Monitor unit calculation for tangential breast treatments: Verification in an anthropomorphic phantom

This paper presents an anthropomorphic phantom study of dose delivered to a specific point during tangential breast irradiation to verify monitor unit calculations. Measurements were made using a 0.6 cc Farmer type cylindrical ionization chamber in the phantom and compared to calculations made on a...

Descripción completa

Detalles Bibliográficos
Autores principales: Howlett, Stephen J., Kron, Tomas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2002
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724595/
https://www.ncbi.nlm.nih.gov/pubmed/12132946
http://dx.doi.org/10.1120/jacmp.v3i3.2568
_version_ 1783285387113791488
author Howlett, Stephen J.
Kron, Tomas
author_facet Howlett, Stephen J.
Kron, Tomas
author_sort Howlett, Stephen J.
collection PubMed
description This paper presents an anthropomorphic phantom study of dose delivered to a specific point during tangential breast irradiation to verify monitor unit calculations. Measurements were made using a 0.6 cc Farmer type cylindrical ionization chamber in the phantom and compared to calculations made on a three‐dimensional radiotherapy treatment planning system using single digitized contour through to multi slice CT data. A large breast phantom was used for a single field size with a combination of open and wedged fields for three different energies (4, 6, and 18 MV). Solid flat phantom measurements were also performed for comparison. Results showed a lower calculated dose than the dose measured for a fixed number of monitor units where the variations were within a range of 0.8% to 4.5%. Differences were larger for the anthropomorphic phantom than the flat phantom. We conclude that little accuracy is gained from CT based monitor unit calculations compared to those based on digitised contours for this breast treatment but that the dose distributions will be affected. This type of test is recommended as one of a large set, in the commissioning and testing procedures for treatment planning systems. PACS number(s): 87.52.Df, 87.53.Dq, 87.53.Xd
format Online
Article
Text
id pubmed-5724595
institution National Center for Biotechnology Information
language English
publishDate 2002
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57245952018-04-02 Monitor unit calculation for tangential breast treatments: Verification in an anthropomorphic phantom Howlett, Stephen J. Kron, Tomas J Appl Clin Med Phys Radiation Oncology Physics This paper presents an anthropomorphic phantom study of dose delivered to a specific point during tangential breast irradiation to verify monitor unit calculations. Measurements were made using a 0.6 cc Farmer type cylindrical ionization chamber in the phantom and compared to calculations made on a three‐dimensional radiotherapy treatment planning system using single digitized contour through to multi slice CT data. A large breast phantom was used for a single field size with a combination of open and wedged fields for three different energies (4, 6, and 18 MV). Solid flat phantom measurements were also performed for comparison. Results showed a lower calculated dose than the dose measured for a fixed number of monitor units where the variations were within a range of 0.8% to 4.5%. Differences were larger for the anthropomorphic phantom than the flat phantom. We conclude that little accuracy is gained from CT based monitor unit calculations compared to those based on digitised contours for this breast treatment but that the dose distributions will be affected. This type of test is recommended as one of a large set, in the commissioning and testing procedures for treatment planning systems. PACS number(s): 87.52.Df, 87.53.Dq, 87.53.Xd John Wiley and Sons Inc. 2002-06-01 /pmc/articles/PMC5724595/ /pubmed/12132946 http://dx.doi.org/10.1120/jacmp.v3i3.2568 Text en © 2002 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Howlett, Stephen J.
Kron, Tomas
Monitor unit calculation for tangential breast treatments: Verification in an anthropomorphic phantom
title Monitor unit calculation for tangential breast treatments: Verification in an anthropomorphic phantom
title_full Monitor unit calculation for tangential breast treatments: Verification in an anthropomorphic phantom
title_fullStr Monitor unit calculation for tangential breast treatments: Verification in an anthropomorphic phantom
title_full_unstemmed Monitor unit calculation for tangential breast treatments: Verification in an anthropomorphic phantom
title_short Monitor unit calculation for tangential breast treatments: Verification in an anthropomorphic phantom
title_sort monitor unit calculation for tangential breast treatments: verification in an anthropomorphic phantom
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724595/
https://www.ncbi.nlm.nih.gov/pubmed/12132946
http://dx.doi.org/10.1120/jacmp.v3i3.2568
work_keys_str_mv AT howlettstephenj monitorunitcalculationfortangentialbreasttreatmentsverificationinananthropomorphicphantom
AT krontomas monitorunitcalculationfortangentialbreasttreatmentsverificationinananthropomorphicphantom