The accuracy of dynamic wedge dose computation in the ADAC Pinnacle RTP system

The nonphysical wedge is a modality that uses computer‐controlled jaw motion to generate wedge‐shaped dose distributions. There are Varian enhanced dynamic wedges (EDWs) and Siemens virtual wedges (VWs). We recently commissioned dynamic wedges on both Varian and Siemens LINACs. The beam data, acquir...

Descripción completa

Detalles Bibliográficos
Autores principales: Shao, H., Wu, X., Luo, C., Crooks, S., Bernstein, A., Markoe, A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2004
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5723519/
https://www.ncbi.nlm.nih.gov/pubmed/15738920
http://dx.doi.org/10.1120/jacmp.v5i4.1964
Descripción
Sumario:The nonphysical wedge is a modality that uses computer‐controlled jaw motion to generate wedge‐shaped dose distributions. There are Varian enhanced dynamic wedges (EDWs) and Siemens virtual wedges (VWs). We recently commissioned dynamic wedges on both Varian and Siemens LINACs. The beam data, acquired with a Wellhöfer chamber array and a Sun Nuclear profiler, are used for modeling in the ADAC Pinnacle system. As recommended by ADAC, only a limited number of beam data is measured and used for beam modeling. Therefore, the dose distributions of dynamic wedges generated by Pinnacle must be examined. Following the commissioning of the dynamic wedges, we used Pinnacle to generate a number of dose distributions with different energies, wedge angles, field sizes, and depths. The computed data from Pinnacle are then compared with the measured data. The deviations of the output factor in all square and rectangular fields are mostly within 2.0% for both EDW and VW. For asymmetric fields, the deviations are within 3%. However, exceptions of differences more than 3% have been found in a larger field and large wedge combinations. The precision of the beam profiles generated by Pinnacle is also evaluated. As a result of this investigation, we present a scope of quality assurance tests that are necessary to ensure acceptable consistency between the delivered dose and the associated treatment plan when dynamic wedges are applied. PACS numbers: 8753 Dq, 87.53.Xd