Cargando…

Use of a radioactive check device for redundancy check of ionization chambers

The use of a radioactive check device containing Strontium 90 was investigated to carry out the redundancy checks of Farmer‐type ionization chambers. It was not possible to meet the recommended tolerance limits of the redundancy checks without taking into account the angular response of the ionizati...

Descripción completa

Detalles Bibliográficos
Autores principales: Sidhu, N. P. S., Rouvas, Alkis, Cadman, Patrick
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2000
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5726160/
https://www.ncbi.nlm.nih.gov/pubmed/11674830
http://dx.doi.org/10.1120/jacmp.v1i4.2636
_version_ 1783285686388916224
author Sidhu, N. P. S.
Rouvas, Alkis
Cadman, Patrick
author_facet Sidhu, N. P. S.
Rouvas, Alkis
Cadman, Patrick
author_sort Sidhu, N. P. S.
collection PubMed
description The use of a radioactive check device containing Strontium 90 was investigated to carry out the redundancy checks of Farmer‐type ionization chambers. It was not possible to meet the recommended tolerance limits of the redundancy checks without taking into account the angular response of the ionization chambers. The ionization chambers exhibited a maximum variation of 1% in the angular response in this check device. After accounting for the angular response the maximum variation of the short‐term repeatability was 0.14% with a standard deviation of [Formula: see text]. The long‐term constancy checked over a period of nine months was less than [Formula: see text] for measurements, taking into account the angular response of the ionization chambers. No significant effect of the background radiation levels was observed on the measurements. PACS number(s): 87.53.–j, 87.52.–g
format Online
Article
Text
id pubmed-5726160
institution National Center for Biotechnology Information
language English
publishDate 2000
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-57261602018-04-02 Use of a radioactive check device for redundancy check of ionization chambers Sidhu, N. P. S. Rouvas, Alkis Cadman, Patrick J Appl Clin Med Phys Radiation Oncology Physics The use of a radioactive check device containing Strontium 90 was investigated to carry out the redundancy checks of Farmer‐type ionization chambers. It was not possible to meet the recommended tolerance limits of the redundancy checks without taking into account the angular response of the ionization chambers. The ionization chambers exhibited a maximum variation of 1% in the angular response in this check device. After accounting for the angular response the maximum variation of the short‐term repeatability was 0.14% with a standard deviation of [Formula: see text]. The long‐term constancy checked over a period of nine months was less than [Formula: see text] for measurements, taking into account the angular response of the ionization chambers. No significant effect of the background radiation levels was observed on the measurements. PACS number(s): 87.53.–j, 87.52.–g John Wiley and Sons Inc. 2000-09-01 /pmc/articles/PMC5726160/ /pubmed/11674830 http://dx.doi.org/10.1120/jacmp.v1i4.2636 Text en © 2000 The Authors. This is an open access article under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/3.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Sidhu, N. P. S.
Rouvas, Alkis
Cadman, Patrick
Use of a radioactive check device for redundancy check of ionization chambers
title Use of a radioactive check device for redundancy check of ionization chambers
title_full Use of a radioactive check device for redundancy check of ionization chambers
title_fullStr Use of a radioactive check device for redundancy check of ionization chambers
title_full_unstemmed Use of a radioactive check device for redundancy check of ionization chambers
title_short Use of a radioactive check device for redundancy check of ionization chambers
title_sort use of a radioactive check device for redundancy check of ionization chambers
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5726160/
https://www.ncbi.nlm.nih.gov/pubmed/11674830
http://dx.doi.org/10.1120/jacmp.v1i4.2636
work_keys_str_mv AT sidhunps useofaradioactivecheckdeviceforredundancycheckofionizationchambers
AT rouvasalkis useofaradioactivecheckdeviceforredundancycheckofionizationchambers
AT cadmanpatrick useofaradioactivecheckdeviceforredundancycheckofionizationchambers