Diffraction gratings metrology and ray-tracing results for an XUV Raman spectrometer at FLASH

The extreme-ultraviolet double-stage imaging Raman spectrometer is a permanent experimental endstation at the plane-grating monochromator beamline branch PG1 at FLASH at DESY in Hamburg, Germany. This unique instrument covers the photon energy range from 20 to 200 eV with high energy resolution of a...

Descripción completa

Detalles Bibliográficos
Autores principales: Dziarzhytski, Siarhei, Siewert, Frank, Sokolov, Andrey, Gwalt, Grzegorz, Seliger, Tino, Rübhausen, Michael, Weigelt, Holger, Brenner, Günter
Formato: Online Artículo Texto
Lenguaje:English
Publicado: International Union of Crystallography 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5741130/
https://www.ncbi.nlm.nih.gov/pubmed/29271763
http://dx.doi.org/10.1107/S1600577517013066
_version_ 1783288148209434624
author Dziarzhytski, Siarhei
Siewert, Frank
Sokolov, Andrey
Gwalt, Grzegorz
Seliger, Tino
Rübhausen, Michael
Weigelt, Holger
Brenner, Günter
author_facet Dziarzhytski, Siarhei
Siewert, Frank
Sokolov, Andrey
Gwalt, Grzegorz
Seliger, Tino
Rübhausen, Michael
Weigelt, Holger
Brenner, Günter
author_sort Dziarzhytski, Siarhei
collection PubMed
description The extreme-ultraviolet double-stage imaging Raman spectrometer is a permanent experimental endstation at the plane-grating monochromator beamline branch PG1 at FLASH at DESY in Hamburg, Germany. This unique instrument covers the photon energy range from 20 to 200 eV with high energy resolution of about 2 to 20 meV (design values) featuring an efficient elastic line suppression as well as effective stray light rejection. Such a design enables studies of low-energy excitations like, for example, phonons in solids close to the vicinity of the elastic line. The Raman spectrometer effectively operates with four reflective off-axial parabolic mirrors and two plane-grating units. The optics quality and their precise alignment are crucial to guarantee best performance of the instrument. Here, results on a comprehensive investigation of the quality of the spectrometer diffraction gratings are presented. The gratings have been characterized by ex situ metrology at the BESSY-II Optics Laboratory, employing slope measuring deflectometry and interferometry as well as atomic force microscopy studies. The efficiency of these key optical elements has been measured at the at-wavelength metrology laboratory using the reflectometer at the BESSY-II Optics beamline. Also, the metrology results are discussed with respect to the expected resolving power of the instrument by including them in ray-tracing studies of the instrument.
format Online
Article
Text
id pubmed-5741130
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher International Union of Crystallography
record_format MEDLINE/PubMed
spelling pubmed-57411302018-01-01 Diffraction gratings metrology and ray-tracing results for an XUV Raman spectrometer at FLASH Dziarzhytski, Siarhei Siewert, Frank Sokolov, Andrey Gwalt, Grzegorz Seliger, Tino Rübhausen, Michael Weigelt, Holger Brenner, Günter J Synchrotron Radiat Photondiag2017 Workshop The extreme-ultraviolet double-stage imaging Raman spectrometer is a permanent experimental endstation at the plane-grating monochromator beamline branch PG1 at FLASH at DESY in Hamburg, Germany. This unique instrument covers the photon energy range from 20 to 200 eV with high energy resolution of about 2 to 20 meV (design values) featuring an efficient elastic line suppression as well as effective stray light rejection. Such a design enables studies of low-energy excitations like, for example, phonons in solids close to the vicinity of the elastic line. The Raman spectrometer effectively operates with four reflective off-axial parabolic mirrors and two plane-grating units. The optics quality and their precise alignment are crucial to guarantee best performance of the instrument. Here, results on a comprehensive investigation of the quality of the spectrometer diffraction gratings are presented. The gratings have been characterized by ex situ metrology at the BESSY-II Optics Laboratory, employing slope measuring deflectometry and interferometry as well as atomic force microscopy studies. The efficiency of these key optical elements has been measured at the at-wavelength metrology laboratory using the reflectometer at the BESSY-II Optics beamline. Also, the metrology results are discussed with respect to the expected resolving power of the instrument by including them in ray-tracing studies of the instrument. International Union of Crystallography 2018-01-01 /pmc/articles/PMC5741130/ /pubmed/29271763 http://dx.doi.org/10.1107/S1600577517013066 Text en © Siarhei Dziarzhytski et al. 2018 http://creativecommons.org/licenses/by/2.0/uk/ This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.http://creativecommons.org/licenses/by/2.0/uk/
spellingShingle Photondiag2017 Workshop
Dziarzhytski, Siarhei
Siewert, Frank
Sokolov, Andrey
Gwalt, Grzegorz
Seliger, Tino
Rübhausen, Michael
Weigelt, Holger
Brenner, Günter
Diffraction gratings metrology and ray-tracing results for an XUV Raman spectrometer at FLASH
title Diffraction gratings metrology and ray-tracing results for an XUV Raman spectrometer at FLASH
title_full Diffraction gratings metrology and ray-tracing results for an XUV Raman spectrometer at FLASH
title_fullStr Diffraction gratings metrology and ray-tracing results for an XUV Raman spectrometer at FLASH
title_full_unstemmed Diffraction gratings metrology and ray-tracing results for an XUV Raman spectrometer at FLASH
title_short Diffraction gratings metrology and ray-tracing results for an XUV Raman spectrometer at FLASH
title_sort diffraction gratings metrology and ray-tracing results for an xuv raman spectrometer at flash
topic Photondiag2017 Workshop
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5741130/
https://www.ncbi.nlm.nih.gov/pubmed/29271763
http://dx.doi.org/10.1107/S1600577517013066
work_keys_str_mv AT dziarzhytskisiarhei diffractiongratingsmetrologyandraytracingresultsforanxuvramanspectrometeratflash
AT siewertfrank diffractiongratingsmetrologyandraytracingresultsforanxuvramanspectrometeratflash
AT sokolovandrey diffractiongratingsmetrologyandraytracingresultsforanxuvramanspectrometeratflash
AT gwaltgrzegorz diffractiongratingsmetrologyandraytracingresultsforanxuvramanspectrometeratflash
AT seligertino diffractiongratingsmetrologyandraytracingresultsforanxuvramanspectrometeratflash
AT rubhausenmichael diffractiongratingsmetrologyandraytracingresultsforanxuvramanspectrometeratflash
AT weigeltholger diffractiongratingsmetrologyandraytracingresultsforanxuvramanspectrometeratflash
AT brennergunter diffractiongratingsmetrologyandraytracingresultsforanxuvramanspectrometeratflash