Improved visualization of breast tissue on a dedicated breast PET system through ergonomic redesign of the imaging table

BACKGROUND: Breast lesions closer than 2 cm to the chest wall are difficult to position in the field of view of dedicated breast PET (db-PET) systems. This inability to detect such lesions is a significant limitation of these systems. The primary objective of this study was to determine if modificat...

Descripción completa

Detalles Bibliográficos
Autores principales: O’Connor, Michael K., Tran, Thuy D., Swanson, Tiffinee N., Ellingson, Lacey R., Hunt, Katie N., Whaley, Dana H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5736520/
https://www.ncbi.nlm.nih.gov/pubmed/29260333
http://dx.doi.org/10.1186/s13550-017-0351-7
_version_ 1783287366806405120
author O’Connor, Michael K.
Tran, Thuy D.
Swanson, Tiffinee N.
Ellingson, Lacey R.
Hunt, Katie N.
Whaley, Dana H.
author_facet O’Connor, Michael K.
Tran, Thuy D.
Swanson, Tiffinee N.
Ellingson, Lacey R.
Hunt, Katie N.
Whaley, Dana H.
author_sort O’Connor, Michael K.
collection PubMed
description BACKGROUND: Breast lesions closer than 2 cm to the chest wall are difficult to position in the field of view of dedicated breast PET (db-PET) systems. This inability to detect such lesions is a significant limitation of these systems. The primary objective of this study was to determine if modifications to the design of the imaging table and detector used for a db-PET system would enable improved visualization of breast tissue close to the chest wall. All studies were performed on a commercially available db-PET system (Mammi-PET). A central square section of the imaging table, containing the standard 180-mm circular aperture, was modified such that it could be removed and replaced by thinner sections with a larger aperture. Additional changes were made to the cover plate of the detector array and the patient mattress. A total of 60 patients were studied. After administration of F-18 FDG, 30 patients were imaged with a 220-mm-diameter aperture and the standard aperture, and 30 patients with a 200-mm aperture and the standard aperture. On all scans, the length of breast tissue in the field of view was measured as the greatest extent of tissue from the nipple back to the posterior edge of the breast. Image quality and patient comfort were recorded. RESULTS: Averaged over both breasts, relative to the standard aperture, the increase in breast length was 12.5 + 7.7 mm with the 220-mm aperture, and 12.3 + 6.5 mm with the 200-mm aperture (p < 0.05 for both apertures). In ~ 5% of cases, the larger apertures resulted in some degradation in image quality due to closer proximity to cardiac/hepatic activity. In 10–20% of cases, movement of the breast tissue was observed as the detector ring was moved to scan the anterior region of the breast. The patient survey indicated no significant difference in the comfort level between the standard aperture and either of the prototype apertures. CONCLUSIONS: Modifications to the image table and system resulted in a significant gain in the volume of breast tissue that could be imaged on the db-PET system and should allow better visualization of lesions close to the chest wall.
format Online
Article
Text
id pubmed-5736520
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-57365202017-12-20 Improved visualization of breast tissue on a dedicated breast PET system through ergonomic redesign of the imaging table O’Connor, Michael K. Tran, Thuy D. Swanson, Tiffinee N. Ellingson, Lacey R. Hunt, Katie N. Whaley, Dana H. EJNMMI Res Original Research BACKGROUND: Breast lesions closer than 2 cm to the chest wall are difficult to position in the field of view of dedicated breast PET (db-PET) systems. This inability to detect such lesions is a significant limitation of these systems. The primary objective of this study was to determine if modifications to the design of the imaging table and detector used for a db-PET system would enable improved visualization of breast tissue close to the chest wall. All studies were performed on a commercially available db-PET system (Mammi-PET). A central square section of the imaging table, containing the standard 180-mm circular aperture, was modified such that it could be removed and replaced by thinner sections with a larger aperture. Additional changes were made to the cover plate of the detector array and the patient mattress. A total of 60 patients were studied. After administration of F-18 FDG, 30 patients were imaged with a 220-mm-diameter aperture and the standard aperture, and 30 patients with a 200-mm aperture and the standard aperture. On all scans, the length of breast tissue in the field of view was measured as the greatest extent of tissue from the nipple back to the posterior edge of the breast. Image quality and patient comfort were recorded. RESULTS: Averaged over both breasts, relative to the standard aperture, the increase in breast length was 12.5 + 7.7 mm with the 220-mm aperture, and 12.3 + 6.5 mm with the 200-mm aperture (p < 0.05 for both apertures). In ~ 5% of cases, the larger apertures resulted in some degradation in image quality due to closer proximity to cardiac/hepatic activity. In 10–20% of cases, movement of the breast tissue was observed as the detector ring was moved to scan the anterior region of the breast. The patient survey indicated no significant difference in the comfort level between the standard aperture and either of the prototype apertures. CONCLUSIONS: Modifications to the image table and system resulted in a significant gain in the volume of breast tissue that could be imaged on the db-PET system and should allow better visualization of lesions close to the chest wall. Springer Berlin Heidelberg 2017-12-19 /pmc/articles/PMC5736520/ /pubmed/29260333 http://dx.doi.org/10.1186/s13550-017-0351-7 Text en © The Author(s). 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
O’Connor, Michael K.
Tran, Thuy D.
Swanson, Tiffinee N.
Ellingson, Lacey R.
Hunt, Katie N.
Whaley, Dana H.
Improved visualization of breast tissue on a dedicated breast PET system through ergonomic redesign of the imaging table
title Improved visualization of breast tissue on a dedicated breast PET system through ergonomic redesign of the imaging table
title_full Improved visualization of breast tissue on a dedicated breast PET system through ergonomic redesign of the imaging table
title_fullStr Improved visualization of breast tissue on a dedicated breast PET system through ergonomic redesign of the imaging table
title_full_unstemmed Improved visualization of breast tissue on a dedicated breast PET system through ergonomic redesign of the imaging table
title_short Improved visualization of breast tissue on a dedicated breast PET system through ergonomic redesign of the imaging table
title_sort improved visualization of breast tissue on a dedicated breast pet system through ergonomic redesign of the imaging table
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5736520/
https://www.ncbi.nlm.nih.gov/pubmed/29260333
http://dx.doi.org/10.1186/s13550-017-0351-7
work_keys_str_mv AT oconnormichaelk improvedvisualizationofbreasttissueonadedicatedbreastpetsystemthroughergonomicredesignoftheimagingtable
AT tranthuyd improvedvisualizationofbreasttissueonadedicatedbreastpetsystemthroughergonomicredesignoftheimagingtable
AT swansontiffineen improvedvisualizationofbreasttissueonadedicatedbreastpetsystemthroughergonomicredesignoftheimagingtable
AT ellingsonlaceyr improvedvisualizationofbreasttissueonadedicatedbreastpetsystemthroughergonomicredesignoftheimagingtable
AT huntkatien improvedvisualizationofbreasttissueonadedicatedbreastpetsystemthroughergonomicredesignoftheimagingtable
AT whaleydanah improvedvisualizationofbreasttissueonadedicatedbreastpetsystemthroughergonomicredesignoftheimagingtable