The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study.
Low-flow anaesthesia is considered beneficial for the patient and the environment, and it is cost reducing due to reduced anaesthetic gas consumption. An initial high-flow to saturate the circle system ( wash-in) is desirable from a clinical point of view. We measured the wash-in and wash-out times...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
F1000 Research Limited
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5754744/ https://www.ncbi.nlm.nih.gov/pubmed/29333245 http://dx.doi.org/10.12688/f1000research.13064.2 |
_version_ | 1783290483019087872 |
---|---|
author | Leijonhufvud, Fredrik Jöneby, Fredrik Jakobsson, Jan G. |
author_facet | Leijonhufvud, Fredrik Jöneby, Fredrik Jakobsson, Jan G. |
author_sort | Leijonhufvud, Fredrik |
collection | PubMed |
description | Low-flow anaesthesia is considered beneficial for the patient and the environment, and it is cost reducing due to reduced anaesthetic gas consumption. An initial high-flow to saturate the circle system ( wash-in) is desirable from a clinical point of view. We measured the wash-in and wash-out times (time to saturate and to eliminate the anaesthetic agent, AA), for sevoflurane and desflurane, in a test-lung with fixed 3 MAC vaporizer setting at different fresh gas flow (FGF) and calculated the consumption of AA. We tried to find an optimal flow rate for speed and gas consumption, comparing two anaesthesia machines (AMs): Aisys and Flow-i. Time to reach 1 minimal alveolar concentration (MAC) (wash-in) decreased (p<0.05) at higher flow rates (1 – 2 – 4) but plateaued at 4-4.8 l/min. The consumption of AA was at its lowest around 4-4.8 l/min (optimal flow) for all but the Aisys /desflurane group. Wash-out times decreased as FGF increased, until reaching plateau at FGF of 4-6 l/min. Aisys had generally shorter wash-in times at flow rates < 4 l/min as well as lower consumption of AA. At higher flow rates there were little difference between the AMs. The “optimal FGF” for wash-out, elimination of gas from the test-lung and circle system, plateaued with no increase in speed beyond 6 l/min. A fresh gas flow of 4 l/min. seems “optimal” taking speed to reach a 1 MAC ET and gas consumption into account during wash-in with a fixed 3 MAC vaporizer setting, and increasing fresh gas flow beyond 6 l/min does not seem to confirm major benefit during wash-out. |
format | Online Article Text |
id | pubmed-5754744 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | F1000 Research Limited |
record_format | MEDLINE/PubMed |
spelling | pubmed-57547442018-01-11 The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study. Leijonhufvud, Fredrik Jöneby, Fredrik Jakobsson, Jan G. F1000Res Research Note Low-flow anaesthesia is considered beneficial for the patient and the environment, and it is cost reducing due to reduced anaesthetic gas consumption. An initial high-flow to saturate the circle system ( wash-in) is desirable from a clinical point of view. We measured the wash-in and wash-out times (time to saturate and to eliminate the anaesthetic agent, AA), for sevoflurane and desflurane, in a test-lung with fixed 3 MAC vaporizer setting at different fresh gas flow (FGF) and calculated the consumption of AA. We tried to find an optimal flow rate for speed and gas consumption, comparing two anaesthesia machines (AMs): Aisys and Flow-i. Time to reach 1 minimal alveolar concentration (MAC) (wash-in) decreased (p<0.05) at higher flow rates (1 – 2 – 4) but plateaued at 4-4.8 l/min. The consumption of AA was at its lowest around 4-4.8 l/min (optimal flow) for all but the Aisys /desflurane group. Wash-out times decreased as FGF increased, until reaching plateau at FGF of 4-6 l/min. Aisys had generally shorter wash-in times at flow rates < 4 l/min as well as lower consumption of AA. At higher flow rates there were little difference between the AMs. The “optimal FGF” for wash-out, elimination of gas from the test-lung and circle system, plateaued with no increase in speed beyond 6 l/min. A fresh gas flow of 4 l/min. seems “optimal” taking speed to reach a 1 MAC ET and gas consumption into account during wash-in with a fixed 3 MAC vaporizer setting, and increasing fresh gas flow beyond 6 l/min does not seem to confirm major benefit during wash-out. F1000 Research Limited 2017-12-22 /pmc/articles/PMC5754744/ /pubmed/29333245 http://dx.doi.org/10.12688/f1000research.13064.2 Text en Copyright: © 2017 Leijonhufvud F et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Note Leijonhufvud, Fredrik Jöneby, Fredrik Jakobsson, Jan G. The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study. |
title | The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study. |
title_full | The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study. |
title_fullStr | The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study. |
title_full_unstemmed | The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study. |
title_short | The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study. |
title_sort | impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study. |
topic | Research Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5754744/ https://www.ncbi.nlm.nih.gov/pubmed/29333245 http://dx.doi.org/10.12688/f1000research.13064.2 |
work_keys_str_mv | AT leijonhufvudfredrik theimpactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy AT jonebyfredrik theimpactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy AT jakobssonjang theimpactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy AT leijonhufvudfredrik impactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy AT jonebyfredrik impactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy AT jakobssonjang impactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy |