Technical note: The effect of the 4‐mm‐collimator output factor on gamma knife dose distributions
We present results of investigations of the clinical significance of variations in the value of the 4‐mm‐collimator output factor, [Formula: see text]. Changes in treatment volume, dose‐volume histograms (DVHs), and isodose distributions were studied, by varying [Formula: see text] up to 20%. The va...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2003
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724451/ https://www.ncbi.nlm.nih.gov/pubmed/14604431 http://dx.doi.org/10.1120/jacmp.v4i4.2512 |
_version_ | 1783285362198577152 |
---|---|
author | Gibbons, John P. Mihailidis, Dimitris Worthington, Curtis Alkhatib, Hassaan Boulware, Raleigh Clark, Robert Dial, Burke Neglia, William |
author_facet | Gibbons, John P. Mihailidis, Dimitris Worthington, Curtis Alkhatib, Hassaan Boulware, Raleigh Clark, Robert Dial, Burke Neglia, William |
author_sort | Gibbons, John P. |
collection | PubMed |
description | We present results of investigations of the clinical significance of variations in the value of the 4‐mm‐collimator output factor, [Formula: see text]. Changes in treatment volume, dose‐volume histograms (DVHs), and isodose distributions were studied, by varying [Formula: see text] up to 20%. The variations were performed on a sample of clinical patient treatment plans for which the 4 mm collimator was used. Although smaller effects are noted for the prescription isodose line, greater dosimetric changes occur for higher dose regions within the target. PACS number(s): 87.53.–j, 87.52.–g |
format | Online Article Text |
id | pubmed-5724451 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2003 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57244512018-04-02 Technical note: The effect of the 4‐mm‐collimator output factor on gamma knife dose distributions Gibbons, John P. Mihailidis, Dimitris Worthington, Curtis Alkhatib, Hassaan Boulware, Raleigh Clark, Robert Dial, Burke Neglia, William J Appl Clin Med Phys Technical Notes We present results of investigations of the clinical significance of variations in the value of the 4‐mm‐collimator output factor, [Formula: see text]. Changes in treatment volume, dose‐volume histograms (DVHs), and isodose distributions were studied, by varying [Formula: see text] up to 20%. The variations were performed on a sample of clinical patient treatment plans for which the 4 mm collimator was used. Although smaller effects are noted for the prescription isodose line, greater dosimetric changes occur for higher dose regions within the target. PACS number(s): 87.53.–j, 87.52.–g John Wiley and Sons Inc. 2003-09-01 /pmc/articles/PMC5724451/ /pubmed/14604431 http://dx.doi.org/10.1120/jacmp.v4i4.2512 Text en © 2003 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Notes Gibbons, John P. Mihailidis, Dimitris Worthington, Curtis Alkhatib, Hassaan Boulware, Raleigh Clark, Robert Dial, Burke Neglia, William Technical note: The effect of the 4‐mm‐collimator output factor on gamma knife dose distributions |
title | Technical note: The effect of the 4‐mm‐collimator output factor on gamma knife dose distributions |
title_full | Technical note: The effect of the 4‐mm‐collimator output factor on gamma knife dose distributions |
title_fullStr | Technical note: The effect of the 4‐mm‐collimator output factor on gamma knife dose distributions |
title_full_unstemmed | Technical note: The effect of the 4‐mm‐collimator output factor on gamma knife dose distributions |
title_short | Technical note: The effect of the 4‐mm‐collimator output factor on gamma knife dose distributions |
title_sort | technical note: the effect of the 4‐mm‐collimator output factor on gamma knife dose distributions |
topic | Technical Notes |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724451/ https://www.ncbi.nlm.nih.gov/pubmed/14604431 http://dx.doi.org/10.1120/jacmp.v4i4.2512 |
work_keys_str_mv | AT gibbonsjohnp technicalnotetheeffectofthe4mmcollimatoroutputfactorongammaknifedosedistributions AT mihailidisdimitris technicalnotetheeffectofthe4mmcollimatoroutputfactorongammaknifedosedistributions AT worthingtoncurtis technicalnotetheeffectofthe4mmcollimatoroutputfactorongammaknifedosedistributions AT alkhatibhassaan technicalnotetheeffectofthe4mmcollimatoroutputfactorongammaknifedosedistributions AT boulwareraleigh technicalnotetheeffectofthe4mmcollimatoroutputfactorongammaknifedosedistributions AT clarkrobert technicalnotetheeffectofthe4mmcollimatoroutputfactorongammaknifedosedistributions AT dialburke technicalnotetheeffectofthe4mmcollimatoroutputfactorongammaknifedosedistributions AT negliawilliam technicalnotetheeffectofthe4mmcollimatoroutputfactorongammaknifedosedistributions |