Validation of BP devices QardioArm(®) in the general population and Omron M6 Comfort(® )in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP)

BACKGROUND: Following the European Society of Hypertension International Protocol (ESH-IP) Revision 2010, QardioArm(®) and Omron M6 Comfort IT(®) oscillometric devices were evaluated in the general population and in patients with type II diabetes, respectively, for self-blood pressure (BP) measureme...

Descripción completa

Detalles Bibliográficos
Autores principales: Chahine, Mirna N, Topouchian, Jirar, Zelveian, Parounak, Hakobyan, Zoya, Melkonyan, Arevik, Azaki, Alaa, Diab, Reem, Harb, Aya, Asmar, Roland
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5749560/
https://www.ncbi.nlm.nih.gov/pubmed/29343992
http://dx.doi.org/10.2147/MDER.S142126
_version_ 1783289604459200512
author Chahine, Mirna N
Topouchian, Jirar
Zelveian, Parounak
Hakobyan, Zoya
Melkonyan, Arevik
Azaki, Alaa
Diab, Reem
Harb, Aya
Asmar, Roland
author_facet Chahine, Mirna N
Topouchian, Jirar
Zelveian, Parounak
Hakobyan, Zoya
Melkonyan, Arevik
Azaki, Alaa
Diab, Reem
Harb, Aya
Asmar, Roland
author_sort Chahine, Mirna N
collection PubMed
description BACKGROUND: Following the European Society of Hypertension International Protocol (ESH-IP) Revision 2010, QardioArm(®) and Omron M6 Comfort IT(®) oscillometric devices were evaluated in the general population and in patients with type II diabetes, respectively, for self-blood pressure (BP) measurement. METHODS: Both devices, QardioArm(®) and Omron M6 Comfort(®), measure BP at the brachial level. The ESH-IP Revision 2010 includes a total number of 33 subjects. For each measure, the difference between observer and device BP values was calculated. In all, 99 pairs of BP differences are classified into three categories (≤5, ≤10, and ≤15 mmHg). The protocol procedures were followed precisely. RESULTS: QardioArm(®) and Omron M6 Comfort(®) fulfilled the requirements of the ESH-IP and passed the validation process successfully. For QardioArm(®), a total of 69 out of 99 comparisons for systolic blood pressure (SBP) showed an absolute difference within 5 mmHg and 82 out of 99 for diastolic blood pressure (DBP). As for Omron M6 Comfort(®), a total of 83 out of 99 comparisons for SBP showed an absolute difference within 5 mmHg and 77 out of 99 for DBP. The mean differences between the device and mercury readings were 0.7±5.9 mmHg for SBP and 0.3±4.1 mmHg for DBP for QardioArm(®) and −1.4±4.7 mmHg for SBP and −2.1±4.3 mmHg for DBP for Omron M6 Comfort(®). With regard to part 2 of ESH-IP 2010, 27 out of 33 subjects had a minimum of two out of three measurements within 5 mmHg difference for SBP and 31 out of 33 subjects for DBP for the QardioArm(®), and 29 out of 33 patients had a minimum of two out of three measurements within 5 mmHg difference for SBP and 26 out of 33 patients for DBP for Omron M6 Comfort(®). CONCLUSION: QardioArm(®) and Omron M6 Comfort(®) readings differing from the mercury standard by <5, 10, and 15 mmHg fulfill the ESH-IP Revision 2010 requirements. Consequently, these two devices are suitable for use in the general population and non-insulin-dependent type II diabetic patients, respectively.
format Online
Article
Text
id pubmed-5749560
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-57495602018-01-17 Validation of BP devices QardioArm(®) in the general population and Omron M6 Comfort(® )in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP) Chahine, Mirna N Topouchian, Jirar Zelveian, Parounak Hakobyan, Zoya Melkonyan, Arevik Azaki, Alaa Diab, Reem Harb, Aya Asmar, Roland Med Devices (Auckl) Original Research BACKGROUND: Following the European Society of Hypertension International Protocol (ESH-IP) Revision 2010, QardioArm(®) and Omron M6 Comfort IT(®) oscillometric devices were evaluated in the general population and in patients with type II diabetes, respectively, for self-blood pressure (BP) measurement. METHODS: Both devices, QardioArm(®) and Omron M6 Comfort(®), measure BP at the brachial level. The ESH-IP Revision 2010 includes a total number of 33 subjects. For each measure, the difference between observer and device BP values was calculated. In all, 99 pairs of BP differences are classified into three categories (≤5, ≤10, and ≤15 mmHg). The protocol procedures were followed precisely. RESULTS: QardioArm(®) and Omron M6 Comfort(®) fulfilled the requirements of the ESH-IP and passed the validation process