A quality control program for MR‐guided focused ultrasound ablation therapy

In this study, we propose a quality control program for MR‐guided focused ultrasound (FUS) ablation treatment to assess FUS beam positioning accuracy, FUS power delivery accuracy, MR imaging quality, and FUS ablation system safety. A total of 353 sonication points in Lucite cards were measured, the...

Descripción completa

Detalles Bibliográficos
Autores principales: Wu, Tao, Felmlee, Joel P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2002
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724613/
https://www.ncbi.nlm.nih.gov/pubmed/11958655
http://dx.doi.org/10.1120/jacmp.v3i2.2584
_version_ 1783285391438118912
author Wu, Tao
Felmlee, Joel P.
author_facet Wu, Tao
Felmlee, Joel P.
author_sort Wu, Tao
collection PubMed
description In this study, we propose a quality control program for MR‐guided focused ultrasound (FUS) ablation treatment to assess FUS beam positioning accuracy, FUS power delivery accuracy, MR imaging quality, and FUS ablation system safety. A total of 353 sonication points in Lucite cards were measured, the average placement errors were [Formula: see text] in the SI direction and [Formula: see text] in the LR direction. Temperature elevation was calculated from MR phase difference images and the measured water proton chemical shift (WPCS) temperature coefficient. WPCS temperature calibration for phantoms yielded a temperature coefficient of 0.011 ppm/°C. Sixteen experiments were conducted using six different phantoms to test the reliability of FUS power delivery. SNR and RF power calculated from phantom images were analyzed and stored at the MR console. A computer program was developed to integrate the system power delivery and the MR image quality control into one automated process. In the clinical trial at our institution, we expect this quality control program to be carried out before each patient treatment. If measured quality control values exceeds or below the preset values, a system service and retest should be conducted before the treatment. PACS number(s): 87.61.–c, 87.54.–n
format Online
Article
Text
id pubmed-5724613
institution National Center for Biotechnology Information
language English
publishDate 2002
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57246132018-04-02 A quality control program for MR‐guided focused ultrasound ablation therapy Wu, Tao Felmlee, Joel P. J Appl Clin Med Phys Medical Imaging In this study, we propose a quality control program for MR‐guided focused ultrasound (FUS) ablation treatment to assess FUS beam positioning accuracy, FUS power delivery accuracy, MR imaging quality, and FUS ablation system safety. A total of 353 sonication points in Lucite cards were measured, the average placement errors were [Formula: see text] in the SI direction and [Formula: see text] in the LR direction. Temperature elevation was calculated from MR phase difference images and the measured water proton chemical shift (WPCS) temperature coefficient. WPCS temperature calibration for phantoms yielded a temperature coefficient of 0.011 ppm/°C. Sixteen experiments were conducted using six different phantoms to test the reliability of FUS power delivery. SNR and RF power calculated from phantom images were analyzed and stored at the MR console. A computer program was developed to integrate the system power delivery and the MR image quality control into one automated process. In the clinical trial at our institution, we expect this quality control program to be carried out before each patient treatment. If measured quality control values exceeds or below the preset values, a system service and retest should be conducted before the treatment. PACS number(s): 87.61.–c, 87.54.–n John Wiley and Sons Inc. 2002-03-01 /pmc/articles/PMC5724613/ /pubmed/11958655 http://dx.doi.org/10.1120/jacmp.v3i2.2584 Text en © 2002 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Imaging
Wu, Tao
Felmlee, Joel P.
A quality control program for MR‐guided focused ultrasound ablation therapy
title A quality control program for MR‐guided focused ultrasound ablation therapy
title_full A quality control program for MR‐guided focused ultrasound ablation therapy
title_fullStr A quality control program for MR‐guided focused ultrasound ablation therapy
title_full_unstemmed A quality control program for MR‐guided focused ultrasound ablation therapy
title_short A quality control program for MR‐guided focused ultrasound ablation therapy
title_sort quality control program for mr‐guided focused ultrasound ablation therapy
topic Medical Imaging
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724613/
https://www.ncbi.nlm.nih.gov/pubmed/11958655
http://dx.doi.org/10.1120/jacmp.v3i2.2584
work_keys_str_mv AT wutao aqualitycontrolprogramformrguidedfocusedultrasoundablationtherapy
AT felmleejoelp aqualitycontrolprogramformrguidedfocusedultrasoundablationtherapy
AT wutao qualitycontrolprogramformrguidedfocusedultrasoundablationtherapy
AT felmleejoelp qualitycontrolprogramformrguidedfocusedultrasoundablationtherapy