Verification procedure for isocentric alignment of proton beams

We present a technique—based on the Lutz, Winston, and Maleki test used in stereotactic linear accelerator radiosurgery—for verifying whether proton beams are being delivered within the required spatial coincidence with the gantry mechanical isocenter. Our procedure uses a proton beam that is collim...

Descripción completa

Detalles Bibliográficos
Autores principales: Ciangaru, George, Yang, James N., Oliver, Patrick J., Bues, Martin, Zhu, Mengping, Nakagawa, Fumio, Chiba, Hitoshi, Nakamura, Shin, Yoshino, Hirofumi, Umezawa, Mosumi, Smith, Alfred R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722618/
https://www.ncbi.nlm.nih.gov/pubmed/18449154
http://dx.doi.org/10.1120/jacmp.v8i4.2671
_version_ 1783285056330006528
author Ciangaru, George
Yang, James N.
Oliver, Patrick J.
Bues, Martin
Zhu, Mengping
Nakagawa, Fumio
Chiba, Hitoshi
Nakamura, Shin
Yoshino, Hirofumi
Umezawa, Mosumi
Smith, Alfred R.
author_facet Ciangaru, George
Yang, James N.
Oliver, Patrick J.
Bues, Martin
Zhu, Mengping
Nakagawa, Fumio
Chiba, Hitoshi
Nakamura, Shin
Yoshino, Hirofumi
Umezawa, Mosumi
Smith, Alfred R.
author_sort Ciangaru, George
collection PubMed
description We present a technique—based on the Lutz, Winston, and Maleki test used in stereotactic linear accelerator radiosurgery—for verifying whether proton beams are being delivered within the required spatial coincidence with the gantry mechanical isocenter. Our procedure uses a proton beam that is collimated by a circular aperture at its central axis and is then intercepted by a small steel sphere rigidly supported by the patient couch. A laser tracker measurement system and a correction algorithm for couch position assures precise positioning of the steel sphere at the mechanical isocenter of the gantry. A film‐based radiation dosimetry technique, chosen for the good spatial resolution it achieves, records the proton dose distribution for optical image analysis. The optical image obtained presents a circular high‐dose region surrounding a lower‐dose area corresponding to the proton beam absorption by the steel sphere, thereby providing a measure of the beam alignment with the mechanical isocenter. We found the self‐developing Gafchromic EBT film (International Specialty Products, Wayne, NJ) and commercial Epson 10000 XL flatbed scanner (Epson America, Long Beach, CA) to be accurate and efficient tools. The positions of the gantry mechanical and proton beam isocenters, as recorded on film, were clearly identifiable within the scanning resolution used for routine alignment testing (0.17 mm per pixel). The mean displacement of the collimated proton beam from the gantry mechanical isocenter was [Formula: see text] for the gantry positions tested, which was well within the maximum deviation of 0.50 mm accepted at the Proton Therapy Center in Houston. PACS numbers: 87.53.Xd, 87.53.Oc, 87.56.‐v, 87.66.‐a, 87.56.Fc
format Online
Article
Text
id pubmed-5722618
institution National Center for Biotechnology Information
language English
publishDate 2007
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57226182018-04-02 Verification procedure for isocentric alignment of proton beams Ciangaru, George Yang, James N. Oliver, Patrick J. Bues, Martin Zhu, Mengping Nakagawa, Fumio Chiba, Hitoshi Nakamura, Shin Yoshino, Hirofumi Umezawa, Mosumi Smith, Alfred R. J Appl Clin Med Phys Radiation Oncology Physics We present a technique—based on the Lutz, Winston, and Maleki test used in stereotactic linear accelerator radiosurgery—for verifying whether proton beams are being delivered within the required spatial coincidence with the gantry mechanical isocenter. Our procedure uses a proton beam that is collimated by a circular aperture at its central axis and is then intercepted by a small steel sphere rigidly supported by the patient couch. A laser tracker measurement system and a correction algorithm for couch position assures precise positioning of the steel sphere at the mechanical isocenter of the gantry. A film‐based radiation dosimetry technique, chosen for the good spatial resolution it achieves, records the proton dose distribution for optical image analysis. The optical image obtained presents a circular high‐dose region surrounding a lower‐dose area corresponding to the proton beam absorption by the steel sphere, thereby providing a measure of the beam alignment with the mechanical isocenter. We found the self‐developing Gafchromic EBT film (International Specialty Products, Wayne, NJ) and commercial Epson 10000 XL flatbed scanner (Epson America, Long Beach, CA) to be accurate and efficient tools. The positions of the gantry mechanical and proton beam isocenters, as recorded on film, were clearly identifiable within the scanning resolution used for routine alignment testing (0.17 mm per pixel). The mean displacement of the collimated proton beam from the gantry mechanical isocenter was [Formula: see text] for the gantry positions tested, which was well within the maximum deviation of 0.50 mm accepted at the Proton Therapy Center in Houston. PACS numbers: 87.53.Xd, 87.53.Oc, 87.56.‐v, 87.66.‐a, 87.56.Fc John Wiley and Sons Inc. 2007-10-24 /pmc/articles/PMC5722618/ /pubmed/18449154 http://dx.doi.org/10.1120/jacmp.v8i4.2671 Text en © 2007 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Ciangaru, George
Yang, James N.
Oliver, Patrick J.
Bues, Martin
Zhu, Mengping
Nakagawa, Fumio
Chiba, Hitoshi
Nakamura, Shin
Yoshino, Hirofumi
Umezawa, Mosumi
Smith, Alfred R.
Verification procedure for isocentric alignment of proton beams
title Verification procedure for isocentric alignment of proton beams
title_full Verification procedure for isocentric alignment of proton beams
title_fullStr Verification procedure for isocentric alignment of proton beams
title_full_unstemmed Verification procedure for isocentric alignment of proton beams
title_short Verification procedure for isocentric alignment of proton beams
title_sort verification procedure for isocentric alignment of proton beams
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722618/
https://www.ncbi.nlm.nih.gov/pubmed/18449154
http://dx.doi.org/10.1120/jacmp.v8i4.2671
work_keys_str_mv AT ciangarugeorge verificationprocedureforisocentricalignmentofprotonbeams
AT yangjamesn verificationprocedureforisocentricalignmentofprotonbeams
AT oliverpatrickj verificationprocedureforisocentricalignmentofprotonbeams
AT buesmartin verificationprocedureforisocentricalignmentofprotonbeams
AT zhumengping verificationprocedureforisocentricalignmentofprotonbeams
AT nakagawafumio verificationprocedureforisocentricalignmentofprotonbeams
AT chibahitoshi verificationprocedureforisocentricalignmentofprotonbeams
AT nakamurashin verificationprocedureforisocentricalignmentofprotonbeams
AT yoshinohirofumi verificationprocedureforisocentricalignmentofprotonbeams
AT umezawamosumi verificationprocedureforisocentricalignmentofprotonbeams
AT smithalfredr verificationprocedureforisocentricalignmentofprotonbeams