Late Disruption of a Polyethylene Terephthalate Aortic Graft 30 Years after Initial Graft Placement

A 71 year old male who had undergone extra-anatomic bypass grafting between the ascending aorta and the thoraco-abdominal aorta at 41 years of age for aortic coarctation was admitted with back pain and dyspnea. A 16 mm Cooley double velour knitted polyethylene terephthalate (PET) graft was used in t...

Descripción completa

Detalles Bibliográficos
Autores principales: Matsushita, Akihito, Tsunoda, Yu, Hattori, Takashi, Mihara, Wahei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5737948/
https://www.ncbi.nlm.nih.gov/pubmed/29276787
http://dx.doi.org/10.1016/j.ejvssr.2017.11.001
Descripción
Sumario:A 71 year old male who had undergone extra-anatomic bypass grafting between the ascending aorta and the thoraco-abdominal aorta at 41 years of age for aortic coarctation was admitted with back pain and dyspnea. A 16 mm Cooley double velour knitted polyethylene terephthalate (PET) graft was used in the initial operation in 1983. Computed tomography showed disruption of the initial PET graft perforating the right atrium, and a pseudoaneurysm at the distal anastomosis. The patient was in acute cardiac failure because of left to right shunting. A two stage operation was performed. The first stage comprised emergency re-grafting and right atrium repair, and the second stage re-grafting for the pseudoaneurysm. The patient is doing well 48 months post-operatively; however, monitoring of the patient will continue for potential PET graft rupture.