periodic quality assurance (QA) tests. The unit uses an internal and an external scintillation detector to monitor dose...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2006
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722427/ https://www.ncbi.nlm.nih.gov/pubmed/17533340 http://dx.doi.org/10.1120/jacmp.v7i3.2238 |
_version_ | 1783285011778109440 |
---|---|
author | Biggs, Peter J. |
author_facet | Biggs, Peter J. |
author_sort | Biggs, Peter J. |
collection | PubMed |
description | A low‐energy (50 kV) X‐ray tube used for the stereotactic irradiation of intracranial lesions has been in use since 1999. The unit is calibrated prior to every procedure and during periodic quality assurance (QA) tests. The unit uses an internal and an external scintillation detector to monitor dose as well as a conventional timer. The records of these calibrations were reviewed to see whether a change in the output had occurred over that period. Using time as the reference, it was found that both the internal radiation monitor (IRM) and the beam output, determined with a parallel plate ionization chamber, dropped by variable amounts over the given period. The beam output dropped significantly more than the IRM, while the external radiation monitor (ERM) showed no significant deviation from its initial value. The beam output dropped to about 90% of its initial value after about 200 days but remained relatively constant thereafter. The IRM dropped steadily to about 96% to 97% of its initial value at 1000 days, but recovered to about 98% after that. Calibration prior to each procedure is strongly recommended. PACS numbers: 87.53.Dq; 87.56.By |
format | Online Article Text |
id | pubmed-5722427 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2006 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57224272018-04-02 Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation Biggs, Peter J. J Appl Clin Med Phys Radiation Oncology Physics A low‐energy (50 kV) X‐ray tube used for the stereotactic irradiation of intracranial lesions has been in use since 1999. The unit is calibrated prior to every procedure and during periodic quality assurance (QA) tests. The unit uses an internal and an external scintillation detector to monitor dose as well as a conventional timer. The records of these calibrations were reviewed to see whether a change in the output had occurred over that period. Using time as the reference, it was found that both the internal radiation monitor (IRM) and the beam output, determined with a parallel plate ionization chamber, dropped by variable amounts over the given period. The beam output dropped significantly more than the IRM, while the external radiation monitor (ERM) showed no significant deviation from its initial value. The beam output dropped to about 90% of its initial value after about 200 days but remained relatively constant thereafter. The IRM dropped steadily to about 96% to 97% of its initial value at 1000 days, but recovered to about 98% after that. Calibration prior to each procedure is strongly recommended. PACS numbers: 87.53.Dq; 87.56.By John Wiley and Sons Inc. 2006-08-24 /pmc/articles/PMC5722427/ /pubmed/17533340 http://dx.doi.org/10.1120/jacmp.v7i3.2238 Text en © 2006 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Biggs, Peter J. Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation |
title | Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation |
title_full | Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation |
title_fullStr | Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation |
title_full_unstemmed | Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation |
title_short | Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation |
title_sort | long‐term stability of a 50‐kv x‐ray unit for stereotactic irradiation |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722427/ https://www.ncbi.nlm.nih.gov/pubmed/17533340 http://dx.doi.org/10.1120/jacmp.v7i3.2238 |
work_keys_str_mv | AT biggspeterj longtermstabilityofa50kvxrayunitforstereotacticirradiation |