Accuracy and reliability of the sensory test performed using the laryngopharyngeal endoscopic esthesiometer and rangefinder in patients with suspected obstructive sleep apnoea hypopnoea: protocol for a prospective double-blinded, randomised, exploratory study

INTRODUCTION: Patients with obstructive sleep apnoea hypopnoea syndrome (OSA) might have varying degrees of laryngopharyngeal mechanical hyposensitivity that might impair the brain’s capacity to prevent airway collapse during sleep. However, this knowledge about sensory compromises in OSA comes from...

Descripción completa

Detalles Bibliográficos
Autores principales: Giraldo-Cadavid, Luis Fernando, Bastidas, Alirio Rodrigo, Padilla-Ortiz, Diana Marcela, Concha-Galan, Diana Carolina, Bazurto, María Angelica, Vargas, Leslie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BMJ Publishing Group 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724088/
https://www.ncbi.nlm.nih.gov/pubmed/28827237
http://dx.doi.org/10.1136/bmjopen-2016-015235
_version_ 1783285294716420096
author Giraldo-Cadavid, Luis Fernando
Bastidas, Alirio Rodrigo
Padilla-Ortiz, Diana Marcela
Concha-Galan, Diana Carolina
Bazurto, María Angelica
Vargas, Leslie
author_facet Giraldo-Cadavid, Luis Fernando
Bastidas, Alirio Rodrigo
Padilla-Ortiz, Diana Marcela
Concha-Galan, Diana Carolina
Bazurto, María Angelica
Vargas, Leslie
author_sort Giraldo-Cadavid, Luis Fernando
collection PubMed
description INTRODUCTION: Patients with obstructive sleep apnoea hypopnoea syndrome (OSA) might have varying degrees of laryngopharyngeal mechanical hyposensitivity that might impair the brain’s capacity to prevent airway collapse during sleep. However, this knowledge about sensory compromises in OSA comes from studies performed using methods with little evidence of their validity. Hence, the purpose of this study is to assess the reliability and accuracy of the measurement of laryngopharyngeal mechanosensitivity in patients with OSA using a recently developed laryngopharyngeal endoscopic esthesiometer and rangefinder (LPEER). METHODS AND ANALYSIS: The study will be prospective and double blinded, with a randomised crossover assignment of raters performing the sensory tests. Subjects will be recruited from patients with suspected OSA referred for baseline polysomnography to a university hospital sleep laboratory. Intra-rater and inter-rater reliability will be evaluated using the Bland–Altman’s limits of agreement plot, the intraclass correlation coefficient, and the Pearson or Spearman correlation coefficient, depending on the distribution of the variables. Diagnostic accuracy will be evaluated plotting ROC curves using standard baseline polysomnography as a reference. The sensory threshold values ​​for patients with mild, moderate and severe OSA will be determined and compared using ANOVA or the Kruskal–Wallis test, depending on the distribution of the variables. The LPEER could be a new tool for evaluating and monitoring laryngopharyngeal sensory impairment in patients with OSA. If it is shown to be valid, it could help to increase our understanding of the pathophysiological mechanisms of this condition and potentially help in finding new therapeutic interventions for OSA. ETHICS AND DISSEMINATION: The protocol has been approved by the Institutional Review Board of Fundacion Neumologica Colombiana. The results will be disseminated through conference presentations and peer-reviewed publication. TRIAL REGISTRATION: This trial was registered at Clinical Trials Accuracy of the sensory test using the lLaryngopharyngeal endoscopic esthesiometer in obstructive sleep apnea. Protocol ID: 201611-22405. ClinicalTrials.gov ID: NCT03109171.
format Online
Article
Text
id pubmed-5724088
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher BMJ Publishing Group
record_format MEDLINE/PubMed
spelling pubmed-57240882017-12-19 Accuracy and reliability of the sensory test performed using the laryngopharyngeal endoscopic esthesiometer and rangefinder in patients with suspected obstructive sleep apnoea hypopnoea: protocol for a prospective double-blinded, randomised, exploratory study Giraldo-Cadavid, Luis Fernando Bastidas, Alirio Rodrigo Padilla-Ortiz, Diana Marcela Concha-Galan, Diana Carolina Bazurto, María Angelica Vargas, Leslie BMJ Open Respiratory Medicine INTRODUCTION: Patients with obstructive sleep apnoea hypopnoea syndrome (OSA) might have varying degrees of laryngopharyngeal mechanical hyposensitivity that might impair the brain’s capacity to prevent airway collapse during sleep. However, this knowledge about sensory compromises in OSA comes from studies performed using methods with little evidence of their validity. Hence, the purpose of this study is to assess the reliability and accuracy of the measurement of laryngopharyngeal mechanosensitivity in patients with OSA using a recently developed laryngopharyngeal endoscopic esthesiometer and rangefinder (LPEER). METHODS AND ANALYSIS: The study will be prospective and double blinded, with a randomised crossover assignment of raters performing the sensory tests. Subjects will be recruited from patients with suspected OSA referred for baseline polysomnography to a university hospital sleep laboratory. Intra-rater and inter-rater reliability will be evaluated using the Bland–Altman’s limits of agreement