Cargando…

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

The checking of monitor unit calculations is recognized as a vital component of quality assurance in radiotherapy. Using straightforward but detailed computer‐based verification calculations it is possible to achieve a precision of 1% when compared with a three‐dimensional (3D) treatment planning sy...

Descripción completa

Detalles Bibliográficos
Autores principales: Leszczynski, Konrad W., Dunscombe, Peter B.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2000
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5726156/
https://www.ncbi.nlm.nih.gov/pubmed/11674827
http://dx.doi.org/10.1120/jacmp.v1i4.2633
Descripción
Sumario:The checking of monitor unit calculations is recognized as a vital component of quality assurance in radiotherapy. Using straightforward but detailed computer‐based verification calculations it is possible to achieve a precision of 1% when compared with a three‐dimensional (3D) treatment planning system monitor unit calculation. The method is sufficiently sensitive to identify significant errors and is consistent with current recommendations on the magnitude of uncertainties in clinical dosimetry. Moreover, the approach is accurate in the sense of being highly consistent with the validated 3D treatment planning system's calculations. PACS number(s): 87.53.–j, 87.52.–g