Shielding evaluation and acceptance testing of a prefabricated, modular, temporary radiation therapy treatment facility
We have recently commissioned a temporary radiation therapy facility that is novel in two aspects: it was constructed using modular components, and the LINAC was installed in one of the modular sections before it was lifted into position. Additional steel and granular fill was added to the modular s...