Pacing failure caused by automatic pacing threshold adjustment system

Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety...

Descripción completa

Detalles Bibliográficos
Autores principales: Kishihara, Jun, Niwano, Shinichi, Fukaya, Hidehira, Nishinarita, Ryo, Horiguchi, Ai, Nakamura, Hironori, Igarashi, Tazuru, Ishizue, Naruya, Fujiishi, Tamami, Yoshizawa, Tomoharu, Oikawa, Jun, Satoh, Akira, Murakami, Masami, Ako, Junya
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5729025/
https://www.ncbi.nlm.nih.gov/pubmed/29255515
http://dx.doi.org/10.1016/j.joa.2017.05.005
_version_ 1783286131649937408
author Kishihara, Jun
Niwano, Shinichi
Fukaya, Hidehira
Nishinarita, Ryo
Horiguchi, Ai
Nakamura, Hironori
Igarashi, Tazuru
Ishizue, Naruya
Fujiishi, Tamami
Yoshizawa, Tomoharu
Oikawa, Jun
Satoh, Akira
Murakami, Masami
Ako, Junya
author_facet Kishihara, Jun
Niwano, Shinichi
Fukaya, Hidehira
Nishinarita, Ryo
Horiguchi, Ai
Nakamura, Hironori
Igarashi, Tazuru
Ishizue, Naruya
Fujiishi, Tamami
Yoshizawa, Tomoharu
Oikawa, Jun
Satoh, Akira
Murakami, Masami
Ako, Junya
author_sort Kishihara, Jun
collection PubMed
description Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient.
format Online
Article
Text
id pubmed-5729025
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-57290252017-12-18 Pacing failure caused by automatic pacing threshold adjustment system Kishihara, Jun Niwano, Shinichi Fukaya, Hidehira Nishinarita, Ryo Horiguchi, Ai Nakamura, Hironori Igarashi, Tazuru Ishizue, Naruya Fujiishi, Tamami Yoshizawa, Tomoharu Oikawa, Jun Satoh, Akira Murakami, Masami Ako, Junya J Arrhythm Case Report Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient. Elsevier 2017-12 2017-10-13 /pmc/articles/PMC5729025/ /pubmed/29255515 http://dx.doi.org/10.1016/j.joa.2017.05.005 Text en © 2017 Japanese Heart Rhythm Society. Published by Elsevier B.V. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Case Report
Kishihara, Jun
Niwano, Shinichi
Fukaya, Hidehira
Nishinarita, Ryo
Horiguchi, Ai
Nakamura, Hironori
Igarashi, Tazuru
Ishizue, Naruya
Fujiishi, Tamami
Yoshizawa, Tomoharu
Oikawa, Jun
Satoh, Akira
Murakami, Masami
Ako, Junya
Pacing failure caused by automatic pacing threshold adjustment system
title Pacing failure caused by automatic pacing threshold adjustment system
title_full Pacing failure caused by automatic pacing threshold adjustment system
title_fullStr Pacing failure caused by automatic pacing threshold adjustment system
title_full_unstemmed Pacing failure caused by automatic pacing threshold adjustment system
title_short Pacing failure caused by automatic pacing threshold adjustment system
title_sort pacing failure caused by automatic pacing threshold adjustment system
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5729025/
https://www.ncbi.nlm.nih.gov/pubmed/29255515
http://dx.doi.org/10.1016/j.joa.2017.05.005
work_keys_str_mv AT kishiharajun pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT niwanoshinichi pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT fukayahidehira pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT nishinaritaryo pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT horiguchiai pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT nakamurahironori pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT igarashitazuru pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT ishizuenaruya pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT fujiishitamami pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT yoshizawatomoharu pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT oikawajun pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT satohakira pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT murakamimasami pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem
AT akojunya pacingfailurecausedbyautomaticpacingthresholdadjustmentsystem