Cargando…

Shielding evaluation for IMRT implementation in an existing accelerator vault

A formalism is developed for evaluating the shielding in an existing vault to be used for IMRT. Existing exposure rate measurements are utilized as well as a newly developed effective modulation scaling factor. Examples are given for vaults housing 6, 10 and 18 MV linear accelerators. The use of an...

Descripción completa

Detalles Bibliográficos
Autores principales: Price, R. A., Chibani, O., Ma, C.‐M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724448/
https://www.ncbi.nlm.nih.gov/pubmed/12841794
http://dx.doi.org/10.1120/jacmp.v4i3.2520
_version_ 1783285361463525376
author Price, R. A.
Chibani, O.
Ma, C.‐M.
author_facet Price, R. A.
Chibani, O.
Ma, C.‐M.
author_sort Price, R. A.
collection PubMed
description A formalism is developed for evaluating the shielding in an existing vault to be used for IMRT. Existing exposure rate measurements are utilized as well as a newly developed effective modulation scaling factor. Examples are given for vaults housing 6, 10 and 18 MV linear accelerators. The use of an 18 MV Siemens linear accelerator is evaluated for IMRT delivery with respect to neutron production and the effects on individual patients. A modified modulation scaling factor is developed and the risk of the incurrence of fatal secondary malignancies is estimated. The difference in neutron production between 18 MV Varian and Siemens accelerators is estimated using Monte Carlo results. The neutron production from the Siemens accelerator is found to be approximately 4 times less than that of the Varian accelerator resulting in a risk of fatal secondary malignancy occurrence of approximately 1.6% when using the SMLC delivery technique and our measured modulation scaling factors. This compares with a previously published value of 1.6% for routine 3D CRT delivery on the Varian accelerator. PACS number(s): 87.52.Ga, 87.52.Px, 87.53.Qc, 87.53.Wz
format Online
Article
Text
id pubmed-5724448
institution National Center for Biotechnology Information
language English
publishDate 2003
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57244482018-04-02 Shielding evaluation for IMRT implementation in an existing accelerator vault Price, R. A. Chibani, O. Ma, C.‐M. J Appl Clin Med Phys Radiation Protection & Regulations A formalism is developed for evaluating the shielding in an existing vault to be used for IMRT. Existing exposure rate measurements are utilized as well as a newly developed effective modulation scaling factor. Examples are given for vaults housing 6, 10 and 18 MV linear accelerators. The use of an 18 MV Siemens linear accelerator is evaluated for IMRT delivery with respect to neutron production and the effects on individual patients. A modified modulation scaling factor is developed and the risk of the incurrence of fatal secondary malignancies is estimated. The difference in neutron production between 18 MV Varian and Siemens accelerators is estimated using Monte Carlo results. The neutron production from the Siemens accelerator is found to be approximately 4 times less than that of the Varian accelerator resulting in a risk of fatal secondary malignancy occurrence of approximately 1.6% when using the SMLC delivery technique and our measured modulation scaling factors. This compares with a previously published value of 1.6% for routine 3D CRT delivery on the Varian accelerator. PACS number(s): 87.52.Ga, 87.52.Px, 87.53.Qc, 87.53.Wz John Wiley and Sons Inc. 2003-06-01 /pmc/articles/PMC5724448/ /pubmed/12841794 http://dx.doi.org/10.1120/jacmp.v4i3.2520 Text en © 2003 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Protection & Regulations
Price, R. A.
Chibani, O.
Ma, C.‐M.
Shielding evaluation for IMRT implementation in an existing accelerator vault
title Shielding evaluation for IMRT implementation in an existing accelerator vault
title_full Shielding evaluation for IMRT implementation in an existing accelerator vault
title_fullStr Shielding evaluation for IMRT implementation in an existing accelerator vault
title_full_unstemmed Shielding evaluation for IMRT implementation in an existing accelerator vault
title_short Shielding evaluation for IMRT implementation in an existing accelerator vault
title_sort shielding evaluation for imrt implementation in an existing accelerator vault
topic Radiation Protection & Regulations
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724448/
https://www.ncbi.nlm.nih.gov/pubmed/12841794
http://dx.doi.org/10.1120/jacmp.v4i3.2520
work_keys_str_mv AT pricera shieldingevaluationforimrtimplementationinanexistingacceleratorvault
AT chibanio shieldingevaluationforimrtimplementationinanexistingacceleratorvault
AT macm shieldingevaluationforimrtimplementationinanexistingacceleratorvault