In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device
PURPOSE: We recently developed a novel, open-source in vivo dosimetry that uses the electronic portal imaging device to detect dose delivery discrepancies. We applied our method on patients with rectal cancer treated on a belly board device. METHODS: In vivo dosimetry was performed on 10 patients wi...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5762054/ https://www.ncbi.nlm.nih.gov/pubmed/28585490 http://dx.doi.org/10.1177/1533034617711519 |
_version_ | 1783291625874653184 |
---|---|
author | Peca, Stefano Sinha, Richie Siddhartha Brown, Derek Wilson Smith, Wendy Lani |
author_facet | Peca, Stefano Sinha, Richie Siddhartha Brown, Derek Wilson Smith, Wendy Lani |
author_sort | Peca, Stefano |
collection | PubMed |
description | PURPOSE: We recently developed a novel, open-source in vivo dosimetry that uses the electronic portal imaging device to detect dose delivery discrepancies. We applied our method on patients with rectal cancer treated on a belly board device. METHODS: In vivo dosimetry was performed on 10 patients with rectal cancer treated prone on the belly board with a 4-field box arrangement. Portal images were acquired approximately once per week from each treatment beam. Our dosimetry method used these images along with the planning CT to reconstruct patient planar dose at isocenter depth. RESULTS: Our algorithm proved sensitive to dose discrepancies and detected discordances in 7 patients. The majority of these were due to soft tissue differences between planning and treatment, present despite matching to bony anatomy. As a result of this work, quality assurance procedures have been implemented for our immobilization devices. CONCLUSION: In vivo dosimetry is a powerful quality assurance tool that can detect delivery discrepancies, including changes in patient setup and position. The added information on actual dose delivery may be used to evaluate equipment and process quality and to guide for adaptive radiotherapy. |
format | Online Article Text |
id | pubmed-5762054 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-57620542018-01-17 In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device Peca, Stefano Sinha, Richie Siddhartha Brown, Derek Wilson Smith, Wendy Lani Technol Cancer Res Treat Original Articles PURPOSE: We recently developed a novel, open-source in vivo dosimetry that uses the electronic portal imaging device to detect dose delivery discrepancies. We applied our method on patients with rectal cancer treated on a belly board device. METHODS: In vivo dosimetry was performed on 10 patients with rectal cancer treated prone on the belly board with a 4-field box arrangement. Portal images were acquired approximately once per week from each treatment beam. Our dosimetry method used these images along with the planning CT to reconstruct patient planar dose at isocenter depth. RESULTS: Our algorithm proved sensitive to dose discrepancies and detected discordances in 7 patients. The majority of these were due to soft tissue differences between planning and treatment, present despite matching to bony anatomy. As a result of this work, quality assurance procedures have been implemented for our immobilization devices. CONCLUSION: In vivo dosimetry is a powerful quality assurance tool that can detect delivery discrepancies, including changes in patient setup and position. The added information on actual dose delivery may be used to evaluate equipment and process quality and to guide for adaptive radiotherapy. SAGE Publications 2017-06-06 2017-12 /pmc/articles/PMC5762054/ /pubmed/28585490 http://dx.doi.org/10.1177/1533034617711519 Text en © The Author(s) 2017 http://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Articles Peca, Stefano Sinha, Richie Siddhartha Brown, Derek Wilson Smith, Wendy Lani In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device |
title |
In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device |
title_full |
In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device |
title_fullStr |
In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device |
title_full_unstemmed |
In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device |
title_short |
In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device |
title_sort | in vivo portal imaging dosimetry identifies delivery errors in rectal cancer radiotherapy on the belly board device |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5762054/ https://www.ncbi.nlm.nih.gov/pubmed/28585490 http://dx.doi.org/10.1177/1533034617711519 |
work_keys_str_mv | AT pecastefano invivoportalimagingdosimetryidentifiesdeliveryerrorsinrectalcancerradiotherapyonthebellyboarddevice AT sinharichiesiddhartha invivoportalimagingdosimetryidentifiesdeliveryerrorsinrectalcancerradiotherapyonthebellyboarddevice AT brownderekwilson invivoportalimagingdosimetryidentifiesdeliveryerrorsinrectalcancerradiotherapyonthebellyboarddevice AT smithwendylani invivoportalimagingdosimetryidentifiesdeliveryerrorsinrectalcancerradiotherapyonthebellyboarddevice |