Dose calculation accuracy of lung planning with a commercial IMRT treatment planning system
The dose calculation accuracy of a commercial pencil beam IMRT planning system is evaluated by comparison with Monte Carlo calculations and measurements in an anthropomorphic phantom. The target volume is in the right lung and mediastinum and thus significant tissue inhomogeneities are present. The...
Autores principales: | McDermott, Patrick N., He, Tongming, DeYoung, A. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2003
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724454/ https://www.ncbi.nlm.nih.gov/pubmed/14604424 http://dx.doi.org/10.1120/jacmp.v4i4.2505 |