A method for checking high dose rate treatment times for vaginal applicators
A method is presented for checking the treatment time calculation for high dose rate (HDR) vaginal cylinder treatments. The method represents an independent check of the HDR planning system and can take into account nonuniform isodose line coverage around the cylinder. Only the air kerma strength of...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2001
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5726014/ https://www.ncbi.nlm.nih.gov/pubmed/11686739 http://dx.doi.org/10.1120/jacmp.v2i4.2595 |
Sumario: | A method is presented for checking the treatment time calculation for high dose rate (HDR) vaginal cylinder treatments. The method represents an independent check of the HDR planning system and can take into account nonuniform isodose line coverage around the cylinder. Only the air kerma strength of the source and information that is available from the written directive are required. The maximum discrepancy for a representative set of cylinder plans done on a Nucletron unit was 5%. A working HTML JavaScript program is included in the Appendix. PACS number(s): 87.53.–j, 87.90.+y |
---|