Verifying monitor unit calculations for tangential breast fields
An essential component of quality assurance in radiation therapy is verifying the accuracy of monitor unit calculations. Differences between sophisticated algorithms using 2.5D or 3D calculations and simpler monitor unit check algorithms or hand calculations assuming a flat water phantom must be exp...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2006
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722436/ https://www.ncbi.nlm.nih.gov/pubmed/17533323 http://dx.doi.org/10.1120/jacmp.v7i2.2177 |
_version_ | 1783285014014722048 |
---|---|
author | Kay, Ian Dunscombe, Peter |
author_facet | Kay, Ian Dunscombe, Peter |
author_sort | Kay, Ian |
collection | PubMed |
description | An essential component of quality assurance in radiation therapy is verifying the accuracy of monitor unit calculations. Differences between sophisticated algorithms using 2.5D or 3D calculations and simpler monitor unit check algorithms or hand calculations assuming a flat water phantom must be expected. For many anatomical sites, such differences are small and of little or no consequence in the context of expected clinical impact. However, for tangential breast fields the discrepancies are considerably larger than those that would generally be considered acceptable. A simple model to reconcile the differences between sophisticated and simple algorithms is presented, based on replacing the breast contour with a triangular or elliptical contour and using this to estimate an equivalent rectangular prism providing equivalent scatter to the prescription point. The elliptical approximation reconciles the observed differences in calculated monitor units. The analysis we present can assist the treatment planning physicist in selecting a method and tolerance window for verifying monitor unit calculations for tangential breast fields. PACS: 87.53.Kn, 87.53.Tf, 87.53.Xd |
format | Online Article Text |
id | pubmed-5722436 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2006 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57224362018-04-02 Verifying monitor unit calculations for tangential breast fields Kay, Ian Dunscombe, Peter J Appl Clin Med Phys Radiation Oncology Physics An essential component of quality assurance in radiation therapy is verifying the accuracy of monitor unit calculations. Differences between sophisticated algorithms using 2.5D or 3D calculations and simpler monitor unit check algorithms or hand calculations assuming a flat water phantom must be expected. For many anatomical sites, such differences are small and of little or no consequence in the context of expected clinical impact. However, for tangential breast fields the discrepancies are considerably larger than those that would generally be considered acceptable. A simple model to reconcile the differences between sophisticated and simple algorithms is presented, based on replacing the breast contour with a triangular or elliptical contour and using this to estimate an equivalent rectangular prism providing equivalent scatter to the prescription point. The elliptical approximation reconciles the observed differences in calculated monitor units. The analysis we present can assist the treatment planning physicist in selecting a method and tolerance window for verifying monitor unit calculations for tangential breast fields. PACS: 87.53.Kn, 87.53.Tf, 87.53.Xd John Wiley and Sons Inc. 2006-05-25 /pmc/articles/PMC5722436/ /pubmed/17533323 http://dx.doi.org/10.1120/jacmp.v7i2.2177 Text en © 2006 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Radiation Oncology Physics Kay, Ian Dunscombe, Peter Verifying monitor unit calculations for tangential breast fields |
title | Verifying monitor unit calculations for tangential breast fields |
title_full | Verifying monitor unit calculations for tangential breast fields |
title_fullStr | Verifying monitor unit calculations for tangential breast fields |
title_full_unstemmed | Verifying monitor unit calculations for tangential breast fields |
title_short | Verifying monitor unit calculations for tangential breast fields |
title_sort | verifying monitor unit calculations for tangential breast fields |
topic | Radiation Oncology Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722436/ https://www.ncbi.nlm.nih.gov/pubmed/17533323 http://dx.doi.org/10.1120/jacmp.v7i2.2177 |
work_keys_str_mv | AT kayian verifyingmonitorunitcalculationsfortangentialbreastfields AT dunscombepeter verifyingmonitorunitcalculationsfortangentialbreastfields |