Cargando…

Reference photon dosimetry data: A preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator

The dosimetric characteristics for modern computer‐controlled linear accelerators with the same make, model, and nominal energy are known to be very similar, as long as the machines are unaltered from the manufacturer's original specifications. In this preliminary study, a quantitative investig...

Descripción completa

Detalles Bibliográficos
Autores principales: Cho, S. H., Ibbott, G. S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2003
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724467/
https://www.ncbi.nlm.nih.gov/pubmed/14604419
http://dx.doi.org/10.1120/jacmp.v4i4.2500
_version_ 1783285366049996800
author Cho, S. H.
Ibbott, G. S.
author_facet Cho, S. H.
Ibbott, G. S.
author_sort Cho, S. H.
collection PubMed
description The dosimetric characteristics for modern computer‐controlled linear accelerators with the same make, model, and nominal energy are known to be very similar, as long as the machines are unaltered from the manufacturer's original specifications. In this preliminary study, a quantitative investigation of the similarity in the basic photon dosimetry data from the Siemens Primus linear accelerators at eight different institutions is reported. The output factor, percentage depth dose (PDD), and in‐air off‐axis factor (OAF) for the 6 and 18 MV photon beams measured or verified by the Radiological Physics Center (RPC) were analyzed. The RPC‐measured output factors varied by less than about 2% for each field size. The difference between the maximum and minimum RPC‐verified PDD values at each depth was less than about 3%. The difference between the maximum and minimum RPC‐measured in‐air OAF was no more than 4% at all off‐axis distances considered in this study. These results strongly suggest that it is feasible to establish a reference photon dosimetry data set for each make, model, and nominal energy, universally applicable to those machines unaltered from the manufacturers' original specifications, within a clinically acceptable tolerance (e.g., [Formula: see text]). PACS number(s): 87.53.–j, 87.66.–a
format Online
Article
Text
id pubmed-5724467
institution National Center for Biotechnology Information
language English
publishDate 2003
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57244672018-04-02 Reference photon dosimetry data: A preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator Cho, S. H. Ibbott, G. S. J Appl Clin Med Phys Radiation Oncology Physics The dosimetric characteristics for modern computer‐controlled linear accelerators with the same make, model, and nominal energy are known to be very similar, as long as the machines are unaltered from the manufacturer's original specifications. In this preliminary study, a quantitative investigation of the similarity in the basic photon dosimetry data from the Siemens Primus linear accelerators at eight different institutions is reported. The output factor, percentage depth dose (PDD), and in‐air off‐axis factor (OAF) for the 6 and 18 MV photon beams measured or verified by the Radiological Physics Center (RPC) were analyzed. The RPC‐measured output factors varied by less than about 2% for each field size. The difference between the maximum and minimum RPC‐verified PDD values at each depth was less than about 3%. The difference between the maximum and minimum RPC‐measured in‐air OAF was no more than 4% at all off‐axis distances considered in this study. These results strongly suggest that it is feasible to establish a reference photon dosimetry data set for each make, model, and nominal energy, universally applicable to those machines unaltered from the manufacturers' original specifications, within a clinically acceptable tolerance (e.g., [Formula: see text]). PACS number(s): 87.53.–j, 87.66.–a John Wiley and Sons Inc. 2003-09-01 /pmc/articles/PMC5724467/ /pubmed/14604419 http://dx.doi.org/10.1120/jacmp.v4i4.2500 Text en © 2003 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Cho, S. H.
Ibbott, G. S.
Reference photon dosimetry data: A preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator
title Reference photon dosimetry data: A preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator
title_full Reference photon dosimetry data: A preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator
title_fullStr Reference photon dosimetry data: A preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator
title_full_unstemmed Reference photon dosimetry data: A preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator
title_short Reference photon dosimetry data: A preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator
title_sort reference photon dosimetry data: a preliminary study of in‐air off‐axis factor, percentage depth dose, and output factor of the siemens primus linear accelerator
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724467/
https://www.ncbi.nlm.nih.gov/pubmed/14604419
http://dx.doi.org/10.1120/jacmp.v4i4.2500
work_keys_str_mv AT chosh referencephotondosimetrydataapreliminarystudyofinairoffaxisfactorpercentagedepthdoseandoutputfactorofthesiemensprimuslinearaccelerator
AT ibbottgs referencephotondosimetrydataapreliminarystudyofinairoffaxisfactorpercentagedepthdoseandoutputfactorofthesiemensprimuslinearaccelerator