The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria

BACKGROUND: Malaria rapid diagnostic tests (RDTs) are becoming widely adopted for case management at community level. However, reports and anecdotal observations indicate that the blood transfer step poses a significant challenge to many users. This study sought to evaluate the inverted cup device i...

Descripción completa

Detalles Bibliográficos
Autores principales: Incardona, Sandra, Mwancha-Kwasa, Magoma, Rees-Channer, Roxanne R., Albertini, Audrey, Havumaki, Joshua, Chiodini, Peter, Oyibo, Wellington, Gonzalez, Iveth J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5769227/
https://www.ncbi.nlm.nih.gov/pubmed/29334955
http://dx.doi.org/10.1186/s12936-018-2173-0
_version_ 1783292859752906752
author Incardona, Sandra
Mwancha-Kwasa, Magoma
Rees-Channer, Roxanne R.
Albertini, Audrey
Havumaki, Joshua
Chiodini, Peter
Oyibo, Wellington
Gonzalez, Iveth J.
author_facet Incardona, Sandra
Mwancha-Kwasa, Magoma
Rees-Channer, Roxanne R.
Albertini, Audrey
Havumaki, Joshua
Chiodini, Peter
Oyibo, Wellington
Gonzalez, Iveth J.
author_sort Incardona, Sandra
collection PubMed
description BACKGROUND: Malaria rapid diagnostic tests (RDTs) are becoming widely adopted for case management at community level. However, reports and anecdotal observations indicate that the blood transfer step poses a significant challenge to many users. This study sought to evaluate the inverted cup device in the hands of health workers in everyday clinical practice, in comparison with the plastic pipette, and to determine the volume accuracy of the device made of a lower-cost plastic. METHODS: The volume accuracy of inverted cup devices made of two plastics, PMMA and SBC, was compared by transferring blood 150 times onto filter paper and comparing the blood spot areas with those produced by 20 reference transfers with a calibrated micropipette. The ease of use, safety and acceptability of the inverted cup device and the pipette were evaluated by 50 health workers in Nigeria. Observations were recorded on pre-designed questionnaires, by the health workers themselves and by trained observers. Focus group discussions were also conducted. RESULTS: The volume accuracy assessment showed that the device made from the low-cost material (SBC) delivered a more accurate volume (mean 5.4 μL, SD 0.48 μL, range 4.5–7.0 μL) than the PMMA device (mean 5.9 μL, SD 0.48 μL, range 4.9–7.2 μL). The observational evaluation demonstrated that the inverted cup device performed better than the pipette in all aspects, e.g. higher proportions of health workers achieved successful blood collection (96%, vs. 66%), transfer of the required blood volume (90%, vs. 58%), and blood deposit without any loss (95%, vs. 50%). Majority of health workers also considered it’ very easy’ to use (81%),’very appropriate’ for everyday use (78%), and 50% of them reported that it was their preferred BTD. CONCLUSIONS: The good volume accuracy and high acceptability of the inverted cup device shown in this study, along with observed ease of use and safety in hands of health workers, further strengthens prior findings which demonstrated its higher accuracy as compared with other BTDs in a laboratory setting. Altogether, these studies suggest that the inverted cup device should replace other types of devices for use in day-to-day malaria diagnosis with RDTs.
format Online
Article
Text
id pubmed-5769227
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-57692272018-01-25 The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria Incardona, Sandra Mwancha-Kwasa, Magoma Rees-Channer, Roxanne R. Albertini, Audrey Havumaki, Joshua Chiodini, Peter Oyibo, Wellington Gonzalez, Iveth J. Malar J Research BACKGROUND: Malaria rapid diagnostic tests (RDTs) are becoming widely adopted for case management at community level. However, reports and anecdotal observations indicate that the blood transfer step poses a significant challenge to many users. This study sought to evaluate the inverted cup device in the hands of health workers in everyday clinical practice, in comparison with the plastic pipette, and to determine the volume accuracy of the device made of a lower-cost plastic. METHODS: The volume accuracy of inverted cup devices made of two plastics, PMMA and SBC, was compared by transferring blood 150 times onto filter paper and comparing the blood spot areas with those produced by 20 reference transfers with a calibrated micropipette. The ease of use, safety and acceptability of the inverted cup device and the pipette were evaluated by 50 health workers in Nigeria. Observations were recorded on pre-designed questionnaires, by the health workers themselves and by trained observers. Focus group discussions were also conducted. RESULTS: The volume accuracy assessment showed that the device made from the low-cost material (SBC) delivered a more accurate volume (mean 5.4 μL, SD 0.48 μL, range 4.5–7.0 μL) than the PMMA device (mean 5.9 μL, SD 0.48 μL, range 4.9–7.2 μL). The observational evaluation demonstrated that the inverted cup device performed better than the pipette in all aspects, e.g. higher proportions of health workers achieved successful blood collection (96%, vs. 66%), transfer of the required blood volume (90%, vs. 58%), and blood deposit without any loss (95%, vs. 50%). Majority of health workers also considered it’ very easy’ to use (81%),’very appropriate’ for everyday use (78%), and 50% of them reported that it was their preferred BTD. CONCLUSIONS: The good volume accuracy and high acceptability of the inverted cup device shown in this study, along with observed ease of use and safety in hands of health workers, further strengthens prior findings which demonstrated its higher accuracy as compared with other BTDs in a laboratory setting. Altogether, these studies suggest that the inverted cup device should replace other types of devices for use in day-to-day malaria diagnosis with RDTs. BioMed Central 2018-01-15 /pmc/articles/PMC5769227/ /pubmed/29334955 http://dx.doi.org/10.1186/s12936-018-2173-0 Text en © The Author(s) 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Incardona, Sandra
Mwancha-Kwasa, Magoma
Rees-Channer, Roxanne R.
Albertini, Audrey
Havumaki, Joshua
Chiodini, Peter
Oyibo, Wellington
Gonzalez, Iveth J.
The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria
title The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria
title_full The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria
title_fullStr The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria
title_full_unstemmed The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria
title_short The inverted cup device for blood transfer on malaria RDTs: ease of use, acceptability and safety in routine use by health workers in Nigeria
title_sort inverted cup device for blood transfer on malaria rdts: ease of use, acceptability and safety in routine use by health workers in nigeria
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5769227/
https://www.ncbi.nlm.nih.gov/pubmed/29334955
http://dx.doi.org/10.1186/s12936-018-2173-0
work_keys_str_mv AT incardonasandra theinvertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT mwanchakwasamagoma theinvertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT reeschannerroxanner theinvertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT albertiniaudrey theinvertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT havumakijoshua theinvertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT chiodinipeter theinvertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT oyibowellington theinvertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT gonzalezivethj theinvertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT incardonasandra invertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT mwanchakwasamagoma invertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT reeschannerroxanner invertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT albertiniaudrey invertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT havumakijoshua invertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT chiodinipeter invertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT oyibowellington invertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria
AT gonzalezivethj invertedcupdeviceforbloodtransferonmalariardtseaseofuseacceptabilityandsafetyinroutineusebyhealthworkersinnigeria