A quality control program for MR‐guided focused ultrasound ablation therapy
In this study, we propose a quality control program for MR‐guided focused ultrasound (FUS) ablation treatment to assess FUS beam positioning accuracy, FUS power delivery accuracy, MR imaging quality, and FUS ablation system safety. A total of 353 sonication points in Lucite cards were measured, the...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2002
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724613/ https://www.ncbi.nlm.nih.gov/pubmed/11958655 http://dx.doi.org/10.1120/jacmp.v3i2.2584 |
_version_ | 1783285391438118912 |
---|---|
author | Wu, Tao Felmlee, Joel P. |
author_facet | Wu, Tao Felmlee, Joel P. |
author_sort | Wu, Tao |
collection | PubMed |
description | In this study, we propose a quality control program for MR‐guided focused ultrasound (FUS) ablation treatment to assess FUS beam positioning accuracy, FUS power delivery accuracy, MR imaging quality, and FUS ablation system safety. A total of 353 sonication points in Lucite cards were measured, the average placement errors were [Formula: see text] in the SI direction and [Formula: see text] in the LR direction. Temperature elevation was calculated from MR phase difference images and the measured water proton chemical shift (WPCS) temperature coefficient. WPCS temperature calibration for phantoms yielded a temperature coefficient of 0.011 ppm/°C. Sixteen experiments were conducted using six different phantoms to test the reliability of FUS power delivery. SNR and RF power calculated from phantom images were analyzed and stored at the MR console. A computer program was developed to integrate the system power delivery and the MR image quality control into one automated process. In the clinical trial at our institution, we expect this quality control program to be carried out before each patient treatment. If measured quality control values exceeds or below the preset values, a system service and retest should be conducted before the treatment. PACS number(s): 87.61.–c, 87.54.–n |
format | Online Article Text |
id | pubmed-5724613 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2002 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-57246132018-04-02 A quality control program for MR‐guided focused ultrasound ablation therapy Wu, Tao Felmlee, Joel P. J Appl Clin Med Phys Medical Imaging In this study, we propose a quality control program for MR‐guided focused ultrasound (FUS) ablation treatment to assess FUS beam positioning accuracy, FUS power delivery accuracy, MR imaging quality, and FUS ablation system safety. A total of 353 sonication points in Lucite cards were measured, the average placement errors were [Formula: see text] in the SI direction and [Formula: see text] in the LR direction. Temperature elevation was calculated from MR phase difference images and the measured water proton chemical shift (WPCS) temperature coefficient. WPCS temperature calibration for phantoms yielded a temperature coefficient of 0.011 ppm/°C. Sixteen experiments were conducted using six different phantoms to test the reliability of FUS power delivery. SNR and RF power calculated from phantom images were analyzed and stored at the MR console. A computer program was developed to integrate the system power delivery and the MR image quality control into one automated process. In the clinical trial at our institution, we expect this quality control program to be carried out before each patient treatment. If measured quality control values exceeds or below the preset values, a system service and retest should be conducted before the treatment. PACS number(s): 87.61.–c, 87.54.–n John Wiley and Sons Inc. 2002-03-01 /pmc/articles/PMC5724613/ /pubmed/11958655 http://dx.doi.org/10.1120/jacmp.v3i2.2584 Text en © 2002 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Medical Imaging Wu, Tao Felmlee, Joel P. A quality control program for MR‐guided focused ultrasound ablation therapy |
title | A quality control program for MR‐guided focused ultrasound ablation therapy |
title_full | A quality control program for MR‐guided focused ultrasound ablation therapy |
title_fullStr | A quality control program for MR‐guided focused ultrasound ablation therapy |
title_full_unstemmed | A quality control program for MR‐guided focused ultrasound ablation therapy |
title_short | A quality control program for MR‐guided focused ultrasound ablation therapy |
title_sort | quality control program for mr‐guided focused ultrasound ablation therapy |
topic | Medical Imaging |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724613/ https://www.ncbi.nlm.nih.gov/pubmed/11958655 http://dx.doi.org/10.1120/jacmp.v3i2.2584 |
work_keys_str_mv | AT wutao aqualitycontrolprogramformrguidedfocusedultrasoundablationtherapy AT felmleejoelp aqualitycontrolprogramformrguidedfocusedultrasoundablationtherapy AT wutao qualitycontrolprogramformrguidedfocusedultrasoundablationtherapy AT felmleejoelp qualitycontrolprogramformrguidedfocusedultrasoundablationtherapy |