Efficacy of non-compliant balloon post-dilation in optimization of contemporary stents: A digital stent enhancement study
BACKGROUND: There is no evidence from randomized trials for the benefit of routine non-compliant balloon (NCB) post-dilation after stent deployment. Despite being the gold standard, intravascular ultrasound is infrequently performed due to time and cost constraints and a suitable alternative technol...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5801274/ https://www.ncbi.nlm.nih.gov/pubmed/29450169 http://dx.doi.org/10.1016/j.ijchv.2014.03.006 |
Sumario: | BACKGROUND: There is no evidence from randomized trials for the benefit of routine non-compliant balloon (NCB) post-dilation after stent deployment. Despite being the gold standard, intravascular ultrasound is infrequently performed due to time and cost constraints and a suitable alternative technology is required for routine assessment of stent expansion. The purpose of this study was to assess the contribution of NCB post-dilation in optimizing contemporary stents by using digital stent enhancement (DSE). METHODS: We treated 120 patients with stent insertion and assessed the stents with DSE before and after NCB use. Optimal expansion was defined as the minimum stent diameter (MSD) ≥ 90% of the nominal stent diameter, an adaptation of the MUSIC and POSTIT trial criteria. Stent deployment was performed at 12 atm pressure followed by routine NCB post-dilation at ≥ 14 atm. RESULTS: The mean reference diameter on QCA was 2.75 mm (SD 0.63) and mean stent diameter was 3.15 mm (SD 0.46). At a mean stent deployment pressure of 11.7 atm (SD 2.4), only 21% of stents were optimally expanded. After NCB inflation at a mean of 16.9 atm (SD 2.8), MSD increased by 0.26 mm (SD 0.24), optimal stent expansion increased from 21% to 58% and mean stent symmetry ratio increased from 0.83 to 0.87 (p < 0.0001). CONCLUSIONS: Contemporary stents are sub-optimally expanded in the majority of cases after standard deployment compared with nominal sizes. Adjunctive NCB post-dilation optimized an additional 37% of stents. DSE analysis can assist in qualitative and quantitative stent assessments and can potentially facilitate a selective NCB post-dilation strategy to achieve optimal stent expansion. |
---|