Quantifying the effect of air gap, depth, and range shifter thickness on TPS dosimetric accuracy in superficial PBS proton therapy
This study quantifies the dosimetric accuracy of a commercial treatment planning system as functions of treatment depth, air gap, and range shifter thickness for superficial pencil beam scanning proton therapy treatments. The RayStation 6 pencil beam and Monte Carlo dose engines were each used to ca...
Autores principales: | Shirey, Robert J., Wu, Hsinshun Terry |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2017
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5768007/ https://www.ncbi.nlm.nih.gov/pubmed/29239528 http://dx.doi.org/10.1002/acm2.12241 |
Ejemplares similares
-
A Model for Secondary Monitor Unit Calculations of PBS Proton Therapy Treatment Plans
por: Schimke, Greg, et al.
Publicado: (2019) -
Morphological, Structural, Thermal, Permeability, and Antimicrobial Activity of PBS and PBS/TPS Films Incorporated with Biomaster-Silver for Food Packaging Application
por: Aziman, Nurain, et al.
Publicado: (2021) -
Evaluation of the range shifter model for proton pencil‐beam scanning for the Eclipse v.11 treatment planning system
por: Matysiak, Witold, et al.
Publicado: (2016) -
On the sensitivity of TG‐119 and IROC credentialing to TPS commissioning errors
por: McVicker, Drew, et al.
Publicado: (2016) -
Evaluation of six TPS algorithms in computing entrance and exit doses
por: Tan, Yun I., et al.
Publicado: (2014)