Cargando…

Development of an LCD-Based Visual Field System

BACKGROUND: The present study investigated the diagnostic effectiveness of an LCD-based visual field testing system (LVF) in comparison with the standard automated perimetry Humphrey Field Analyzer II-750i (HFA). METHODS: A randomized controlled crossover study was conducted with 202 normal and 128...

Descripción completa

Detalles Bibliográficos
Autores principales: Joo, Jin-Ho, Lee, Jihyoung, You, Heecheon, Kang, Jaheon
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Medical Sciences 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5729647/
https://www.ncbi.nlm.nih.gov/pubmed/29215806
http://dx.doi.org/10.3346/jkms.2018.33.e19
Descripción
Sumario:BACKGROUND: The present study investigated the diagnostic effectiveness of an LCD-based visual field testing system (LVF) in comparison with the standard automated perimetry Humphrey Field Analyzer II-750i (HFA). METHODS: A randomized controlled crossover study was conducted with 202 normal and 128 glaucomatous eyes using both LVF and HFA. The visual field testing systems were compared in terms of mean deviation (MD), pattern standard deviation (PSD), and area under the receiver operating characteristics curve (AUC) of MD and PSD differentiating the normal and glaucomatous eyes. RESULTS: Significant correlations were found between MD measurements from LVF and those from HFA for normal eyes (r = 0.342) and glaucomatous eyes (r = 0.796); slightly higher significant correlations were identified between PSD measurements from LVF and those from HFA for normal eyes (r = 0.363) and glaucomatous eyes (r = 0.828). Furthermore, high AUCs of MD were found as 0.786 for LVF and 0.868 for HFA and AUCs of PSD as 0.913 for LVF and 0.932 for HFA. CONCLUSION: The comparison results of the present study support the competence of LVF compared with HFA in visual field testing for early detection of glaucoma.