Release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study

In the present study, amount of titanium (Ti) released into the surrounding bone during placement of implants with different surface structure was investigated. Quantification of Ti released during insertion from three different implants was performed in this ex vivo study. Jaw bone from pigs was us...

Descripción completa

Detalles Bibliográficos
Autores principales: Pettersson, Mattias, Pettersson, Jean, Molin Thorén, Margareta, Johansson, Anders
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Taylor & Francis 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724801/
https://www.ncbi.nlm.nih.gov/pubmed/29242814
http://dx.doi.org/10.1080/23337931.2017.1399270
_version_ 1783285415715799040
author Pettersson, Mattias
Pettersson, Jean
Molin Thorén, Margareta
Johansson, Anders
author_facet Pettersson, Mattias
Pettersson, Jean
Molin Thorén, Margareta
Johansson, Anders
author_sort Pettersson, Mattias
collection PubMed
description In the present study, amount of titanium (Ti) released into the surrounding bone during placement of implants with different surface structure was investigated. Quantification of Ti released during insertion from three different implants was performed in this ex vivo study. Jaw bone from pigs was used as model for installation of the implants and Inductively Coupled Plasma Atomic Emission Spectroscopy (ICP-AES) was used for analysis of the released Ti. Implant surface were examined with scanning electron microscopy (SEM), before and after the placement into the bone. Ti was abraded to the surrounding bone upon insertion of a dental implant and the surface roughness of the implant increased the amount of Ti found. Diameter and total area of the implant were of less importance for the Ti released to the bone. No visible damages to the implant surfaces could be identified in SEM after placement.
format Online
Article
Text
id pubmed-5724801
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Taylor & Francis
record_format MEDLINE/PubMed
spelling pubmed-57248012017-12-14 Release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study Pettersson, Mattias Pettersson, Jean Molin Thorén, Margareta Johansson, Anders Acta Biomater Odontol Scand Original Article In the present study, amount of titanium (Ti) released into the surrounding bone during placement of implants with different surface structure was investigated. Quantification of Ti released during insertion from three different implants was performed in this ex vivo study. Jaw bone from pigs was used as model for installation of the implants and Inductively Coupled Plasma Atomic Emission Spectroscopy (ICP-AES) was used for analysis of the released Ti. Implant surface were examined with scanning electron microscopy (SEM), before and after the placement into the bone. Ti was abraded to the surrounding bone upon insertion of a dental implant and the surface roughness of the implant increased the amount of Ti found. Diameter and total area of the implant were of less importance for the Ti released to the bone. No visible damages to the implant surfaces could be identified in SEM after placement. Taylor & Francis 2017-11-14 /pmc/articles/PMC5724801/ /pubmed/29242814 http://dx.doi.org/10.1080/23337931.2017.1399270 Text en © 2017 The Author(s). Published by Informa UK Limited, trading as Taylor & Francis Group. http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Pettersson, Mattias
Pettersson, Jean
Molin Thorén, Margareta
Johansson, Anders
Release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study
title Release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study
title_full Release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study
title_fullStr Release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study
title_full_unstemmed Release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study
title_short Release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study
title_sort release of titanium after insertion of dental implants with different surface characteristics – an ex vivo animal study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5724801/
https://www.ncbi.nlm.nih.gov/pubmed/29242814
http://dx.doi.org/10.1080/23337931.2017.1399270
work_keys_str_mv AT petterssonmattias releaseoftitaniumafterinsertionofdentalimplantswithdifferentsurfacecharacteristicsanexvivoanimalstudy
AT petterssonjean releaseoftitaniumafterinsertionofdentalimplantswithdifferentsurfacecharacteristicsanexvivoanimalstudy
AT molinthorenmargareta releaseoftitaniumafterinsertionofdentalimplantswithdifferentsurfacecharacteristicsanexvivoanimalstudy
AT johanssonanders releaseoftitaniumafterinsertionofdentalimplantswithdifferentsurfacecharacteristicsanexvivoanimalstudy