Corrections to traditional methods of verifying tangential‐breast 3D monitor‐unit calculations: Use of an equivalent triangle to estimate effective fields

This paper describes an innovative method for correctly estimating the effective field size of tangential‐breast fields. The method uses an “equivalent triangle” to verify intact breast tangential field monitor‐unit settings calculated by a 3D planning system to within 2%. The effects on verificatio...

Descripción completa

Detalles Bibliográficos
Autores principales: Prado, Karl L., Kirsner, Steven M., Erice, Rolly C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724436/
https://www.ncbi.nlm.nih.gov/pubmed/12540818
http://dx.doi.org/10.1120/jacmp.v4i1.2541
_version_ 1783285358562115584
author Prado, Karl L.
Kirsner, Steven M.
Erice, Rolly C.
author_facet Prado, Karl L.
Kirsner, Steven M.
Erice, Rolly C.
author_sort Prado, Karl L.
collection PubMed
description This paper describes an innovative method for correctly estimating the effective field size of tangential‐breast fields. The method uses an “equivalent triangle” to verify intact breast tangential field monitor‐unit settings calculated by a 3D planning system to within 2%. The effects on verification calculations of loss of full scatter due to beam oblique incidence, proximity to field boundaries, and reduced scattering volumes are handled properly. The methodology is validated by comparing calculations performed by the 3D planning system with the respective verification estimates. The accuracy of this technique is established for dose calculations both with and without heterogeneity corrections. © 2003 American College of Medical Physics. PACS number(s): 87.53.– j
format Online
Article
Text
id pubmed-5724436
institution National Center for Biotechnology Information
language English
publishDate 2003
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57244362018-04-02 Corrections to traditional methods of verifying tangential‐breast 3D monitor‐unit calculations: Use of an equivalent triangle to estimate effective fields Prado, Karl L. Kirsner, Steven M. Erice, Rolly C. J Appl Clin Med Phys Radiation Oncology Physics This paper describes an innovative method for correctly estimating the effective field size of tangential‐breast fields. The method uses an “equivalent triangle” to verify intact breast tangential field monitor‐unit settings calculated by a 3D planning system to within 2%. The effects on verification calculations of loss of full scatter due to beam oblique incidence, proximity to field boundaries, and reduced scattering volumes are handled properly. The methodology is validated by comparing calculations performed by the 3D planning system with the respective verification estimates. The accuracy of this technique is established for dose calculations both with and without heterogeneity corrections. © 2003 American College of Medical Physics. PACS number(s): 87.53.– j John Wiley and Sons Inc. 2003-01-01 /pmc/articles/PMC5724436/ /pubmed/12540818 http://dx.doi.org/10.1120/jacmp.v4i1.2541 Text en © 2003 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Prado, Karl L.
Kirsner, Steven M.
Erice, Rolly C.
Corrections to traditional methods of verifying tangential‐breast 3D monitor‐unit calculations: Use of an equivalent triangle to estimate effective fields
title Corrections to traditional methods of verifying tangential‐breast 3D monitor‐unit calculations: Use of an equivalent triangle to estimate effective fields
title_full Corrections to traditional methods of verifying tangential‐breast 3D monitor‐unit calculations: Use of an equivalent triangle to estimate effective fields
title_fullStr Corrections to traditional methods of verifying tangential‐breast 3D monitor‐unit calculations: Use of an equivalent triangle to estimate effective fields
title_full_unstemmed Corrections to traditional methods of verifying tangential‐breast 3D monitor‐unit calculations: Use of an equivalent triangle to estimate effective fields
title_short Corrections to traditional methods of verifying tangential‐breast 3D monitor‐unit calculations: Use of an equivalent triangle to estimate effective fields
title_sort corrections to traditional methods of verifying tangential‐breast 3d monitor‐unit calculations: use of an equivalent triangle to estimate effective fields
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724436/
https://www.ncbi.nlm.nih.gov/pubmed/12540818
http://dx.doi.org/10.1120/jacmp.v4i1.2541
work_keys_str_mv AT pradokarll correctionstotraditionalmethodsofverifyingtangentialbreast3dmonitorunitcalculationsuseofanequivalenttriangletoestimateeffectivefields
AT kirsnerstevenm correctionstotraditionalmethodsofverifyingtangentialbreast3dmonitorunitcalculationsuseofanequivalenttriangletoestimateeffectivefields
AT ericerollyc correctionstotraditionalmethodsofverifyingtangentialbreast3dmonitorunitcalculationsuseofanequivalenttriangletoestimateeffectivefields