Cargando…

Knee Implant Loosening Detection: A Vibration Analysis Investigation

Knee implant loosening is mainly caused by the weakness of the prosthesis-bone interface and is the main reason for surgical revisions. However, pre-operative diagnosis is difficult due to lack of accurate tests. In this study, we developed a vibration-based system to detect the loosening of the tib...

Descripción completa

Detalles Bibliográficos
Autores principales: Arami, Arash, Delaloye, Jean-Romain, Rouhani, Hossein, Jolles, Brigitte M., Aminian, Kamiar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer US 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5754435/
https://www.ncbi.nlm.nih.gov/pubmed/29067562
http://dx.doi.org/10.1007/s10439-017-1941-2
_version_ 1783290416123084800
author Arami, Arash
Delaloye, Jean-Romain
Rouhani, Hossein
Jolles, Brigitte M.
Aminian, Kamiar
author_facet Arami, Arash
Delaloye, Jean-Romain
Rouhani, Hossein
Jolles, Brigitte M.
Aminian, Kamiar
author_sort Arami, Arash
collection PubMed
description Knee implant loosening is mainly caused by the weakness of the prosthesis-bone interface and is the main reason for surgical revisions. However, pre-operative diagnosis is difficult due to lack of accurate tests. In this study, we developed a vibration-based system to detect the loosening of the tibial implant of an instrumented knee prosthesis. The proposed system includes an instrumented vibrator for transcutaneous stimulation of the bone in a repeatable manner, and accelerometer sensors integrated into the implants to measure the propagated vibration. A coherence-based detection technique was proposed to distinguish the loosened implants from the secure ones. Fourteen ex vivo lower limbs were used, on which the knee prosthesis was implanted, and harmonic-forced vibration was applied on the tibia. The input–output coherence measure provided 92.26% accuracy, a high sensitivity (91.67%) and specificity (92.86%). This technique was benchmarked against power spectrum based analysis of the propagated vibration to the implant. In particular, loosening detection based on new peak appearance, peak shift, and peak flattening in power spectra showed inferior performance to the proposed coherence-based technique. As such, application of vibration on our instrumented knee prosthesis together with input–output coherence analysis enabled us to distinguish the secure from loose implants.
format Online
Article
Text
id pubmed-5754435
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher Springer US
record_format MEDLINE/PubMed
spelling pubmed-57544352018-01-22 Knee Implant Loosening Detection: A Vibration Analysis Investigation Arami, Arash Delaloye, Jean-Romain Rouhani, Hossein Jolles, Brigitte M. Aminian, Kamiar Ann Biomed Eng Article Knee implant loosening is mainly caused by the weakness of the prosthesis-bone interface and is the main reason for surgical revisions. However, pre-operative diagnosis is difficult due to lack of accurate tests. In this study, we developed a vibration-based system to detect the loosening of the tibial implant of an instrumented knee prosthesis. The proposed system includes an instrumented vibrator for transcutaneous stimulation of the bone in a repeatable manner, and accelerometer sensors integrated into the implants to measure the propagated vibration. A coherence-based detection technique was proposed to distinguish the loosened implants from the secure ones. Fourteen ex vivo lower limbs were used, on which the knee prosthesis was implanted, and harmonic-forced vibration was applied on the tibia. The input–output coherence measure provided 92.26% accuracy, a high sensitivity (91.67%) and specificity (92.86%). This technique was benchmarked against power spectrum based analysis of the propagated vibration to the implant. In particular, loosening detection based on new peak appearance, peak shift, and peak flattening in power spectra showed inferior performance to the proposed coherence-based technique. As such, application of vibration on our instrumented knee prosthesis together with input–output coherence analysis enabled us to distinguish the secure from loose implants. Springer US 2017-10-24 2018 /pmc/articles/PMC5754435/ /pubmed/29067562 http://dx.doi.org/10.1007/s10439-017-1941-2 Text en © The Author(s) 2017 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Article
Arami, Arash
Delaloye, Jean-Romain
Rouhani, Hossein
Jolles, Brigitte M.
Aminian, Kamiar
Knee Implant Loosening Detection: A Vibration Analysis Investigation
title Knee Implant Loosening Detection: A Vibration Analysis Investigation
title_full Knee Implant Loosening Detection: A Vibration Analysis Investigation
title_fullStr Knee Implant Loosening Detection: A Vibration Analysis Investigation
title_full_unstemmed Knee Implant Loosening Detection: A Vibration Analysis Investigation
title_short Knee Implant Loosening Detection: A Vibration Analysis Investigation
title_sort knee implant loosening detection: a vibration analysis investigation
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5754435/
https://www.ncbi.nlm.nih.gov/pubmed/29067562
http://dx.doi.org/10.1007/s10439-017-1941-2
work_keys_str_mv AT aramiarash kneeimplantlooseningdetectionavibrationanalysisinvestigation
AT delaloyejeanromain kneeimplantlooseningdetectionavibrationanalysisinvestigation
AT rouhanihossein kneeimplantlooseningdetectionavibrationanalysisinvestigation
AT jollesbrigittem kneeimplantlooseningdetectionavibrationanalysisinvestigation
AT aminiankamiar kneeimplantlooseningdetectionavibrationanalysisinvestigation