Skin Temperature Measurement Using Contact Thermometry: A Systematic Review of Setup Variables and Their Effects on Measured Values

Background: Skin temperature (T(skin)) is commonly measured using T(skin) sensors affixed directly to the skin surface, although the influence of setup variables on the measured outcome requires clarification. Objectives: The two distinct objectives of this systematic review were (1) to examine meas...

Descripción completa

Detalles Bibliográficos
Autores principales: MacRae, Braid A., Annaheim, Simon, Spengler, Christina M., Rossi, René M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5797625/
https://www.ncbi.nlm.nih.gov/pubmed/29441024
http://dx.doi.org/10.3389/fphys.2018.00029
_version_ 1783297719383621632
author MacRae, Braid A.
Annaheim, Simon
Spengler, Christina M.
Rossi, René M.
author_facet MacRae, Braid A.
Annaheim, Simon
Spengler, Christina M.
Rossi, René M.
author_sort MacRae, Braid A.
collection PubMed
description Background: Skin temperature (T(skin)) is commonly measured using T(skin) sensors affixed directly to the skin surface, although the influence of setup variables on the measured outcome requires clarification. Objectives: The two distinct objectives of this systematic review were (1) to examine measurements from contact T(skin) sensors considering equilibrium temperature and temperature disturbance, sensor attachments, pressure, environmental temperature, and sensor type, and (2) to characterise the contact T(skin) sensors used, conditions of use, and subsequent reporting in studies investigating sports, exercise, and other physical activity. Data sources and study selection: For the measurement comparison objective, Ovid Medline and Scopus were used (1960 to July 2016) and studies comparing contact T(skin) sensor measurements in vivo or using appropriate physical models were included. For the survey of use, Ovid Medline was used (2011 to July 2016) and studies using contact temperature sensors for the measurement of human T(skin) in vivo during sport, exercise, and other physical activity were included. Study appraisal and synthesis methods: For measurement comparisons, assessments of risk of bias were made according to an adapted version of the Cochrane Collaboration's risk of bias tool. Comparisons of temperature measurements were expressed, where possible, as mean difference and 95% limits of agreement (LoA). Meta-analyses were not performed due to the lack of a common reference condition. For the survey of use, extracted information was summarised in text and tabular form. Results: For measurement comparisons, 21 studies were included. Results from these studies indicated minor (<0.5°C) to practically meaningful (>0.5°C) measurement bias within the subgroups of attachment type, applied pressure, environmental conditions, and sensor type. The 95% LoA were often within 1.0°C for in vivo studies and 0.5°C for physical models. For the survey of use, 172 studies were included. Details about T(skin) sensor setup were often poorly reported and, from those reporting setup information, it was evident that setups widely varied in terms of type of sensors, attachments, and locations used. Conclusions: Setup variables and conditions of use can influence the measured temperature from contact T(skin) sensors and thus key setup variables need to be appropriately considered and consistently reported.
format Online
Article
Text
id pubmed-5797625
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-57976252018-02-13 Skin Temperature Measurement Using Contact Thermometry: A Systematic Review of Setup Variables and Their Effects on Measured Values MacRae, Braid A. Annaheim, Simon Spengler, Christina M. Rossi, René M. Front Physiol Physiology Background: Skin temperature (T(skin)) is commonly measured using T(skin) sensors affixed directly to the skin surface, although the influence of setup variables on the measured outcome requires clarification. Objectives: The two distinct objectives of this systematic review were (1) to examine measurements from contact T(skin) sensors considering equilibrium temperature and temperature disturbance, sensor attachments, pressure, environmental temperature, and sensor type, and (2) to characterise the contact T(skin) sensors used, conditions of use, and subsequent reporting in studies investigating sports, exercise, and other physical activity. Data sources and study selection: For the measurement comparison objective, Ovid Medline and Scopus were used (1960 to July 2016) and studies comparing contact T(skin) sensor measurements in vivo or using appropriate physical models were included. For the survey of use, Ovid Medline was used (2011 to July 2016) and studies using contact temperature sensors for the measurement of human T(skin) in vivo during sport, exercise, and other physical activity were included. Study appraisal and synthesis methods: For measurement comparisons, assessments of risk of bias were made according to an adapted version of the Cochrane Collaboration's risk of bias tool. Comparisons of temperature measurements were expressed, where possible, as mean difference and 95% limits of agreement (LoA). Meta-analyses were not performed due to the lack of a common reference condition. For the survey of use, extracted information was summarised in text and tabular form. Results: For measurement comparisons, 21 studies were included. Results from these studies indicated minor (<0.5°C) to practically meaningful (>0.5°C) measurement bias within the subgroups of attachment type, applied pressure, environmental conditions, and sensor type. The 95% LoA were often within 1.0°C for in vivo studies and 0.5°C for physical models. For the survey of use, 172 studies were included. Details about T(skin) sensor setup were often poorly reported and, from those reporting setup information, it was evident that setups widely varied in terms of type of sensors, attachments, and locations used. Conclusions: Setup variables and conditions of use can influence the measured temperature from contact T(skin) sensors and thus key setup variables need to be appropriately considered and consistently reported. Frontiers Media S.A. 2018-01-30 /pmc/articles/PMC5797625/ /pubmed/29441024 http://dx.doi.org/10.3389/fphys.2018.00029 Text en Copyright © 2018 MacRae, Annaheim, Spengler and Rossi. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Physiology
MacRae, Braid A.
Annaheim, Simon
Spengler, Christina M.
Rossi, René M.
Skin Temperature Measurement Using Contact Thermometry: A Systematic Review of Setup Variables and Their Effects on Measured Values
title Skin Temperature Measurement Using Contact Thermometry: A Systematic Review of Setup Variables and Their Effects on Measured Values
title_full Skin Temperature Measurement Using Contact Thermometry: A Systematic Review of Setup Variables and Their Effects on Measured Values
title_fullStr Skin Temperature Measurement Using Contact Thermometry: A Systematic Review of Setup Variables and Their Effects on Measured Values
title_full_unstemmed Skin Temperature Measurement Using Contact Thermometry: A Systematic Review of Setup Variables and Their Effects on Measured Values
title_short Skin Temperature Measurement Using Contact Thermometry: A Systematic Review of Setup Variables and Their Effects on Measured Values
title_sort skin temperature measurement using contact thermometry: a systematic review of setup variables and their effects on measured values
topic Physiology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5797625/
https://www.ncbi.nlm.nih.gov/pubmed/29441024
http://dx.doi.org/10.3389/fphys.2018.00029
work_keys_str_mv AT macraebraida skintemperaturemeasurementusingcontactthermometryasystematicreviewofsetupvariablesandtheireffectsonmeasuredvalues
AT annaheimsimon skintemperaturemeasurementusingcontactthermometryasystematicreviewofsetupvariablesandtheireffectsonmeasuredvalues
AT spenglerchristinam skintemperaturemeasurementusingcontactthermometryasystematicreviewofsetupvariablesandtheireffectsonmeasuredvalues
AT rossirenem skintemperaturemeasurementusingcontactthermometryasystematicreviewofsetupvariablesandtheireffectsonmeasuredvalues