Cargando…

Negative interferences by calcium dobesilate in the detection of five serum analytes involving Trinder reaction-based assays

Previously, we reported the strong negative interference of calcium dobesilate, a vasoprotective agent, in creatinine assays involving the Trinder reaction. It is hypothesized that a similar effect occurs in the detection of uric acid (UA), total cholesterol (TC), triglycerides (TG), high-density li...

Descripción completa

Detalles Bibliográficos
Autores principales: Guo, Xiuzhi, Hou, Li’an, Yin, Yicong, Wu, Jie, Zhao, Fang, Xia, Liangyu, Cheng, Xinqi, Liu, Qian, Liu, Li, Xu, Ermu, Qiu, Ling
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5809042/
https://www.ncbi.nlm.nih.gov/pubmed/29432460
http://dx.doi.org/10.1371/journal.pone.0192440
Descripción
Sumario:Previously, we reported the strong negative interference of calcium dobesilate, a vasoprotective agent, in creatinine assays involving the Trinder reaction. It is hypothesized that a similar effect occurs in the detection of uric acid (UA), total cholesterol (TC), triglycerides (TG), high-density lipoprotein cholesterol (HDL-C), and low-density lipoprotein cholesterol (LDL-C). The interferences of calcium dobesilate during the detection of the five serum analytes were investigated on automated systems/analysers, and the effects were compared among eight different assay systems for each analyte. A calcium dobesilate standard was added into two sets of the blank serum pools of each analyte at final concentrations of 0, 2, 4, 8, 16, 32, and 64 μg/mL. The percentage deviation of each analyte value was calculated between each drug concentration and the drug-free samples. The clinically acceptable error levels for UA, TC, TG, HDL-C, and LDL-C were defined as ±4.87%, ±4.1%, ±9.57%, ±5.61%, and ±5.46%, respectively. The observed interference was concentration dependent for each analyte. In the presence of 16 μg/mL calcium dobesilate, which was within the therapeutic range, all seven Trinder reaction-based UA assay systems, two TG assay systems, two HDL-C assay systems and one TC assay system exhibited negative drug interferences. Calcium dobesilate negatively interferes with the detection of UA, TG, TC, and HDL-C in assay systems based on the Trinder reaction. The effect was most significant in UA and TG detection.