Effects of temperature and wavelength choice on in-situ dissolution test of Cimetidine tablets

The effects of temperature and wavelength choice on in-situ dissolution test instrument of Cimetidine were studied. Absorbance (A)<1.0 is required when using a fiber-optic dissolution test system. The detection wavelength of λ(max) (218 nm) was replaced by 244 nm to carry out this test. The absor...

Descripción completa

Detalles Bibliográficos
Autores principales: Li, Xin-Xia, Wang, Yan, Xu, Ping-Ping, Zhang, Qi-Zhou, Nie, Kun, Hu, Xu, Kong, Bin, Li, Li, Chen, Jian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Xi'an Jiaotong University 2013
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5760944/
https://www.ncbi.nlm.nih.gov/pubmed/29403799
http://dx.doi.org/10.1016/j.jpha.2012.06.001
Descripción
Sumario:The effects of temperature and wavelength choice on in-situ dissolution test instrument of Cimetidine were studied. Absorbance (A)<1.0 is required when using a fiber-optic dissolution test system. The detection wavelength of λ(max) (218 nm) was replaced by 244 nm to carry out this test. The absorbance of Cimetidine solution at different temperature showed an obvious change. Calibration of Cimetidine solution should be tested at the same temperature (37° C) with the test solution. A suitable wavelength with smaller tangent slope could be chosen for in-situ dissolution test of Cimetidine tablets.