Dacroscintigraphy by Pediatric Dropper Technique: A User-friendly Instillation Procedure of Radiotracer

CONTEXT: Instillation or application of the radiotracer over the tear film is one of the important parts of dacroscintigraphy. Our study explains the value of an improvised dropper technique so that desired amount of radioactivity in desired volume can be instilled over the tear film. AIM: The aim o...

Descripción completa

Detalles Bibliográficos
Autores principales: Jain, Anurag, Aheer, Deepak, Arora, Amit, John, Arun Ravi, Solanki, KP, Kishore, Braj, Vishnoi, MG, Mahto, Abhishek, Prakash, Surya, Pandit, AG
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5798097/
https://www.ncbi.nlm.nih.gov/pubmed/29430114
http://dx.doi.org/10.4103/ijnm.IJNM_98_17
Descripción
Sumario:CONTEXT: Instillation or application of the radiotracer over the tear film is one of the important parts of dacroscintigraphy. Our study explains the value of an improvised dropper technique so that desired amount of radioactivity in desired volume can be instilled over the tear film. AIM: The aim of the study is to introduce a simple and convenient instillation method based on dropper technique for performing dacroscintigraphy. This improvised dropper technique can be used so that desired amount of radioactivity in desired volume can be instilled over the tear film. The objectives of this experiment are to measure the volume and activity of each drop from the dropper. SETTINGS AND DESIGN: Experiment of volume and activity measurement standardization was carried out in two parts. In the first part, we calculated the volume of each drop indirectly to standardize the volume of drop. In the second part, we standardized the activity in each drop by measuring it in a dose calibrator. SUBJECTS AND METHODS: In this study, we used a common pediatric dropper of approximately 1 ml capacity, radioactivity ((99m)TcO4 pertechnetate), sample vial (container), vial holder, a pair of nonsterile gloves, dose calibrator, etc., Experiments of volume and radioactivity standardization were carried out in two parts. The first part of experiment calculated volume of each drop indirectly, and in the second, we standardized the activity in each drop by measuring it in dose calibrator. Statistical analysis used: Analysis of variance test was used to calculate the correlation of readings by same individual as well as among the three individuals RESULTS: After analysis of result obtained, it was understood that there was no significant difference found in volume and activity of each drop in the readings recorded by same individual as well as among the three individuals. The calculated activity and observed activity were 86.64 and 79.16 μCi, respectively. The difference was only 8.63% lying within acceptable limits.