Using Auditory Steady-State Responses for Measuring Hearing Protector Occlusion Effect

INTRODUCTION: The currently available methods for measuring the occlusion effect (OE) of hearing protection devices (HPDs) have limitations. Objective microphonic measurements do not assess bone-conducted sounds directly transmitted to the cochlea. Psychophysical measurements at threshold are biased...

Descripción completa

Detalles Bibliográficos
Autores principales: Valentin, Olivier, Laville, Frédéric
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5771060/
https://www.ncbi.nlm.nih.gov/pubmed/29319012
http://dx.doi.org/10.4103/nah.NAH_13_17
Descripción
Sumario:INTRODUCTION: The currently available methods for measuring the occlusion effect (OE) of hearing protection devices (HPDs) have limitations. Objective microphonic measurements do not assess bone-conducted sounds directly transmitted to the cochlea. Psychophysical measurements at threshold are biased due to the low-frequency masking effects from test participants’ physiological noise and the variability of measurements based on subjective responses. An auditory steady-state responses (ASSRs) procedure is used as a technique that might overcome these limitations. PARTICIPANTS AND METHODS: Pure-tone stimuli (250 and 500 Hz), with amplitude modulated at 40 Hz, were presented to twelve adults with normal hearing through a bone vibrator at three levels in 10-dB steps. The following two conditions were assessed: the unoccluded ear canal and occluded ear canal. ASSR amplitude data as a function of the stimulation level were linearized using least-square regressions. The ASSR-based “physiological” OE was then calculated as the average difference between the two measurements. RESULTS: A significant statistical difference was found between the average threshold-based psychophysical OE and the average ASSR-based OE. CONCLUSION: This study successfully ascertained that it is possible to objectively measure the OE of HPD using ASSRs collected on the same participant both with and without protectors.