In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device

PURPOSE: We recently developed a novel, open-source in vivo dosimetry that uses the electronic portal imaging device to detect dose delivery discrepancies. We applied our method on patients with rectal cancer treated on a belly board device. METHODS: In vivo dosimetry was performed on 10 patients wi...

Descripción completa

Detalles Bibliográficos
Autores principales: Peca, Stefano, Sinha, Richie Siddhartha, Brown, Derek Wilson, Smith, Wendy Lani
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5762054/
https://www.ncbi.nlm.nih.gov/pubmed/28585490
http://dx.doi.org/10.1177/1533034617711519
_version_ 1783291625874653184
author Peca, Stefano
Sinha, Richie Siddhartha
Brown, Derek Wilson
Smith, Wendy Lani
author_facet Peca, Stefano
Sinha, Richie Siddhartha
Brown, Derek Wilson
Smith, Wendy Lani
author_sort Peca, Stefano
collection PubMed
description PURPOSE: We recently developed a novel, open-source in vivo dosimetry that uses the electronic portal imaging device to detect dose delivery discrepancies. We applied our method on patients with rectal cancer treated on a belly board device. METHODS: In vivo dosimetry was performed on 10 patients with rectal cancer treated prone on the belly board with a 4-field box arrangement. Portal images were acquired approximately once per week from each treatment beam. Our dosimetry method used these images along with the planning CT to reconstruct patient planar dose at isocenter depth. RESULTS: Our algorithm proved sensitive to dose discrepancies and detected discordances in 7 patients. The majority of these were due to soft tissue differences between planning and treatment, present despite matching to bony anatomy. As a result of this work, quality assurance procedures have been implemented for our immobilization devices. CONCLUSION: In vivo dosimetry is a powerful quality assurance tool that can detect delivery discrepancies, including changes in patient setup and position. The added information on actual dose delivery may be used to evaluate equipment and process quality and to guide for adaptive radiotherapy.
format Online
Article
Text
id pubmed-5762054
institution National Center for Biotechnology Information
language English
publishDate 2017
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-57620542018-01-17 In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device Peca, Stefano Sinha, Richie Siddhartha Brown, Derek Wilson Smith, Wendy Lani Technol Cancer Res Treat Original Articles PURPOSE: We recently developed a novel, open-source in vivo dosimetry that uses the electronic portal imaging device to detect dose delivery discrepancies. We applied our method on patients with rectal cancer treated on a belly board device. METHODS: In vivo dosimetry was performed on 10 patients with rectal cancer treated prone on the belly board with a 4-field box arrangement. Portal images were acquired approximately once per week from each treatment beam. Our dosimetry method used these images along with the planning CT to reconstruct patient planar dose at isocenter depth. RESULTS: Our algorithm proved sensitive to dose discrepancies and detected discordances in 7 patients. The majority of these were due to soft tissue differences between planning and treatment, present despite matching to bony anatomy. As a result of this work, quality assurance procedures have been implemented for our immobilization devices. CONCLUSION: In vivo dosimetry is a powerful quality assurance tool that can detect delivery discrepancies, including changes in patient setup and position. The added information on actual dose delivery may be used to evaluate equipment and process quality and to guide for adaptive radiotherapy. SAGE Publications 2017-06-06 2017-12 /pmc/articles/PMC5762054/ /pubmed/28585490 http://dx.doi.org/10.1177/1533034617711519 Text en © The Author(s) 2017 http://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Articles
Peca, Stefano
Sinha, Richie Siddhartha
Brown, Derek Wilson
Smith, Wendy Lani
In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device
title In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device
title_full In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device
title_fullStr In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device
title_full_unstemmed In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device
title_short In vivo Portal Imaging Dosimetry Identifies Delivery Errors in Rectal Cancer Radiotherapy on the Belly Board Device
title_sort in vivo portal imaging dosimetry identifies delivery errors in rectal cancer radiotherapy on the belly board device
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5762054/
https://www.ncbi.nlm.nih.gov/pubmed/28585490
http://dx.doi.org/10.1177/1533034617711519
work_keys_str_mv AT pecastefano invivoportalimagingdosimetryidentifiesdeliveryerrorsinrectalcancerradiotherapyonthebellyboarddevice
AT sinharichiesiddhartha invivoportalimagingdosimetryidentifiesdeliveryerrorsinrectalcancerradiotherapyonthebellyboarddevice
AT brownderekwilson invivoportalimagingdosimetryidentifiesdeliveryerrorsinrectalcancerradiotherapyonthebellyboarddevice
AT smithwendylani invivoportalimagingdosimetryidentifiesdeliveryerrorsinrectalcancerradiotherapyonthebellyboarddevice