Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method

Monte Carlo calculations of the absorbed dose at various points of a brachytherapy anthropomorphic phantom are presented. The phantom walls and internal structures are made of polymethylmethacrylate and its external shape was taken from a female Alderson phantom. A complete Fletcher‐Green type appli...

Descripción completa

Detalles Bibliográficos
Autores principales: Rodríguez, Miguel L., deAlmeida, Carlos E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2002
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724533/
https://www.ncbi.nlm.nih.gov/pubmed/12383048
http://dx.doi.org/10.1120/jacmp.v3i4.2552
_version_ 1783285378314141696
author Rodríguez, Miguel L.
deAlmeida, Carlos E.
author_facet Rodríguez, Miguel L.
deAlmeida, Carlos E.
author_sort Rodríguez, Miguel L.
collection PubMed
description Monte Carlo calculations of the absorbed dose at various points of a brachytherapy anthropomorphic phantom are presented. The phantom walls and internal structures are made of polymethylmethacrylate and its external shape was taken from a female Alderson phantom. A complete Fletcher‐Green type applicator with the uterine tandem was fixed at the bottom of the phantom reproducing a typical geometrical configuration as that attained in a gynecological brachytherapy treatment. The dose rate produced by an array of five (137)Cs CDC‐J type sources placed in the applicator colpostats and the uterine tandem was evaluated by Monte Carlo simulations using the code penelope at three points: point A, the rectum, and the bladder. The influence of the applicator in the dose rate was evaluated by comparing Monte Carlo simulations of the sources alone and the sources inserted in the applicator. Differences up to 56% in the dose may be observed for the two cases in the planes including the rectum and bladder. The results show a reduction of the dose of 15.6%, 14.0%, and 5.6% in the rectum, bladder, and point A respectively, when the applicator wall and shieldings are considered. PACS number(s): 87.53Jw, 87.53.Wz, 87.53.Vb, 87.66.Xa
format Online
Article
Text
id pubmed-5724533
institution National Center for Biotechnology Information
language English
publishDate 2002
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57245332018-04-02 Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method Rodríguez, Miguel L. deAlmeida, Carlos E. J Appl Clin Med Phys Radiation Oncology Physics Monte Carlo calculations of the absorbed dose at various points of a brachytherapy anthropomorphic phantom are presented. The phantom walls and internal structures are made of polymethylmethacrylate and its external shape was taken from a female Alderson phantom. A complete Fletcher‐Green type applicator with the uterine tandem was fixed at the bottom of the phantom reproducing a typical geometrical configuration as that attained in a gynecological brachytherapy treatment. The dose rate produced by an array of five (137)Cs CDC‐J type sources placed in the applicator colpostats and the uterine tandem was evaluated by Monte Carlo simulations using the code penelope at three points: point A, the rectum, and the bladder. The influence of the applicator in the dose rate was evaluated by comparing Monte Carlo simulations of the sources alone and the sources inserted in the applicator. Differences up to 56% in the dose may be observed for the two cases in the planes including the rectum and bladder. The results show a reduction of the dose of 15.6%, 14.0%, and 5.6% in the rectum, bladder, and point A respectively, when the applicator wall and shieldings are considered. PACS number(s): 87.53Jw, 87.53.Wz, 87.53.Vb, 87.66.Xa John Wiley and Sons Inc. 2002-09-01 /pmc/articles/PMC5724533/ /pubmed/12383048 http://dx.doi.org/10.1120/jacmp.v3i4.2552 Text en © 2002 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Rodríguez, Miguel L.
deAlmeida, Carlos E.
Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method
title Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method
title_full Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method
title_fullStr Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method
title_full_unstemmed Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method
title_short Absorbed dose calculations in a brachytherapy pelvic phantom using the Monte Carlo method
title_sort absorbed dose calculations in a brachytherapy pelvic phantom using the monte carlo method
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724533/
https://www.ncbi.nlm.nih.gov/pubmed/12383048
http://dx.doi.org/10.1120/jacmp.v3i4.2552
work_keys_str_mv AT rodriguezmiguell absorbeddosecalculationsinabrachytherapypelvicphantomusingthemontecarlomethod
AT dealmeidacarlose absorbeddosecalculationsinabrachytherapypelvicphantomusingthemontecarlomethod