Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography
OBJECTIVES: This study aimed to assess the relation between dimensional changes and object location in the field of view (FOV) using cone beam computed tomography (CBCT). MATERIALS AND METHODS: A custom-made phantom was fabricated from base plate wax. To analyze the accuracy of measurements in horiz...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Tehran University of Medical Sciences
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5748456/ https://www.ncbi.nlm.nih.gov/pubmed/29296114 |
_version_ | 1783289403410481152 |
---|---|
author | Panjnoush, Mehrdad Kheirandish, Yasaman Zeini, Negar |
author_facet | Panjnoush, Mehrdad Kheirandish, Yasaman Zeini, Negar |
author_sort | Panjnoush, Mehrdad |
collection | PubMed |
description | OBJECTIVES: This study aimed to assess the relation between dimensional changes and object location in the field of view (FOV) using cone beam computed tomography (CBCT). MATERIALS AND METHODS: A custom-made phantom was fabricated from base plate wax. To analyze the accuracy of measurements in horizontal and longitudinal dimensions, aluminum squares (0.5 mm thickness, 10×10 mm dimensions) were constructed and placed in three levels (upper, middle, and lower) and five positions (central, right, left, anterior and posterior). This phantom was scanned using Asahi, Planmeca and NewTom CBCT systems. CBCT scans were measured three times by use of their corresponding software. Statistical analysis was performed using one-way ANOVA, post-hoc test and two-way ANOVA (P<0.05). RESULTS: The differences between the mean horizontal dimensions of different systems were not significant (P=0.296). However, the differences between the mean longitudinal dimensions of different systems were significant (P=0.039). The differences between the different positions and the mean horizontal and longitudinal dimensions were significant (P<0.001, and P<0.001, respectively). The differences between the mean horizontal dimensions and different levels were not significant (P=0.51), but the differences between the mean longitudinal dimensions and different levels were significant (P<0.001). The interaction effect of level and position on the accuracy of horizontal and longitudinal measurements was significant (P<0.0001). CONCLUSIONS: We found statistically significant differences in most of our comparisons; however, these differences were not clinically significant. Therefore, CBCT could be an accurate device for measurement of dimensions of objects placed in different positions in the FOV. |
format | Online Article Text |
id | pubmed-5748456 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | Tehran University of Medical Sciences |
record_format | MEDLINE/PubMed |
spelling | pubmed-57484562018-01-02 Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography Panjnoush, Mehrdad Kheirandish, Yasaman Zeini, Negar J Dent (Tehran) Original Article OBJECTIVES: This study aimed to assess the relation between dimensional changes and object location in the field of view (FOV) using cone beam computed tomography (CBCT). MATERIALS AND METHODS: A custom-made phantom was fabricated from base plate wax. To analyze the accuracy of measurements in horizontal and longitudinal dimensions, aluminum squares (0.5 mm thickness, 10×10 mm dimensions) were constructed and placed in three levels (upper, middle, and lower) and five positions (central, right, left, anterior and posterior). This phantom was scanned using Asahi, Planmeca and NewTom CBCT systems. CBCT scans were measured three times by use of their corresponding software. Statistical analysis was performed using one-way ANOVA, post-hoc test and two-way ANOVA (P<0.05). RESULTS: The differences between the mean horizontal dimensions of different systems were not significant (P=0.296). However, the differences between the mean longitudinal dimensions of different systems were significant (P=0.039). The differences between the different positions and the mean horizontal and longitudinal dimensions were significant (P<0.001, and P<0.001, respectively). The differences between the mean horizontal dimensions and different levels were not significant (P=0.51), but the differences between the mean longitudinal dimensions and different levels were significant (P<0.001). The interaction effect of level and position on the accuracy of horizontal and longitudinal measurements was significant (P<0.0001). CONCLUSIONS: We found statistically significant differences in most of our comparisons; however, these differences were not clinically significant. Therefore, CBCT could be an accurate device for measurement of dimensions of objects placed in different positions in the FOV. Tehran University of Medical Sciences 2017-09 /pmc/articles/PMC5748456/ /pubmed/29296114 Text en Copyright© Dental Research Center, Tehran University of Medical Sciences http://creativecommons.org/licenses/by/3.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Panjnoush, Mehrdad Kheirandish, Yasaman Zeini, Negar Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography |
title | Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography |
title_full | Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography |
title_fullStr | Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography |
title_full_unstemmed | Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography |
title_short | Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography |
title_sort | effect of spatial position in the field of view on dimensional changes in cone beam computed tomography |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5748456/ https://www.ncbi.nlm.nih.gov/pubmed/29296114 |
work_keys_str_mv | AT panjnoushmehrdad effectofspatialpositioninthefieldofviewondimensionalchangesinconebeamcomputedtomography AT kheirandishyasaman effectofspatialpositioninthefieldofviewondimensionalchangesinconebeamcomputedtomography AT zeininegar effectofspatialpositioninthefieldofviewondimensionalchangesinconebeamcomputedtomography |