Shielding evaluation and acceptance testing of a prefabricated, modular, temporary radiation therapy treatment facility

We have recently commissioned a temporary radiation therapy facility that is novel in two aspects: it was constructed using modular components, and the LINAC was installed in one of the modular sections before it was lifted into position. Additional steel and granular fill was added to the modular s...

Descripción completa

Detalles Bibliográficos
Autor principal: Ezzell, Gary A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2004
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5723524/
https://www.ncbi.nlm.nih.gov/pubmed/15738926
http://dx.doi.org/10.1120/jacmp.v5i4.2025
Descripción
Sumario:We have recently commissioned a temporary radiation therapy facility that is novel in two aspects: it was constructed using modular components, and the LINAC was installed in one of the modular sections before it was lifted into position. Additional steel and granular fill was added to the modular sections on‐site during construction. The building will be disassembled and removed when no longer needed. This paper describes the radiation shielding specifications and survey of the facility, as well as the ramifications for acceptance testing occasioned by the novel installation procedure. The LINAC is a Varian 21EX operating at 6 MV and 18 MV. The radiation levels outside the vault satisfied the design criteria, and no anomalous leakage was detected along the joints of the modular structure. At 18 MV and 600 monitor units (MU) per minute, the radiation level outside the primary barrier walls was [Formula: see text] of photons; there were no detectable neutrons. Outside the direct‐shielded door, the levels were [Formula: see text] of photons and [Formula: see text] of neutrons. The isocentricity of the accelerator met the acceptance criteria and was not affected by its preinstallation into an integrated baseframe and subsequent transport to the building site. PACS numbers: 87.52.Df, 87.52.Ga