successfully. For QardioArm(®), a total of 69 out of 99 comparisons for systolic blood pressure (SBP) showed an absolute difference within 5 mmHg and 82 out of 99 for diastolic blood pressure (DBP). As for Omron M6 Comfort(®), a total of 83 out of 99 comparisons for SBP showed an absolute difference within 5 mmHg and 77 out of 99 for DBP. The mean differences between the device and mercury readings were 0.7±5.9 mmHg for SBP and 0.3±4.1 mmHg for DBP for QardioArm(®) and −1.4±4.7 mmHg for SBP and −2.1±4.3 mmHg for DBP for Omron M6 Comfort(®). With regard to part 2 of ESH-IP 2010, 27 out of 33 subjects had a minimum of two out of three measurements within 5 mmHg difference for SBP and 31 out of 33 subjects for DBP for the QardioArm(®), and 29 out of 33 patients had a minimum of two out of three measurements within 5 mmHg difference for SBP and 26 out of 33 patients for DBP for Omron M6 Comfort(®). CONCLUSION: QardioArm(®) and Omron M6 Comfort(®) readings differing from the mercury standard by <5, 10, and 15 mmHg fulfill the ESH-IP Revision 2010 requirements. Consequently, these two devices are suitable for use in the general population and non-insulin-dependent type II diabetic patients, respectively. Dove Medical Press 2017-12-27 /pmc/articles/PMC5749560/ /pubmed/29343992 http://dx.doi.org/10.2147/MDER.S142126 Text en © 2018 Chahine et al. This work is published and licensed by Dove Medical Press Limited The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed.
spellingShingle Original Research
Chahine, Mirna N
Topouchian, Jirar
Zelveian, Parounak
Hakobyan, Zoya
Melkonyan, Arevik
Azaki, Alaa
Diab, Reem
Harb, Aya
Asmar, Roland
Validation of BP devices QardioArm(®) in the general population and Omron M6 Comfort(® )in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP)
title Validation of BP devices QardioArm(®) in the general population and Omron M6 Comfort(® )in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP)
title_full Validation of BP devices QardioArm(®) in the general population and Omron M6 Comfort(® )in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP)
title_fullStr Validation of BP devices QardioArm(®) in the general population and Omron M6 Comfort(® )in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP)
title_full_unstemmed Validation of BP devices QardioArm(®) in the general population and Omron M6 Comfort(® )in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP)
title_short Validation of BP devices QardioArm(®) in the general population and Omron M6 Comfort(® )in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP)
title_sort validation of bp devices qardioarm(®) in the general population and omron m6 comfort(® )in type ii diabetic patients according to the european society of hypertension international protocol (esh-ip)
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5749560/
https://www.ncbi.nlm.nih.gov/pubmed/29343992
http://dx.doi.org/10.2147/MDER.S142126
work_keys_str_mv AT chahinemirnan validationofbpdevicesqardioarminthegeneralpopulationandomronm6comfortintypeiidiabeticpatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocoleship
AT topouchianjirar validationofbpdevicesqardioarminthegeneralpopulationandomronm6comfortintypeiidiabeticpatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocoleship
AT zelveianparounak validationofbpdevicesqardioarminthegeneralpopulationandomronm6comfortintypeiidiabeticpatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocoleship
AT hakobyanzoya validationofbpdevicesqardioarminthegeneralpopulationandomronm6comfortintypeiidiabeticpatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocoleship
AT melkonyanarevik validationofbpdevicesqardioarminthegeneralpopulationandomronm6comfortintypeiidiabeticpatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocoleship
AT azakialaa validationofbpdevicesqardioarminthegeneralpopulationandomronm6comfortintypeiidiabeticpatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocoleship
AT diabreem validationofbpdevicesqardioarminthegeneralpopulationandomronm6comfortintypeiidiabeticpatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocoleship
AT harbaya validationofbpdevicesqardioarminthegeneralpopulationandomronm6comfortintypeiidiabeticpatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocoleship
AT asmarroland validationofbpdevicesqardioarminthegeneralpopulationandomronm6comfortintypeiidiabeticpatientsaccordingtotheeuropeansocietyofhypertensioninternationalprotocoleship