Cargando…

Corrections to traditional methods of verifying tangential‐breast 3D monitor‐unit calculations: Use of an equivalent triangle to estimate effective fields

This paper describes an innovative method for correctly estimating the effective field size of tangential‐breast fields. The method uses an “equivalent triangle” to verify intact breast tangential field monitor‐unit settings calculated by a 3D planning system to within 2%. The effects on verificatio...

Descripción completa

Detalles Bibliográficos
Autores principales: Prado, Karl L., Kirsner, Steven M., Erice, Rolly C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724436/
https://www.ncbi.nlm.nih.gov/pubmed/12540818
http://dx.doi.org/10.1120/jacmp.v4i1.2541
Descripción
Sumario:This paper describes an innovative method for correctly estimating the effective field size of tangential‐breast fields. The method uses an “equivalent triangle” to verify intact breast tangential field monitor‐unit settings calculated by a 3D planning system to within 2%. The effects on verification calculations of loss of full scatter due to beam oblique incidence, proximity to field boundaries, and reduced scattering volumes are handled properly. The methodology is validated by comparing calculations performed by the 3D planning system with the respective verification estimates. The accuracy of this technique is established for dose calculations both with and without heterogeneity corrections. © 2003 American College of Medical Physics. PACS number(s): 87.53.– j