Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation

A low‐energy (50 kV) X‐ray tube used for the stereotactic irradiation of intracranial lesions has been in use since 1999. The unit is calibrated prior to every procedure and during periodic quality assurance (QA) tests. The unit uses an internal and an external scintillation detector to monitor dose...

Descripción completa

Detalles Bibliográficos
Autor principal: Biggs, Peter J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2006
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722427/
https://www.ncbi.nlm.nih.gov/pubmed/17533340
http://dx.doi.org/10.1120/jacmp.v7i3.2238
_version_ 1783285011778109440
author Biggs, Peter J.
author_facet Biggs, Peter J.
author_sort Biggs, Peter J.
collection PubMed
description A low‐energy (50 kV) X‐ray tube used for the stereotactic irradiation of intracranial lesions has been in use since 1999. The unit is calibrated prior to every procedure and during periodic quality assurance (QA) tests. The unit uses an internal and an external scintillation detector to monitor dose as well as a conventional timer. The records of these calibrations were reviewed to see whether a change in the output had occurred over that period. Using time as the reference, it was found that both the internal radiation monitor (IRM) and the beam output, determined with a parallel plate ionization chamber, dropped by variable amounts over the given period. The beam output dropped significantly more than the IRM, while the external radiation monitor (ERM) showed no significant deviation from its initial value. The beam output dropped to about 90% of its initial value after about 200 days but remained relatively constant thereafter. The IRM dropped steadily to about 96% to 97% of its initial value at 1000 days, but recovered to about 98% after that. Calibration prior to each procedure is strongly recommended. PACS numbers: 87.53.Dq; 87.56.By
format Online
Article
Text
id pubmed-5722427
institution National Center for Biotechnology Information
language English
publishDate 2006
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57224272018-04-02 Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation Biggs, Peter J. J Appl Clin Med Phys Radiation Oncology Physics A low‐energy (50 kV) X‐ray tube used for the stereotactic irradiation of intracranial lesions has been in use since 1999. The unit is calibrated prior to every procedure and during periodic quality assurance (QA) tests. The unit uses an internal and an external scintillation detector to monitor dose as well as a conventional timer. The records of these calibrations were reviewed to see whether a change in the output had occurred over that period. Using time as the reference, it was found that both the internal radiation monitor (IRM) and the beam output, determined with a parallel plate ionization chamber, dropped by variable amounts over the given period. The beam output dropped significantly more than the IRM, while the external radiation monitor (ERM) showed no significant deviation from its initial value. The beam output dropped to about 90% of its initial value after about 200 days but remained relatively constant thereafter. The IRM dropped steadily to about 96% to 97% of its initial value at 1000 days, but recovered to about 98% after that. Calibration prior to each procedure is strongly recommended. PACS numbers: 87.53.Dq; 87.56.By John Wiley and Sons Inc. 2006-08-24 /pmc/articles/PMC5722427/ /pubmed/17533340 http://dx.doi.org/10.1120/jacmp.v7i3.2238 Text en © 2006 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Biggs, Peter J.
Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation
title Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation
title_full Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation
title_fullStr Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation
title_full_unstemmed Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation
title_short Long‐term stability of a 50‐kV X‐ray unit for stereotactic irradiation
title_sort long‐term stability of a 50‐kv x‐ray unit for stereotactic irradiation
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722427/
https://www.ncbi.nlm.nih.gov/pubmed/17533340
http://dx.doi.org/10.1120/jacmp.v7i3.2238
work_keys_str_mv AT biggspeterj longtermstabilityofa50kvxrayunitforstereotacticirradiation