Accuracy and Repeatability of Automated Injector Versus Manual Administration of an MRI Contrast Agent—Results of a Laboratory Study

OBJECTIVE: The aim of this study was to compare flow rates over time and the deviations from the target flow rate of a magnetic resonance imaging contrast agent achieved by an automated injector versus manual injection. MATERIALS AND METHODS: In this laboratory study, the magnetic resonance contrast...

Descripción completa

Detalles Bibliográficos
Autores principales: Endrikat, Jan, Barbati, Ron, Scarpa, Marcella, Jost, Gregor, (Ned) Uber, Arthur E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5728584/
https://www.ncbi.nlm.nih.gov/pubmed/28796722
http://dx.doi.org/10.1097/RLI.0000000000000403
Descripción
Sumario:OBJECTIVE: The aim of this study was to compare flow rates over time and the deviations from the target flow rate of a magnetic resonance imaging contrast agent achieved by an automated injector versus manual injection. MATERIALS AND METHODS: In this laboratory study, the magnetic resonance contrast agent gadobutrol was repeatedly injected by an injector and by 10 experienced technologists. Six scenarios with 2 different target flow rates (1 and 5 mL/s), 2 different contrast volumes (10 and 20 mL), and 2 different intravenous (IV) catheters (22 gauge and 20 gauge) were tested. The flow rates over time were recorded. The target variable was the average absolute deviation and average absolute percentage deviation from the target flow rate. RESULTS: The flow rates over time achieved by an injector were almost identical. Slight deviations from the target flow rate occurred during ramp-up and ramp-down only. Those of manual injection showed high variability over the whole course of the injection. In the 1 mL/s scenarios, the injector deviated from the target flow rate by 0.06 mL/s or less (≤6%) and in the 5 mL/s scenarios by 1.02 mL/s or less (<20%). For the manual injections at the same flow rates, these figures were 0.35 mL/s or less (≤35%) and 3.1 mL/s or less (≤62%). CONCLUSIONS: Injector administration of a magnetic resonance contrast agent minimally deviated from the target flow rate, whereas manual injection varied widely. Injector administration is more accurate and repeatable.