Setup in a clinical workflow and impact on radiotherapy routine of an in vivo dosimetry procedure with an electronic portal imaging device

High conformal techniques such as intensity-modulated radiation therapy and volumetric-modulated arc therapy are widely used in overloaded radiotherapy departments. In vivo dosimetric screening is essential in this environment to avoid important dosimetric errors. This work examines the feasibility...

Descripción completa

Detalles Bibliográficos
Autores principales: Li, Jie, Piermattei, Angelo, Wang, Pei, Kang, Shengwei, Xiao, Mingyong, Tang, Bin, Liao, Xiongfei, Xin, Xin, Grusio, Mattia, Orlandini, Lucia Clara
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5809064/
https://www.ncbi.nlm.nih.gov/pubmed/29432473
http://dx.doi.org/10.1371/journal.pone.0192686
_version_ 1783299526127255552
author Li, Jie
Piermattei, Angelo
Wang, Pei
Kang, Shengwei
Xiao, Mingyong
Tang, Bin
Liao, Xiongfei
Xin, Xin
Grusio, Mattia
Orlandini, Lucia Clara
author_facet Li, Jie
Piermattei, Angelo
Wang, Pei
Kang, Shengwei
Xiao, Mingyong
Tang, Bin
Liao, Xiongfei
Xin, Xin
Grusio, Mattia
Orlandini, Lucia Clara
author_sort Li, Jie
collection PubMed
description High conformal techniques such as intensity-modulated radiation therapy and volumetric-modulated arc therapy are widely used in overloaded radiotherapy departments. In vivo dosimetric screening is essential in this environment to avoid important dosimetric errors. This work examines the feasibility of introducing in vivo dosimetry (IVD) checks in a radiotherapy routine. The causes of dosimetric disagreements between delivered and planned treatments were identified and corrected during the course of treatment. The efficiency of the corrections performed and the added workload needed for the entire procedure were evaluated. The IVD procedure was based on an electronic portal imaging device. A total of 3682 IVD tests were performed for 147 patients who underwent head and neck, abdomen, pelvis, breast, and thorax radiotherapy treatments. Two types of indices were evaluated and used to determine if the IVD tests were within tolerance levels: the ratio R between the reconstructed and planned isocentre doses and a transit dosimetry based on the γ-analysis of the electronic portal images. The causes of test outside tolerance level were investigated and corrected and IVD test was repeated during subsequent fraction. The time needed for each step of the IVD procedure was registered. Pelvis, abdomen, and head and neck treatments had 10% of tests out of tolerance whereas breast and thorax treatments accounted for up to 25%. The patient setup was the main cause of 90% of the IVD tests out of tolerance and the remaining 10% was due to patient morphological changes. An average time of 42 min per day was sufficient to monitor a daily workload of 60 patients in treatment. This work shows that IVD performed with an electronic portal imaging device is feasible in an overloaded department and enables the timely realignment of the treatment quality indices in order to achieve a patient’s final treatment compliant with the one prescribed.
format Online
Article
Text
id pubmed-5809064
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-58090642018-02-28 Setup in a clinical workflow and impact on radiotherapy routine of an in vivo dosimetry procedure with an electronic portal imaging device Li, Jie Piermattei, Angelo Wang, Pei Kang, Shengwei Xiao, Mingyong Tang, Bin Liao, Xiongfei Xin, Xin Grusio, Mattia Orlandini, Lucia Clara PLoS One Research Article High conformal techniques such as intensity-modulated radiation therapy and volumetric-modulated arc therapy are widely used in overloaded radiotherapy departments. In vivo dosimetric screening is essential in this environment to avoid important dosimetric errors. This work examines the feasibility of introducing in vivo dosimetry (IVD) checks in a radiotherapy routine. The causes of dosimetric disagreements between delivered and planned treatments were identified and corrected during the course of treatment. The efficiency of the corrections performed and the added workload needed for the entire procedure were evaluated. The IVD procedure was based on an electronic portal imaging device. A total of 3682 IVD tests were performed for 147 patients who underwent head and neck, abdomen, pelvis, breast, and thorax radiotherapy treatments. Two types of indices were evaluated and used to determine if the IVD tests were within tolerance levels: the ratio R between the reconstructed and planned isocentre doses and a transit dosimetry based on the γ-analysis of the electronic portal images. The causes of test outside tolerance level were investigated and corrected and IVD test was repeated during subsequent fraction. The time needed for each step of the IVD procedure was registered. Pelvis, abdomen, and head and neck treatments had 10% of tests out of tolerance whereas breast and thorax treatments accounted for up to 25%. The patient setup was the main cause of 90% of the IVD tests out of tolerance and the remaining 10% was due to patient morphological changes. An average time of 42 min per day was sufficient to monitor a daily workload of 60 patients in treatment. This work shows that IVD performed with an electronic portal imaging device is feasible in an overloaded department and enables the timely realignment of the treatment quality indices in order to achieve a patient’s final treatment compliant with the one prescribed. Public Library of Science 2018-02-12 /pmc/articles/PMC5809064/ /pubmed/29432473 http://dx.doi.org/10.1371/journal.pone.0192686 Text en © 2018 Li et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Li, Jie
Piermattei, Angelo
Wang, Pei
Kang, Shengwei
Xiao, Mingyong
Tang, Bin
Liao, Xiongfei
Xin, Xin
Grusio, Mattia
Orlandini, Lucia Clara
Setup in a clinical workflow and impact on radiotherapy routine of an in vivo dosimetry procedure with an electronic portal imaging device
title Setup in a clinical workflow and impact on radiotherapy routine of an in vivo dosimetry procedure with an electronic portal imaging device
title_full Setup in a clinical workflow and impact on radiotherapy routine of an in vivo dosimetry procedure with an electronic portal imaging device
title_fullStr Setup in a clinical workflow and impact on radiotherapy routine of an in vivo dosimetry procedure with an electronic portal imaging device
title_full_unstemmed Setup in a clinical workflow and impact on radiotherapy routine of an in vivo dosimetry procedure with an electronic portal imaging device
title_short Setup in a clinical workflow and impact on radiotherapy routine of an in vivo dosimetry procedure with an electronic portal imaging device
title_sort setup in a clinical workflow and impact on radiotherapy routine of an in vivo dosimetry procedure with an electronic portal imaging device
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5809064/
https://www.ncbi.nlm.nih.gov/pubmed/29432473
http://dx.doi.org/10.1371/journal.pone.0192686
work_keys_str_mv AT lijie setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice
AT piermatteiangelo setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice
AT wangpei setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice
AT kangshengwei setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice
AT xiaomingyong setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice
AT tangbin setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice
AT liaoxiongfei setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice
AT xinxin setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice
AT grusiomattia setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice
AT orlandiniluciaclara setupinaclinicalworkflowandimpactonradiotherapyroutineofaninvivodosimetryprocedurewithanelectronicportalimagingdevice