Verification of Dosimetric Commissioning Accuracy of Intensity Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Delivery using Task Group-119 Guidelines
AIM: The purpose of this study is to verify the accuracy of the commissioning of intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) based on the recommendation of the American Association of Physicists in Medicine Task Group 119 (TG-119). MATERIALS AND METHODS:...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5744455/ https://www.ncbi.nlm.nih.gov/pubmed/29296041 http://dx.doi.org/10.4103/jmp.JMP_16_17 |
_version_ | 1783288748761415680 |
---|---|
author | Kaviarasu, Karunakaran Nambi Raj, N. Arunai Hamid, Misba Giri Babu, A. Ananda Sreenivas, Lingampally Murthy, Kammari Krishna |
author_facet | Kaviarasu, Karunakaran Nambi Raj, N. Arunai Hamid, Misba Giri Babu, A. Ananda Sreenivas, Lingampally Murthy, Kammari Krishna |
author_sort | Kaviarasu, Karunakaran |
collection | PubMed |
description | AIM: The purpose of this study is to verify the accuracy of the commissioning of intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) based on the recommendation of the American Association of Physicists in Medicine Task Group 119 (TG-119). MATERIALS AND METHODS: TG-119 proposes a set of clinical test cases to verify the accuracy of IMRT planning and delivery system. For these test cases, we generated two sets of treatment plans, the first plan using 7–9 IMRT fields and a second plan utilizing two-arc VMAT technique for both 6 MV and 15 MV photon beams. The template plans of TG-119 were optimized and calculated by Varian Eclipse Treatment Planning System (version 13.5). Dose prescription and planning objectives were set according to the TG-119 goals. The point dose (mean dose to the contoured chamber volume) at the specified positions/locations was measured using compact (CC-13) ion chamber. The composite planar dose was measured with IMatriXX Evaluation 2D array with OmniPro IMRT Software (version 1.7b). The per-field relative gamma was measured using electronic portal imaging device in a way similar to the routine pretreatment patient-specific quality assurance. RESULTS: Our planning results are compared with the TG-119 data. Point dose and fluence comparison data where within the acceptable confident limit. CONCLUSION: From the obtained data in this study, we conclude that the commissioning of IMRT and VMAT delivery were found within the limits of TG-119. |
format | Online Article Text |
id | pubmed-5744455 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-57444552018-01-02 Verification of Dosimetric Commissioning Accuracy of Intensity Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Delivery using Task Group-119 Guidelines Kaviarasu, Karunakaran Nambi Raj, N. Arunai Hamid, Misba Giri Babu, A. Ananda Sreenivas, Lingampally Murthy, Kammari Krishna J Med Phys Technical Note AIM: The purpose of this study is to verify the accuracy of the commissioning of intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) based on the recommendation of the American Association of Physicists in Medicine Task Group 119 (TG-119). MATERIALS AND METHODS: TG-119 proposes a set of clinical test cases to verify the accuracy of IMRT planning and delivery system. For these test cases, we generated two sets of treatment plans, the first plan using 7–9 IMRT fields and a second plan utilizing two-arc VMAT technique for both 6 MV and 15 MV photon beams. The template plans of TG-119 were optimized and calculated by Varian Eclipse Treatment Planning System (version 13.5). Dose prescription and planning objectives were set according to the TG-119 goals. The point dose (mean dose to the contoured chamber volume) at the specified positions/locations was measured using compact (CC-13) ion chamber. The composite planar dose was measured with IMatriXX Evaluation 2D array with OmniPro IMRT Software (version 1.7b). The per-field relative gamma was measured using electronic portal imaging device in a way similar to the routine pretreatment patient-specific quality assurance. RESULTS: Our planning results are compared with the TG-119 data. Point dose and fluence comparison data where within the acceptable confident limit. CONCLUSION: From the obtained data in this study, we conclude that the commissioning of IMRT and VMAT delivery were found within the limits of TG-119. Medknow Publications & Media Pvt Ltd 2017 /pmc/articles/PMC5744455/ /pubmed/29296041 http://dx.doi.org/10.4103/jmp.JMP_16_17 Text en Copyright: © 2017 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms. |
spellingShingle | Technical Note Kaviarasu, Karunakaran Nambi Raj, N. Arunai Hamid, Misba Giri Babu, A. Ananda Sreenivas, Lingampally Murthy, Kammari Krishna Verification of Dosimetric Commissioning Accuracy of Intensity Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Delivery using Task Group-119 Guidelines |
title | Verification of Dosimetric Commissioning Accuracy of Intensity Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Delivery using Task Group-119 Guidelines |
title_full | Verification of Dosimetric Commissioning Accuracy of Intensity Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Delivery using Task Group-119 Guidelines |
title_fullStr | Verification of Dosimetric Commissioning Accuracy of Intensity Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Delivery using Task Group-119 Guidelines |
title_full_unstemmed | Verification of Dosimetric Commissioning Accuracy of Intensity Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Delivery using Task Group-119 Guidelines |
title_short | Verification of Dosimetric Commissioning Accuracy of Intensity Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Delivery using Task Group-119 Guidelines |
title_sort | verification of dosimetric commissioning accuracy of intensity modulated radiation therapy and volumetric modulated arc therapy delivery using task group-119 guidelines |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5744455/ https://www.ncbi.nlm.nih.gov/pubmed/29296041 http://dx.doi.org/10.4103/jmp.JMP_16_17 |
work_keys_str_mv | AT kaviarasukarunakaran verificationofdosimetriccommissioningaccuracyofintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapydeliveryusingtaskgroup119guidelines AT nambirajnarunai verificationofdosimetriccommissioningaccuracyofintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapydeliveryusingtaskgroup119guidelines AT hamidmisba verificationofdosimetriccommissioningaccuracyofintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapydeliveryusingtaskgroup119guidelines AT giribabuaananda verificationofdosimetriccommissioningaccuracyofintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapydeliveryusingtaskgroup119guidelines AT sreenivaslingampally verificationofdosimetriccommissioningaccuracyofintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapydeliveryusingtaskgroup119guidelines AT murthykammarikrishna verificationofdosimetriccommissioningaccuracyofintensitymodulatedradiationtherapyandvolumetricmodulatedarctherapydeliveryusingtaskgroup119guidelines |