The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study.

Low-flow anaesthesia is considered beneficial for the patient and the environment, and it is cost reducing due to reduced anaesthetic gas consumption. An initial high-flow to saturate the circle system ( wash-in) is desirable from a clinical point of view. We measured the wash-in and wash-out times...

Descripción completa

Detalles Bibliográficos
Autores principales: Leijonhufvud, Fredrik, Jöneby, Fredrik, Jakobsson, Jan G.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: F1000 Research Limited 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5754744/
https://www.ncbi.nlm.nih.gov/pubmed/29333245
http://dx.doi.org/10.12688/f1000research.13064.2
_version_ 1783290483019087872
author Leijonhufvud, Fredrik
Jöneby, Fredrik
Jakobsson, Jan G.
author_facet Leijonhufvud, Fredrik
Jöneby, Fredrik
Jakobsson, Jan G.
author_sort Leijonhufvud, Fredrik
collection PubMed
description Low-flow anaesthesia is considered beneficial for the patient and the environment, and it is cost reducing due to reduced anaesthetic gas consumption. An initial high-flow to saturate the circle system ( wash-in) is desirable from a clinical point of view. We measured the wash-in and wash-out times (time to saturate and to eliminate the anaesthetic agent, AA), for sevoflurane and desflurane, in a test-lung with fixed 3 MAC vaporizer setting at different fresh gas flow (FGF) and calculated the consumption of AA. We tried to find an optimal flow rate for speed and gas consumption, comparing two anaesthesia machines (AMs): Aisys and Flow-i. Time to reach 1 minimal alveolar concentration (MAC) (wash-in) decreased (p<0.05) at higher flow rates (1 – 2 – 4) but plateaued at 4-4.8 l/min. The consumption of AA was at its lowest around 4-4.8 l/min (optimal flow) for all but the Aisys /desflurane group. Wash-out times decreased as FGF increased, until reaching plateau at FGF of 4-6 l/min. Aisys had generally shorter wash-in times at flow rates < 4 l/min as well as lower consumption of AA. At higher flow rates there were little difference between the AMs. The “optimal FGF” for wash-out, elimination of gas from the test-lung and circle system, plateaued with no increase in speed beyond 6 l/min. A fresh gas flow of 4 l/min. seems “optimal” taking speed to reach a 1 MAC ET and gas consumption into account during wash-in with a fixed 3 MAC vaporizer setting, and increasing fresh gas flow beyond 6 l/min does not seem to confirm major benefit during wash-out.
format Online
Article
Text
id pubmed-5754744
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher F1000 Research Limited
record_format MEDLINE/PubMed
spelling pubmed-57547442018-01-11 The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study. Leijonhufvud, Fredrik Jöneby, Fredrik Jakobsson, Jan G. F1000Res Research Note Low-flow anaesthesia is considered beneficial for the patient and the environment, and it is cost reducing due to reduced anaesthetic gas consumption. An initial high-flow to saturate the circle system ( wash-in) is desirable from a clinical point of view. We measured the wash-in and wash-out times (time to saturate and to eliminate the anaesthetic agent, AA), for sevoflurane and desflurane, in a test-lung with fixed 3 MAC vaporizer setting at different fresh gas flow (FGF) and calculated the consumption of AA. We tried to find an optimal flow rate for speed and gas consumption, comparing two anaesthesia machines (AMs): Aisys and Flow-i. Time to reach 1 minimal alveolar concentration (MAC) (wash-in) decreased (p<0.05) at higher flow rates (1 – 2 – 4) but plateaued at 4-4.8 l/min. The consumption of AA was at its lowest around 4-4.8 l/min (optimal flow) for all but the Aisys /desflurane group. Wash-out times decreased as FGF increased, until reaching plateau at FGF of 4-6 l/min. Aisys had generally shorter wash-in times at flow rates < 4 l/min as well as lower consumption of AA. At higher flow rates there were little difference between the AMs. The “optimal FGF” for wash-out, elimination of gas from the test-lung and circle system, plateaued with no increase in speed beyond 6 l/min. A fresh gas flow of 4 l/min. seems “optimal” taking speed to reach a 1 MAC ET and gas consumption into account during wash-in with a fixed 3 MAC vaporizer setting, and increasing fresh gas flow beyond 6 l/min does not seem to confirm major benefit during wash-out. F1000 Research Limited 2017-12-22 /pmc/articles/PMC5754744/ /pubmed/29333245 http://dx.doi.org/10.12688/f1000research.13064.2 Text en Copyright: © 2017 Leijonhufvud F et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Note
Leijonhufvud, Fredrik
Jöneby, Fredrik
Jakobsson, Jan G.
The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study.
title The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study.
title_full The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study.
title_fullStr The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study.
title_full_unstemmed The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study.
title_short The impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study.
title_sort impact of fresh gas flow on wash-in, wash-out time and gas consumption for sevoflurane and desflurane, comparing two anaesthesia machines, a test-lung study.
topic Research Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5754744/
https://www.ncbi.nlm.nih.gov/pubmed/29333245
http://dx.doi.org/10.12688/f1000research.13064.2
work_keys_str_mv AT leijonhufvudfredrik theimpactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy
AT jonebyfredrik theimpactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy
AT jakobssonjang theimpactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy
AT leijonhufvudfredrik impactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy
AT jonebyfredrik impactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy
AT jakobssonjang impactoffreshgasflowonwashinwashouttimeandgasconsumptionforsevofluraneanddesfluranecomparingtwoanaesthesiamachinesatestlungstudy