Cargando…
Verification procedure for isocentric alignment of proton beams
We present a technique—based on the Lutz, Winston, and Maleki test used in stereotactic linear accelerator radiosurgery—for verifying whether proton beams are being delivered within the required spatial coincidence with the gantry mechanical isocenter. Our procedure uses a proton beam that is collim...
Autores principales: | Ciangaru, George, Yang, James N., Oliver, Patrick J., Bues, Martin, Zhu, Mengping, Nakagawa, Fumio, Chiba, Hitoshi, Nakamura, Shin, Yoshino, Hirofumi, Umezawa, Mosumi, Smith, Alfred R. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2007
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5722618/ https://www.ncbi.nlm.nih.gov/pubmed/18449154 http://dx.doi.org/10.1120/jacmp.v8i4.2671 |
Ejemplares similares
-
Validating robotic couch isocentricity with 3D surface imaging
por: El‐Sherif, Omar, et al.
Publicado: (2020) -
Comparing the monoisocentric and dual isocentric techniques in chest wall radiotherapy of mastectomy patients
por: Banaei, Amin, et al.
Publicado: (2015) -
Implementing corrections of isocentric shifts for the stereotactic irradiation of cerebral targets: Clinical validation
por: Wegener, Sonja, et al.
Publicado: (2022) -
Accuracy of low‐dose proton CT image registration for pretreatment alignment verification in reference to planning proton CT
por: Cassetta, Roberto, et al.
Publicado: (2019) -
Conceptual and ion-optical designs of an isocentric gantry for light-ion cancer therapy
por: Vorobev, L G, et al.
Publicado: (1998)