plot, the intraclass correlation coefficient, and the Pearson or Spearman correlation coefficient, depending on the distribution of the variables. Diagnostic accuracy will be evaluated plotting ROC curves using standard baseline polysomnography as a reference. The sensory threshold values ​​for patients with mild, moderate and severe OSA will be determined and compared using ANOVA or the Kruskal–Wallis test, depending on the distribution of the variables. The LPEER could be a new tool for evaluating and monitoring laryngopharyngeal sensory impairment in patients with OSA. If it is shown to be valid, it could help to increase our understanding of the pathophysiological mechanisms of this condition and potentially help in finding new therapeutic interventions for OSA. ETHICS AND DISSEMINATION: The protocol has been approved by the Institutional Review Board of Fundacion Neumologica Colombiana. The results will be disseminated through conference presentations and peer-reviewed publication. TRIAL REGISTRATION: This trial was registered at Clinical Trials Accuracy of the sensory test using the lLaryngopharyngeal endoscopic esthesiometer in obstructive sleep apnea. Protocol ID: 201611-22405. ClinicalTrials.gov ID: NCT03109171. BMJ Publishing Group 2017-08-21 /pmc/articles/PMC5724088/ /pubmed/28827237 http://dx.doi.org/10.1136/bmjopen-2016-015235 Text en © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2017. All rights reserved. No commercial use is permitted unless otherwise expressly granted. This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
spellingShingle Respiratory Medicine
Giraldo-Cadavid, Luis Fernando
Bastidas, Alirio Rodrigo
Padilla-Ortiz, Diana Marcela
Concha-Galan, Diana Carolina
Bazurto, María Angelica
Vargas, Leslie
Accuracy and reliability of the sensory test performed using the laryngopharyngeal endoscopic esthesiometer and rangefinder in patients with suspected obstructive sleep apnoea hypopnoea: protocol for a prospective double-blinded, randomised, exploratory study
title Accuracy and reliability of the sensory test performed using the laryngopharyngeal endoscopic esthesiometer and rangefinder in patients with suspected obstructive sleep apnoea hypopnoea: protocol for a prospective double-blinded, randomised, exploratory study
title_full Accuracy and reliability of the sensory test performed using the laryngopharyngeal endoscopic esthesiometer and rangefinder in patients with suspected obstructive sleep apnoea hypopnoea: protocol for a prospective double-blinded, randomised, exploratory study
title_fullStr Accuracy and reliability of the sensory test performed using the laryngopharyngeal endoscopic esthesiometer and rangefinder in patients with suspected obstructive sleep apnoea hypopnoea: protocol for a prospective double-blinded, randomised, exploratory study
title_full_unstemmed Accuracy and reliability of the sensory test performed using the laryngopharyngeal endoscopic esthesiometer and rangefinder in patients with suspected obstructive sleep apnoea hypopnoea: protocol for a prospective double-blinded, randomised, exploratory study
title_short Accuracy and reliability of the sensory test performed using the laryngopharyngeal endoscopic esthesiometer and rangefinder in patients with suspected obstructive sleep apnoea hypopnoea: protocol for a prospective double-blinded, randomised, exploratory study
title_sort accuracy and reliability of the sensory test performed using the laryngopharyngeal endoscopic esthesiometer and rangefinder in patients with suspected obstructive sleep apnoea hypopnoea: protocol for a prospective double-blinded, randomised, exploratory study
topic Respiratory Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724088/
https://www.ncbi.nlm.nih.gov/pubmed/28827237
http://dx.doi.org/10.1136/bmjopen-2016-015235
work_keys_str_mv AT giraldocadavidluisfernando accuracyandreliabilityofthesensorytestperformedusingthelaryngopharyngealendoscopicesthesiometerandrangefinderinpatientswithsuspectedobstructivesleepapnoeahypopnoeaprotocolforaprospectivedoubleblindedrandomisedexploratorystudy
AT bastidasaliriorodrigo accuracyandreliabilityofthesensorytestperformedusingthelaryngopharyngealendoscopicesthesiometerandrangefinderinpatientswithsuspectedobstructivesleepapnoeahypopnoeaprotocolforaprospectivedoubleblindedrandomisedexploratorystudy
AT padillaortizdianamarcela accuracyandreliabilityofthesensorytestperformedusingthelaryngopharyngealendoscopicesthesiometerandrangefinderinpatientswithsuspectedobstructivesleepapnoeahypopnoeaprotocolforaprospectivedoubleblindedrandomisedexploratorystudy
AT conchagalandianacarolina accuracyandreliabilityofthesensorytestperformedusingthelaryngopharyngealendoscopicesthesiometerandrangefinderinpatientswithsuspectedobstructivesleepapnoeahypopnoeaprotocolforaprospectivedoubleblindedrandomisedexploratorystudy
AT bazurtomariaangelica accuracyandreliabilityofthesensorytestperformedusingthelaryngopharyngealendoscopicesthesiometerandrangefinderinpatientswithsuspectedobstructivesleepapnoeahypopnoeaprotocolforaprospectivedoubleblindedrandomisedexploratorystudy
AT vargasleslie accuracyandreliabilityofthesensorytestperformedusingthelaryngopharyngealendoscopicesthesiometerandrangefinderinpatientswithsuspectedobstructivesleepapnoeahypopnoeaprotocolforaprospectivedoubleblindedrandomisedexploratorystudy